[1]袁翠云,陈辛元,刘晨彬,等.MR模拟定位机的验收测试-依据指南的实践与结果分析[J].中国医学物理学杂志,2024,41(10):1199-1205.[doi:DOI:10.3969/j.issn.1005-202X.2024.10.002]
 YUAN Cuiyun,CHEN Xinyuan,LIU Chenbin,et al.Acceptance testing for MR simulator: guideline-based practice and result analysis[J].Chinese Journal of Medical Physics,2024,41(10):1199-1205.[doi:DOI:10.3969/j.issn.1005-202X.2024.10.002]
点击复制

MR模拟定位机的验收测试-依据指南的实践与结果分析()
分享到:

《中国医学物理学杂志》[ISSN:1005-202X/CN:44-1351/R]

卷:
41卷
期数:
2024年第10期
页码:
1199-1205
栏目:
医学放射物理
出版日期:
2024-10-25

文章信息/Info

Title:
Acceptance testing for MR simulator: guideline-based practice and result analysis
文章编号:
1005-202X(2024)10-1199-07
作者:
袁翠云1陈辛元2刘晨彬1李阳1权恩卓1戴建荣2
1.国家癌症中心/国家肿瘤临床医学研究中心/中国医学科学院北京协和医学院肿瘤医院深圳医院放疗科, 广东 深圳 518116; 2.国家癌症中心/国家肿瘤临床医学研究中心/中国医学科学院北京协和医学院肿瘤医院放疗科, 北京 100021
Author(s):
YUAN Cuiyun1 CHEN Xinyuan2 LIU Chenbin1 LI Yang1 QUAN Enzhuo1 DAI Jianrong2
1. Department of Radiation Oncology, National Cancer Center/National Clinical Research Center for Cancer/Cancer Hospital & Shenzhen Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Shenzhen 518116, China 2. Department of Radiation Oncology, National Cancer Center/National Clinical Research Center for Cancer/Cancer Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing 100021, China
关键词:
磁共振模拟定位验收测试3.0T MR放射治疗
Keywords:
Keywords: magnetic resonance simulator acceptance testing 3.0T magnetic resonance radiotherapy
分类号:
R318;R811.1
DOI:
DOI:10.3969/j.issn.1005-202X.2024.10.002
文献标志码:
A
摘要:
目的:磁共振模拟定位机(MR Sim)是新类型的放疗模拟定位机。验收测试是MR Sim投入使用前的重要质量保证工作。该文报告一台MR Sim的验收测试实施细节和结果分析。方法:依据新近发布的AAPM TG284报告和NCC/T-RT 002-2023指南,制定验收测试方案。针对MR Sim的磁场屏蔽、成像系统基础功能、图像质量指标、射频线圈验收、图像几何精度、放疗外置激光灯、定位平面床及图像传输等,使用ACR(American College of Radiology)大号模体、几何畸变测量模体等质控设备进行测试。结果:MR Sim验收指标中的屏蔽测试在150 MHz的频率下,屏蔽效能均值为105 dB;8项图像质量指标:几何精度、层位置精度、层厚精度、图像均匀性、伪影比、信噪比、高对比空间分辨率和低对比分辨率,均在容许误差范围内;图像几何精度测量在25 cm感兴趣范围内,最大几何畸变<2 mm;激光灯及床的误差均<1 mm,床水平度<1°;射频线圈测试及图像传输测试满足要求。结论:MR Sim是高精度的复杂设备,为确保其在放疗中的精准应用,投入使用前需结合本单位MR Sim的具体特点,遵循指南规范,进行验收测试具体方案的详细设计和执行。
Abstract:
Abstract: Objective Magnetic resonance simulator (MR Sim) is a novel type of simulation equipment utilized in radiotherapy. Acceptance testing is an essential quality assurance procedure prior to the clinical use of the MR Sim. The report provides the detailed procedures and result analysis of acceptance testing for an MR Sim. Methods The acceptance testing scheme was developed following the recently published AAPM TG284 report and the NCC/T-RT 002-2023 guidelines. Quality control equipments such as ACR (American College of Radiology) large phantom and geometric distortion measurement phantom were used for evaluating various aspects of the MR Sim, including the effectiveness of shielding, the functionality of imaging system, the image quality, the performance of radio frequency coils, the geometric accuracy of large field imaging, the precision of external laser markings, the couch movement accuracy, and the image transmission accuracy. Results The shielding effectiveness at a frequency of 150 MHz exhibited an average value of 105 dB. All of 8 image quality indices, namely geometric accuracy, slice position accuracy, slice thickness accuracy, image uniformity, artifact ratio, signal-to-noise ratio, high-contrast spatial resolution, and low-contrast resolution, fell within recommended tolerances. The maximum geometric distortion observed across a 25 cm field of view was less than 2 mm. The errors in external laser markings and couch movement accuracy were both less than 1 mm. The couch levelness was less than 1°. Both radio frequency coils and image transmission passed the required tests. Conclusion MR Sim is high-precision and complex. To ensure its precise application in radiotherapy, the acceptance testing for an MR Sim should be meticulously designed and executed following the established guidelines and accounting for its unique performance characteristics.

相似文献/References:

[1]高兴旺,Wang Yang,梁志强,等.快速容积旋转调强治疗系统的验收测试[J].中国医学物理学杂志,2016,33(6):541.[doi:DOI:10.3969/j.issn.1005-202X.2016.06.001]
 [J].Chinese Journal of Medical Physics,2016,33(10):541.[doi:DOI:10.3969/j.issn.1005-202X.2016.06.001]
[2]苏坤普,周德力,李林山,等.M6型射波刀立体定向放疗系统的验收测试[J].中国医学物理学杂志,2023,40(7):814.[doi:1005-202X(2023)07-0814-08]
 SU Kunpu,ZHOU Deli,LI Linshan,et al.Acceptance test of M6 CyberKnife stereotactic radiotherapy system[J].Chinese Journal of Medical Physics,2023,40(10):814.[doi:1005-202X(2023)07-0814-08]

备注/Memo

备注/Memo:
【收稿日期】2024-05-12 【基金项目】国家自然科学基金(12005301);广东省基础与应用基础研究基金(2022A1515012456);深圳市医疗卫生三名工程项目(SZSM201612063);深圳市医学研究专项资金(B022300257);中国医学科学院肿瘤医院深圳医院院内科研课题(E010321015,E010322028,SZ2020QN013) 【作者简介】袁翠云,硕士,研究实习员,研究方向:放射物理,E-mail: cuiyun.yuan@hotmail.com 【通信作者】权恩卓,博士,研究员,研究方向:放射物理,E-mail: quanenzhuo@cicams-sz.org.cn