Effect of rotational mechanical errors on dose distribution of volumetric modulated arc therapy for multiple intracranial metastases
文章编号:
1005-202X(2020)01-0038-06
作者:
邓官华1林正茂2陈丽霞3戴鹏1张平1罗龙辉1罗日顺1赖名耀1山常国1黄庆1蔡林波1
1.广东三九脑科医院肿瘤综合治疗中心, 广东 广州 510510; 2.中山大学孙逸仙纪念医院, 广东 广州 510120; 3.南方医科大学生物医学工程学院, 广东 广州 510515
Author(s):
DENG Guanhua1 LIN Zhengmao2 CHEN Lixia3DAI Peng1 ZHANG Ping1 LUO Longhui1 LUO Rishun1 LAI Mingyao1 SHAN Changguo1 HUANG Qing1 CAI Linbo1
1. Department of Oncology, Guangdong Sanjiu Brain Hospital, Guangzhou 510510, China; 2. Sun Yat-sen Memorial Hospital, Sun Yat-sen University, Guangzhou 510120, China; 3. School of Biomedical Engineering, Southern Medical University, Guangzhou 510515, China
关键词:
多发脑转移瘤容积旋转调强机械旋转误差单中心
Keywords:
Keywords: multiple intracranial metastases volumetric modulated arc therapy rotational mechanical error single isocenter
分类号:
R811.1;R312
DOI:
DOI:10.3969/j.issn.1005-202X.2020.01.008
文献标志码:
A
摘要:
目的:探讨直线加速器的机械旋转误差对单中心多发转移瘤VMAT计划剂量分布的影响。方法:随机选取21例多发脑转移瘤患者,假定患者均采用非共面VMAT放疗计划,分别将治疗床、准直器的角度旋转偏移[±]0.5°、[±]1.0°、[±]1.5°、[±]2.0°,并通过Eclipse 13.6治疗计划系统模拟机械旋转偏移误差在多发脑转移瘤VMAT计划中对剂量分布的影响。记录并分析不同治疗床、准直器角度旋转偏移下靶区的适形度指数(CI)、剂量梯度跌落指数(GI)、剂量均匀性指数(HI)以及危及器官的最大剂量。结果:当治疗床及准直器角度旋转偏移大于0.5°时,靶区的CI、GI(治疗床除外)及HI指数差异具有统计学意义(P<0.05);危及器官的最大剂量差异无统计学意义(P>0.05)。结论:在设计多发脑转移VMAT计划时应考虑等中心与靶区之间的距离,六维床旋转误差校正阈值为0.5°更为合理。
Abstract:
Abstract: Objective To investigate the effect of rotational mechanical errors of linear accelerator on the dose distribution of single-isocenter volumetric modulated arc therapy (VMAT) for multiple intracranial metastases. Methods A total of 21 patients with multiple intracranial metastases were enrolled in the study. Assuming that patients were treated with non-coplanar VMAT, the treatment couch and collimator angle were shifted by ±0.5°, ±1.0°, ±1.5° and ±2.0°, respectively. Eclipse 13.6 treatment planning system was used to study the effects of the rotational mechanical errors on dose distribution of VMAT for multiple intracranial metastases. With different rotational errors, the conformity index, gradient index and homogeneity index of target areas as well as the maximum dose of organs-at-risk were recorded and analyzed. Results There were significant differences in the conformity index, gradient index (except for couch), and homogeneity index of target areas (P<0.05) when the rotational deviations of treatment couch and collimator angle were greater than 0.5°. No significant difference was found in the maximum dose of organs-at-risk (P>0.05). Conclusion In the planning of VMAT for multiple intracranial metastases, the distance between the isocenter and target areas should be taken into consideration and the reasonable correction threshold for six-dimensional couch rotational errors should be 0.5°.

相似文献/References:

[1]孙小喆,孟慧鹏,史孝伟,等.胸部肿瘤容积旋转调强放射治疗摆位误差分析[J].中国医学物理学杂志,2015,32(01):34.[doi:10.3969/j.issn.1005-202X.2015.01.009]
[2]孙晓欢,周咏春,谭丽娜,等.容积旋转调强与固定野调强在宫颈癌术后放疗中的剂量学比较[J].中国医学物理学杂志,2014,31(01):4604.[doi:10.3969/j.issn.1005-202X.2014.01.003]
[3]张俊俊,邱小平,李奇欣,等.ArcCheck系统在鼻咽癌容积旋转调强剂量验证中的应用[J].中国医学物理学杂志,2014,31(05):5136.[doi:10.3969/j.issn.1005-202X.2014.05.009]
[4]张矛,金海国,苏清秀,等.肺癌静态调强与容积旋转调强放射治疗间比较[J].中国医学物理学杂志,2013,30(05):4364.[doi:10.3969/j.issn.1005-202X.2013.05.006]
[5]谭丽娜,孙晓欢,马奎,等.三维剂量验证系统Delta4在容积旋转调强计划剂量验证中的应用[J].中国医学物理学杂志,2013,30(06):4497.[doi:10.3969/j.issn.1005-202X.2013.06.007]
[6]王沛沛,刘哲铭,李彩虹,等.容积旋转调强与固定野动态调强在直肠癌新辅助放疗中的剂量学比较[J].中国医学物理学杂志,2015,32(06):901.[doi:doi:10.3969/j.issn.1005-202X.2015.06.030]
 [J].Chinese Journal of Medical Physics,2015,32(1):901.[doi:doi:10.3969/j.issn.1005-202X.2015.06.030]
[7]李丹明,王黎,孙新臣,等.中晚期宫颈癌适形调强放疗与容积旋转调强技术临床效果比较[J].中国医学物理学杂志,2016,33(5):478.[doi:10.3969/j.issn.1005-202X.2016.05.010]
 [J].Chinese Journal of Medical Physics,2016,33(1):478.[doi:10.3969/j.issn.1005-202X.2016.05.010]
[8]姚升宇,颜歌,赵国旗,等.容积旋转调强与固定野动态调强低剂量区比较[J].中国医学物理学杂志,2016,33(5):538.[doi:10.3969/j.issn.1005-202X.2016.05.022]
 [J].Chinese Journal of Medical Physics,2016,33(1):538.[doi:10.3969/j.issn.1005-202X.2016.05.022]
[9]高兴旺,Wang Yang,梁志强,等.快速容积旋转调强治疗系统的验收测试[J].中国医学物理学杂志,2016,33(6):541.[doi:DOI:10.3969/j.issn.1005-202X.2016.06.001]
 [J].Chinese Journal of Medical Physics,2016,33(1):541.[doi:DOI:10.3969/j.issn.1005-202X.2016.06.001]
[10]黎旦,宾石珍,程品晶,等.非晶硅电子射野影像系统与Arc CHECK在直肠癌容积旋转调强剂量验证中的应用[J].中国医学物理学杂志,2016,33(7):674.[doi:10.3969/j.issn.1005-202X.2016.07.006]
 [J].Chinese Journal of Medical Physics,2016,33(1):674.[doi:10.3969/j.issn.1005-202X.2016.07.006]

备注/Memo

备注/Memo:
【收稿日期】2019-07-14 【基金项目】广东三九脑科医院新技术项目(2017012) 【作者简介】邓官华,硕士,中级职称,研究方向:肿瘤早期发现及放射治疗,E-mail: gh.deng@foxmail.com 【通信作者】蔡林波,硕士,主任医师,研究方向:肿瘤放射治疗及免疫治疗,E-mail: cailinbo999@163.com
更新日期/Last Update: 2020-01-14