[1]陈彦宇,柏朋刚,陈榕钦,等.剂量预测引导的鼻咽癌放疗计划质量定量评估方法[J].中国医学物理学杂志,2022,39(9):1076-1082.[doi:DOI:10.3969/j.issn.1005-202X.2022.09.004]
 CHEN Yanyu,BAI Penggang,CHEN Rongqin,et al.Dose prediction-guided quantitative assessment of the quality radiotherapy plan for nasopharyngeal cancer[J].Chinese Journal of Medical Physics,2022,39(9):1076-1082.[doi:DOI:10.3969/j.issn.1005-202X.2022.09.004]
点击复制

剂量预测引导的鼻咽癌放疗计划质量定量评估方法()
分享到:

《中国医学物理学杂志》[ISSN:1005-202X/CN:44-1351/R]

卷:
39卷
期数:
2022年第9期
页码:
1076-1082
栏目:
医学放射物理
出版日期:
2022-11-02

文章信息/Info

Title:
Dose prediction-guided quantitative assessment of the quality radiotherapy plan for nasopharyngeal cancer
文章编号:
1005-202X(2022)09-1076-07
作者:
陈彦宇1柏朋刚2陈榕钦2邱小平1陈济鸿2戴艺涛2全科润1周益民1
1.南华大学核科学技术学院, 湖南 衡阳 421001; 2.福建省肿瘤医院放疗科, 福建 福州 350014
Author(s):
CHEN Yanyu1 BAI Penggang2 CHEN Rongqin2 QIU Xiaoping1 CHEN Jihong2 DAI Yitao2 QUAN Kerun1 ZHOU Yimin1
1. School of Nuclear Science and Technology, University of South China, Hengyang 421001, China 2. Department of Radiation Oncology, Fujian Cancer Hospital, Fuzhou 350014, China
关键词:
鼻咽癌剂量预测人工神经网络计划质量测度
Keywords:
Keywords: nasopharyngeal carcinoma dose prediction artificial neural network plan quality metrics
分类号:
R739.6;R811.1
DOI:
DOI:10.3969/j.issn.1005-202X.2022.09.004
文献标志码:
A
摘要:
目的:利用人工神经网络模型对鼻咽癌计划进行剂量预测,根据剂量预测值建立计划质量测度标准(PQM),实现放疗计划个性化定量评估。方法:回顾性分析于福建省肿瘤医院进行放射治疗的鼻咽癌计划114例。提取25项危及器官(OAR)与靶区之间的几何空间关系特征,训练(81例)得到基于人工神经网络的剂量预测模型并测试(23例)验证其准确性。分别用基于剂量限值建立PQM和基于剂量预测值建立PQM两种评估方法对10例临床通过计划进行定量评估,讨论两种评估方法的合理性。结果:11项鼻咽癌主要OAR剂量学指标,预测值与实际值的剂量相关总体平均差值为(-0.07±4.55) Gy,体积相关总体平均差值为-1.06%±3.80%,预测准确性可达90%。10例临床通过的鼻咽癌计划,基于剂量限值建立PQM方法对病例4评估为不合格,其余病例合格。基于剂量预测值建立PQM方法对病例9评估为不合格,其余病例合格。结论:剂量预测引导的鼻咽癌计划质量定量评估方法可以反映计划是否存在继续优化的空间,且该评估方法克服了剂量限值未考虑病例特异性的缺陷,能更科学合理地对放疗计划进行定量评估。
Abstract:
Abstract: Objective To carry out the dose prediction for nasopharyngeal carcinoma radiotherapy using artificial neural network model, and to establish plan quality metrics (PQM) based on dose prediction for achieving personalized quantitative assessment of radiotherapy plan quality. Methods A total of 114 cases of nasopharyngeal carcinoma treated with radiotherapy in Fujian Cancer Hospital were analyzed retrospectively. After extracting 25 geometric spatial relationship features between organs-at-risk and target area, 81 cases were used for training and obtaining an artificial neural network-based dose prediction model, and the other 23 cases were used for validating the accuracy of the prediction model. PQM were developed based on dose limits and dose prediction, separately, for the quantitative assessment of 10 clinical cases, and finally, the rationalities of the two methods were discussed. Results For the 11 main OAR dosimetric indicators of nasopharyngeal carcinoma, the overall mean difference between predictive and actual values was (-0.07±4.55) Gy for dose-related indicators and (-1.06±3.80)% for volume-related indicators and the prediction accuracy reached 90%. For the 10 clinically approved nasopharyngeal carcinoma plans, the method of establishing PQM based on dose limits was assessed as failing for case 4 and passing for the remaining cases, while the method of establishing PQM based on dose prediction was assessed as failing for case 9 and passing for the remaining cases. Conclusion The dose prediction-guided quantitative assessment of the quality radiotherapy plan for nasopharyngeal cancer can reflect whether the plan can be further optimized. Moreover, the dose prediction-guided method overcomes the shortcomings of the dose limits-based assessment which do not take case specificity into account, and can evaluate the radiotherapy plan more scientifically and reasonably.

相似文献/References:

[1]陈广涛,汪 淇,刘亚强,等.TrueBeam加速器射野准直器跟随功能在鼻咽癌固定野调强放疗和容积旋转调强放疗中的应用[J].中国医学物理学杂志,2014,31(03):4865.[doi:10.3969/j.issn.1005-202X.2014.03.005]
[2]潘宇辉,陈榕钦,柏朋刚,等.鼻咽癌调强放射治疗前后腮腺体积及空间几何位置变化研究[J].中国医学物理学杂志,2014,31(05):5120.[doi:10.3969/j.issn.1005-202X.2014.05.005]
[3]张俊俊,邱小平,李奇欣,等.ArcCheck系统在鼻咽癌容积旋转调强剂量验证中的应用[J].中国医学物理学杂志,2014,31(05):5136.[doi:10.3969/j.issn.1005-202X.2014.05.009]
[4]桑 勇,陈维军,狄小云,等.多叶光栅叶片宽度对鼻咽癌剂量学的影响[J].中国医学物理学杂志,2014,31(06):5240.[doi:10.3969/j.issn.1005-202X.2014.06.004]
[5]黄利生,庄婷婷,伍方财,等.鼻咽癌调强放疗Ib区剂量分析[J].中国医学物理学杂志,2014,31(06):5265.[doi:10.3969/j.issn.1005-202X.2014.06.010]
[6]易金玲,金献测,周永强,等.鼻咽癌IMRT与VMAT治疗的计划与剂量验证比较研究[J].中国医学物理学杂志,2013,30(01):3859.[doi:10.3969/j.issn.1005-202X.2013.01.006]
[7]金献测,易金玲,周永强,等.鼻咽癌同步加量调强放射治疗新的计划质量评估指标研究[J].中国医学物理学杂志,2013,30(02):3978.[doi:10.3969/j.issn.1005-202X.2013.02.004]
[8]蒋晓芹,段宝风,艾平,等.基于图谱库的自动轮廓勾画软件(ABAS)在鼻咽癌调强放疗中的应用[J].中国医学物理学杂志,2013,30(02):3997.[doi:10.3969/j.issn.1005-202X.2013.02.008]
[9]张达光,蒋胜鹏,李智华,等.鼻咽癌调强放疗中解剖结构改变及剂量分布变化的研究[J].中国医学物理学杂志,2013,30(04):4239.[doi:10.3969/j.issn.1005-202X.2013.04.006]
[10]杨海燕,王军良,周振山,等.AAA算法和PBC算法在鼻咽癌调强放疗中剂量分布的比较[J].中国医学物理学杂志,2013,30(04):4245.[doi:10.3969/j.issn.1005-202X.2013.04.008]

备注/Memo

备注/Memo:
【收稿日期】2022-02-19 【基金项目】福建省自然科学基金引导性项目(2015Y0010);福建医科大学启航基金(2019QH1193);福建省卫生健康科技项目(2018-ZQN-19) 【作者简介】陈彦宇,在读研究生,研究方向:医学物理,E-mail: 857739933@qq.com 【通信作者】邱小平,教授,研究方向:放射诊疗与医学剂量,E-mail: nh6651@163.com;柏朋刚,高级工程师,研究方向:医学图像处理,E-mail: littwind@sina.cn
更新日期/Last Update: 2022-09-27