Title:
Coil design and simulation analysis of magnetic induction hyperthermia in rats
文章编号:
1005-202X(2024)01-0101-08
作者:
郭思玮逯迈
兰州交通大学光电技术与智能控制教育部重点实验室, 甘肃 兰州 730000
Author(s):
GUO Siwei LU Mai
Key Lab of Optoelectronic Technology and Intelligent Control of Ministry of Education, Lanzhou Jiaotong University, Lanzhou 730000, China
关键词:
磁感应热疗线圈耦合场大鼠模型
Keywords:
Keywords: magnetic induction hyperthermia coil coupling field rat model
分类号:
R318;R730.59
DOI:
DOI:10.3969/j.issn.1005-202X.2024.01.015
文献标志码:
A
摘要:
选用亥姆霍兹线圈作为交变磁场发生装置,以大鼠脑部肿瘤为研究对象,建立相关模型,以Pennes生物传热方程为基础,在COMSOL仿真软件中进行电磁场分布和温度场分布计算,并分析影响磁感应热疗结果的因素。结果表明,大鼠肿瘤治疗区域磁场分布均匀,中心磁感应强度为12.847 mT,能达到进行磁感应热疗的条件;温度场分布符合要求,肿瘤区域升温明显,肿瘤中心温度可达46 ℃以上,肿瘤组织基本能达到治疗温度,能实现比较充分的治疗;线圈匝数、电流、半径及间距、磁场频率等参数均能影响磁感应热疗的效果。本研究结果可为磁感应热疗的临床应用和设备中的线圈设计提供参考。
Abstract:
Abstract: A Helmholtz coil is selected to generate the alternating magnetic field, and the relevant model is established with a rat brain tumor as the research object. Based on the Pennes bioheat transfer equation, the electromagnetic field distribution and temperature field distribution are calculated in COMSOL simulation software, and the factors affecting the outcome of magnetic induction hyperthermia are analyzed. The results show that both magnetic field distribution and temperature field distribution meet the requirements for magnetic induction hyperthermia. The magnetic field distribution in the tumor treatment area is uniform, and the central magnetic induction strength is 12.847 mT. The temperature rise in the tumor area is significant, and the temperature at the tumor center is 46℃ or above, basically reaching the treatment temperature. The therapeutic efficacy of magnetic induction hyperthermia is affected by the number of turns, current, radius and spacing, magnetic field frequency and other parameters. The study provides reference for the clinical application of magnetic induction hyperthermia and the coil design.

相似文献/References:

[1]段海军,金贵,白泽霖,等. 磁感应方法测量家兔脑出血过程中的拐点[J].中国医学物理学杂志,2017,34(11):1150.[doi:DOI:10.3969/j.issn.1005-202X.2017.11.013]
 DUAN Haijun,JIN Gui,BAI Zelin,et al. Turning point during the intracerebral hemorrhage of rabbit monitored with magnetic induction method[J].Chinese Journal of Medical Physics,2017,34(1):1150.[doi:DOI:10.3969/j.issn.1005-202X.2017.11.013]
[2]彭康强,戴心仪,刘辉明,等.自适应图像接收线圈与传统专用线圈在乳腺磁共振检查中的应用比较[J].中国医学物理学杂志,2023,40(9):1091.[doi:DOI:10.3969/j.issn.1005-202X.2023.09.006]
 PENG Kangqiang,DAI Xinyi,LIU Huiming,et al.Comparison of adaptive image receiving coil and traditional coil in breast magnetic resonance imaging[J].Chinese Journal of Medical Physics,2023,40(1):1091.[doi:DOI:10.3969/j.issn.1005-202X.2023.09.006]
[3]张恒,逯迈.磁感应热疗作用于胰腺癌模型热场分布[J].中国医学物理学杂志,2023,40(11):1408.[doi:DOI:10.3969/j.issn.1005-202X.2023.11.015]
 ZHANG Heng,LU Mai.Thermal field distribution of pancreatic cancer model treated with magnetic induction hyperthermia[J].Chinese Journal of Medical Physics,2023,40(1):1408.[doi:DOI:10.3969/j.issn.1005-202X.2023.11.015]

备注/Memo

备注/Memo:
【收稿日期】2023-09-06 【基金项目】国家自然科学基金(51567015,51867014) 【作者简介】郭思玮,硕士研究生,研究方向:磁感应热疗,E-mail: fangfanggsw@163.com 【通信作者】逯迈,博士,博士生导师,教授,研究方向:生物医学电磁学、电磁暴露安全评估,E-mail: mai.lu@hotmail.com
更新日期/Last Update: 2024-01-23