CT扫描电压与CT值-相对电子密度值转换曲线不匹配对放疗计划剂量的影响()
分享到:

《中国医学物理学杂志》[ISSN:1005-202X/CN:44-1351/R]

卷:
38卷
期数:
2021年第3期
页码:
312-316
栏目:
医学放射物理
出版日期:
2021-03-30

文章信息/Info

Title:
Effects of mismatch between CT scanning voltage and CT-RED conversion curve on radiotherapy dose
文章编号:
R318;R811
作者:
孔繁图1陈露茜1金锦辉2董洁1贺亚迪2李政欢1
1.中山大学附属第三医院肿瘤放射治疗科, 广东 广州 510630; 2.南方医科大学生物医学工程学院, 广东 广州 510515
Author(s):
KONG Fantu1 CHEN Luxi1 JIN Jinhui2 DONG Jie1 HE Yadi2 LI Zhenghuan1
1. Department of Radiation Oncology, The Third Affiliated Hospital of Sun Yat-sen University, Guangzhou 510630, China 2.School of Biomedical Engineering, South Medical University, Guangzhou 510515, China
关键词:
CT扫描电压CT值相对电子密度放疗计划剂量计算
Keywords:
Keywords: CT scanning voltage CT number relative electron density radiotherapy plan dose calculation
分类号:
R318;R811
DOI:
1005-202X(2021)03-0312-05
文献标志码:
A
摘要:
目的:探讨选取不同放疗定位CT扫描电压和CT值-相对电子密度值(CT-RED)转换曲线对放疗计划剂量计算的影响。方法:利用大孔径CT模拟定位机在80、100、120和140 kV电压下对CIRS 062M模体进行扫描,分别建立不同电压下的CT-RED转换曲线;在均匀石蜡模体中心嵌入高密度材料,通过选择不同CT扫描电压与CT-RED转换曲线的组合条件,在计划系统(TPS)设计不同照射角度的放疗计划。分别在高密度材料前、材料中和材料后选取3个兴趣点I1、I2、I3,统计高密度材料的Dmean、D2、D98,分析CT扫描电压与CT-RED转换曲线不匹配时对以上3个兴趣点剂量(DI1、DI2、DI3)和高密度材料Dmean、D2、D98的影响。结果:CT扫描电压与CT-RED转换曲线不匹配的组合条件下,放疗计划剂量有较大差异,其中Dmean、D98、DI2、DI3影响较大,差异具有统计学意义(均P<0.05)。结论:放疗患者CT定位的扫描电压需要与TPS计划设计时选取的CT-RED转换曲线匹配,否则会造成高密度组织中的剂量和射野出口处的剂量计算有较大差异,从而影响放疗计划的准确性。
Abstract:
Abstract: Objective To explore the effects of different voltages selected for computed tomography (CT) positioning and CT number-relative electron density (CT-RED) conversion curves on the calculation of radiotherapy dose. Methods Under the voltages of 80, 100, 120 and 140 kV, CIRS 062M phantom were scanned by a large-aperture CT simulator, thereby establishing CT-RED conversion curves under different voltages. A known high-density material was embedded in the center of a uniform paraffin phantom, and by setting a series of combinations of different voltages and CT-RED conversion curves, radiotherapy plans with different irradiation angles were established in treatment planning system. The variations of doses (DI1, DI2 and DI3) delivered to 3 points of interest (I1, I2 and I3) selected at the front, middle and rear of the high-density material and the changes in the Dmean, D2 and D98 of the high-density material were analyzed when voltages mismatched with CT-RED conversion curves. Results Under the mismatch between voltages and CT-RED conversion curves, the doses of radiotherapy plan were significantly different, among which Dmean, D98, DI2, DI3 were most affected, with statistical differences (all P<0.05). Conclusion The voltage for CT positioning needs to match with CT-RED conversion curve selected for radiotherapy planning. Otherwise, it will lead to great differences in the calculations of dose in the high-density material and the dose at the exit of radiation field, thus affecting radiotherapy accuracy.

相似文献/References:

[1]周丁屹,蒋大振,全红,等. 四维CT扫描参数和扫描模式对CT值的影响[J].中国医学物理学杂志,2016,33(9):919.[doi:10.3969/j.issn.1005-202X.2016.09.010]
 [J].Chinese Journal of Medical Physics,2016,33(3):919.[doi:10.3969/j.issn.1005-202X.2016.09.010]
[2]郑旭亮,邢晓汾,崔桐,等. 头颈肩定位体架对放射治疗剂量的影响[J].中国医学物理学杂志,2017,34(10):999.[doi:DOI:10.3969/j.issn.1005-202X.2017.10.006]
 [J].Chinese Journal of Medical Physics,2017,34(3):999.[doi:DOI:10.3969/j.issn.1005-202X.2017.10.006]
[3]吴先想,郭红博,费振乐,等. CT值相对电子密度转换关系影响因素及其对剂量计算的影响分析[J].中国医学物理学杂志,2019,36(2):157.[doi:DOI:10.3969/j.issn.1005-202X.2019.02.007]
 WU Xianxiang,GUO Hongbo,FEI Zhenle,et al. Influence factors of CT value-relative electron density conversion and its effects on dose calculation[J].Chinese Journal of Medical Physics,2019,36(3):157.[doi:DOI:10.3969/j.issn.1005-202X.2019.02.007]
[4]贺瑶瑶,汪立文,孙哲,等. 等效组织材料聚氯乙烯的医学成像特性[J].中国医学物理学杂志,2019,36(9):1091.[doi:DOI:10.3969/j.issn.1005-202X.2019.09.019]
 HE Yaoyao,,et al. Medical imaging properties of tissue equivalent material of polyvinyl chloride[J].Chinese Journal of Medical Physics,2019,36(3):1091.[doi:DOI:10.3969/j.issn.1005-202X.2019.09.019]
[5]麻祯媛,邹莹,廖玉荣,等.不同剂量CT肺动脉成像评估肺动脉高压严重程度[J].中国医学物理学杂志,2020,37(11):1403.[doi:DOI:10.3969/j.issn.1005-202X.2020.11.012]
 MA Zhenyuan,ZOU Ying,LIAO Yurong,et al.Assessment of the severity of pulmonary hypertension by CT pulmonary artery imaging with different doses[J].Chinese Journal of Medical Physics,2020,37(3):1403.[doi:DOI:10.3969/j.issn.1005-202X.2020.11.012]

备注/Memo

备注/Memo:
【收稿日期】2020-09-22 【作者简介】孔繁图,助理工程师,研究方向:放射治疗物理,E-mail: mralonekft@163.com 【通信作者】李政欢,主管技师,研究方向:放射治疗物理,E-mail:huanliz2819@163.com
更新日期/Last Update: 2021-03-30