[1]李师,罗焕丽,谭霞,等.6 MV光子束放射治疗中发泡胶的剂量学影响[J].中国医学物理学杂志,2021,38(4):397-403.[doi:DOI:10.3969/j.issn.1005-202X.2021.04.001]
 LI Shi,LUO Huanli,TAN Xia,et al.Dosimetric effects of styrofoam in 6 MV photon beam radiotherapy[J].Chinese Journal of Medical Physics,2021,38(4):397-403.[doi:DOI:10.3969/j.issn.1005-202X.2021.04.001]
点击复制

6 MV光子束放射治疗中发泡胶的剂量学影响()
分享到:

《中国医学物理学杂志》[ISSN:1005-202X/CN:44-1351/R]

卷:
38卷
期数:
2021年第4期
页码:
397-403
栏目:
医学放射物理
出版日期:
2021-04-29

文章信息/Info

Title:
Dosimetric effects of styrofoam in 6 MV photon beam radiotherapy
文章编号:
1005-202X(2021)04-0397-07
作者:
李师罗焕丽谭霞蒙万里李博李广申正文钟明松王颖靳富
重庆大学附属肿瘤医院肿瘤放射治疗中心, 重庆 400030
Author(s):
LI Shi LUO Huanli TAN Xia MENG Wanli LI Bo LI Guang SHEN Zhengwen ZHONG Mingsong WANG Ying JIN Fu
Radiation Oncology Center, Chongqing University Cancer Hospital, Chongqing 400030, China
关键词:
发泡胶皮肤剂量固定装置剂量衰减放射治疗
Keywords:
Keywords: styrofoam skin dose immobilization device dose attenuation radiotherapy
分类号:
R815.6
DOI:
DOI:10.3969/j.issn.1005-202X.2021.04.001
文献标志码:
A
摘要:
【摘要】目的:评估发泡胶在6 MV光子束放疗中的剂量学影响。方法:制作10块2 cm厚的发泡胶模体,获取CT影像、导入放疗计划系统(TPS),计算其平均线性衰减系数,并用电离室进行测量验证。在固体等效水上表面叠加上述模体成2、4、6、8、10 cm厚度,计算并测量不同厚度发泡胶的剂量学影响。随机选取8例由发泡胶辅助固定的鼻咽癌患者,制定9野均分调强计划,分别计算包含与不包含发泡胶的剂量分布。结果:TPS计算得到的平均线性衰减系数为0.16×10-2 cm-1。发泡胶厚度>4 cm时,γ通过率(1%/1 mm)<90%,且厚度越厚通过率越低。相较于剂量跌落区,发泡胶对剂量建成区的影响更大,仅加入2 cm发泡胶时,表面剂量由16.22%增加到63.73%。发泡胶使靶区PTVnd+nx、PTV1、PTV2的平均剂量分别从(72.61±0.98) Gy减少到(72.12±1.00) Gy、(69.11±0.79) Gy减少到(68.72±0.77) Gy、(62.61±1.04) Gy减少到(62.22±1.03) Gy,100%处方剂量靶区覆盖率分别从0.97%±0.01%降低到0.95%±0.03%、0.98%±0.01%降低到0.97%±0.01%、0.94%±0.04%降低到0.92%±0.05%,总剂量γ通过率(1%/1 mm)为94.06%±0.86%。结论:发泡胶厚度越厚对剂量扰动越大,并导致建成区和表面剂量的显著增加,其衰减作用降低了靶区的平均剂量和处方剂量覆盖率。在临床应用时建议发泡胶厚度不超过4 cm,并应将发泡胶勾画在外轮廓中以评估皮肤剂量是否安全。
Abstract:
Abstract: Objective To evaluate the dosimetric effects of styrofoam in 6 MV photon beam radiotherapy. Methods Ten pieces of styrofoam phantoms with 2 cm thick were made, and their CT images were imported into treatment planning system for calculating the average linear attenuation coefficient which was then verified by ionization chamber measurement. The styrofoam phantoms of 2, 4, 6, 8, 10 cm thick which were formed by superimposing one over one another was placed on the surface of a solid equivalent water phantom to estimate the dosimetric effects caused by Styrofoam of different thicknesses. Furthermore, the doses distributions of 9-field intensity-modulated radiotherapy plan for 8 nasopharyngeal cancer patients immobilized with Styrofoam were calculated in the case of body structure contour with styrofoam or not. Results The average linear attenuation coefficient calculated by treatment planning system was (0.16×10-2) cm-1. The gamma passing rate (1% /1 mm) was less than 90% when styrofoam thickness was larger than 4 cm, and the thicker the Styrofoam was, the lower the gamma passing rate was. Furthermore, styrofoam had larger effects on dose built-up area than dose-drop area, and the relative surface dose was increased from 16.22% to 63.73% when only a piece of Styrofoam of 2 cm thick was placed on the surface of a solid equivalent water phantom. Due to styrofoam, the mean doses of PTVnd+nx, PTV1 and PTV2 were reduced from (72.61±0.98) Gy to (72.12±1.00) Gy, (69.11±0.79) Gy to (68.72±0.77) Gy, (62.61±1.04) Gy to (62.22±1.03) Gy, respectively, and the corresponding 100% prescription dose coverage rate was decreased from 0.97%±0.01% to 0.95%±0.03%, 0.98%±0.01% to 0.97±0.01%, 0.94%±0.04% to 0.92%±0.05%, respectively. The gamma passing rate (1%/1 mm) of overall dose was 94.06%±0.86%. Conclusion The dosimetric effects is increasing with the increasing of styrofoam thickness, which led to a significant dose increase in dose built-up area and surface dose. The attenuation effects of styrofoam reduce both mean dose sand 100% prescription dose coverage rates at target areas. It is recommended that the thickness of styrofoam should no exceed 4 cm, and that styrofoam should be included in body structure contour to assess whether the skin dose is safe.

相似文献/References:

[1]刘巍巍,王美娇,杨敬贤,等. 热塑膜联合发泡胶固定技术在宫颈癌放疗中的应用价值[J].中国医学物理学杂志,2019,36(6):636.[doi:DOI:10.3969/j.issn.1005-202X.2019.06.004]
 LIU Weiwei,WANG Meijiao,YANG Jingxian,et al. Application value of thermoplastic film combined with styrofoam fixation technology in radiotherapy for cervical cancer[J].Chinese Journal of Medical Physics,2019,36(4):636.[doi:DOI:10.3969/j.issn.1005-202X.2019.06.004]

备注/Memo

备注/Memo:
【收稿日期】2020-11-23 【基金项目】国家自然科学基金(11575038, 11805025, 81972857);2018年重庆市院士牵头科技创新引导专项 【作者简介】李师,初级工程师,研究方向:放射医学物理,E-mail: 2272482753@qq.com 【通信作者】靳富,高级工程师,研究方向:放射医学物理,E-mail: jfazj@126.com
更新日期/Last Update: 2021-04-29