[1]杨立旺,欧海峰,汪金龙,等.用于像素化瞬发伽马探测器的质子射程验证算法[J].中国医学物理学杂志,2025,42(3):281-287.[doi:10.3969/j.issn.1005-202X.2025.03.001]
 YANG Liwang,OU Haifeng,WANG Jinlong,et al.Proton beam range verification algorithm for pixelated prompt gamma-ray imaging detector[J].Chinese Journal of Medical Physics,2025,42(3):281-287.[doi:10.3969/j.issn.1005-202X.2025.03.001]
点击复制

用于像素化瞬发伽马探测器的质子射程验证算法()
分享到:

《中国医学物理学杂志》[ISSN:1005-202X/CN:44-1351/R]

卷:
42
期数:
2025年第3期
页码:
281-287
栏目:
医学放射物理
出版日期:
2025-03-20

文章信息/Info

Title:
Proton beam range verification algorithm for pixelated prompt gamma-ray imaging detector
文章编号:
1005-202X(2025)03-0281-07
作者:
杨立旺 1欧海峰 1汪金龙 2吴晓光 2贺子阳 2邹佳楠 2郑云 2李聪博 2关少雄 2施婧 2李金泽 2李韵秋 2 洪锐 2常浩恩 1王梦婷 1魏凯杰 1
1. 河南工业大学物理学院,河南 郑州 450001;2. 中国原子能科学研究院核物理研究所,北京 102413
Author(s):
YANG Liwang1 OU Haifeng1 WANG Jinlong2 WU Xiaoguang2 HE Ziyang2 ZOU Jian’an2 ZHENG Yun2 LI Congbo2 GUAN Shaoxiong2 SHI Jing2 LI Jinze2 LI Yunqiu2 HONG Rui2 CHANG Hao’en1 WANG Mengting1 WEI Kaijie1
1. School of Physics, Henan University of Technology, Zhengzhou 450001, China; 2. Nuclear Physics Institute, China Institute of Atomic Energy, Beijing 102413, China
关键词:
瞬发伽马质子治疗射程验证像素化
Keywords:
prompt gamma proton therapy range verification pixelated
分类号:
R318R811
DOI:
10.3969/j.issn.1005-202X.2025.03.001
文献标志码:
A
摘要:
在质子治疗中,瞬发伽马成像被认为是评估质子射程最有前途的方法之一。瞬发伽马成像探测器根据瞬发伽马成 像系统获得的瞬发伽马分布来评估质子射程,可实现质子束射程的高精度测量。本文为新开发的射程验证探测器原型: 像素化瞬发伽马成像探测器设计质子射程验证算法,该算法通过蒙特卡罗模拟验证该原型机在均匀介质的不同体模、不 同 能 量 下 的 射 程 估 计 准 确 度 。 结 果 表 明 ,该 质 子 射 程 验 证 算 法 得 出 的 射 程 估 计 精 度 与 布 拉 格 峰 安 全 裕 度 误 差 在 0.5 mm 之内,并且随着质子数量的增加,测量精度显著提高,表明该原型机算法用于质子射程验证是可行的。
Abstract:
In proton therapy, prompt gamma-ray imaging is considered as one of the most promising methods for assessing proton beam range. Prompt gamma-ray imaging detector evaluates the proton beam range based on the prompt gamma-ray distribution obtained by the prompt gamma-ray imaging system, which enables high-precision measurement of the proton beam range. Herein a proton beam range verification algorithm is designed for the newly developed prototype of the range verification detector (pixelated prompt gamma-ray imaging detector), which verifies the range estimation accuracy of the prototype for different phantoms and different energies of homogeneous media through Monte Carlo simulation. The results show that the accuracy of the proton beam range verification algorithm is within 0.5 mm of the safety margin error of the Bragg peak, and the measurement accuracy is significantly improved with the increase of the number of protons, indicating that the prototype algorithm is feasible for proton beam range verification.

相似文献/References:

[1]常城,王若曦,杜乙,等.一种新型质子治疗剂量递送系统的设计研究及模拟验证[J].中国医学物理学杂志,2019,36(7):745.[doi:DOI:10.3969/j.issn.1005-202X.2019.07.001]
 CHANG Cheng,WANG Ruoxi,DU Yi,et al.Design and verification of a new compact dose delivery system for proton therapy[J].Chinese Journal of Medical Physics,2019,36(3):745.[doi:DOI:10.3969/j.issn.1005-202X.2019.07.001]
[2]汪金龙,屈卫卫,谢树青,等. 扫描治疗头的蒙特卡罗模型研究[J].中国医学物理学杂志,2019,36(9):1001.[doi:DOI:10.3969/j.issn.1005-202X.2019.09.002]
 WANG Jinlong,QU Weiwei,XIE Shuqing,et al. Investigation of a Monte Carlo model of scanning nozzle[J].Chinese Journal of Medical Physics,2019,36(3):1001.[doi:DOI:10.3969/j.issn.1005-202X.2019.09.002]
[3]闫卓,徐榭,陈志.基于蒙特卡洛方法和数字体模的质子治疗设施屏蔽优化设计[J].中国医学物理学杂志,2020,37(12):1482.[doi:DOI:10.3969/j.issn.1005-202X.2020.12.002]
 YAN Zhuo,X. George XU,CHEN Zhi,et al.Shielding optimization of proton therapy facilities based on Monte Carlo simulations and voxelized human phantoms[J].Chinese Journal of Medical Physics,2020,37(3):1482.[doi:DOI:10.3969/j.issn.1005-202X.2020.12.002]
[4]崔相利,张建光,卢晓明,等.质子治疗技术的发展及其挑战[J].中国医学物理学杂志,2022,39(4):404.[doi:DOI:10.3969/j.issn.1005-202X.2022.04.002]
 CUI Xiangli,ZHANG Jianguang,et al.Development and challenges of proton therapy[J].Chinese Journal of Medical Physics,2022,39(3):404.[doi:DOI:10.3969/j.issn.1005-202X.2022.04.002]
[5]李国栋,王明,薛景硕,等.边缘场存在时磁共振图像引导质子治疗中束流轨迹校正与剂量分布[J].中国医学物理学杂志,2024,41(6):661.[doi:DOI:10.3969/j.issn.1005-202X.2024.06.001]
 LI Guodong,WANG Ming,XUE Jingshuo,et al.Beam trajectory correction and dose distribution in the presence of fringe fields in magnetic resonance imaging-guided proton therapy[J].Chinese Journal of Medical Physics,2024,41(3):661.[doi:DOI:10.3969/j.issn.1005-202X.2024.06.001]

备注/Memo

备注/Memo:
【收稿日期】2024-10-20 【基金项目】国 家 自 然 科 学 基 金(U1932209, 11975315, U1867210, 11905134) 【作者简介】杨立旺,硕士研究生,研究方向:粒子治疗中的瞬发伽马成 像,E-mail: yangliwang@sohu.com 【通信作者】欧 海 峰 ,副 教 授 ,研 究 方 向 :离 子 束 治 疗 技 术 基 础 研 究 , E-mail: ouhaifeng@haut.edu.com;吴晓光,研究员,研究方 向 :粒 子 治 疗 中 的 瞬 发 伽 马 成 像 ,E-mail: xingkong1973 @sina.com
更新日期/Last Update: 2025-03-27