[1]肖亮杰,方涌文,王宇留,等.基于锥形束CT分析食管癌和肺癌患者摆位误差[J].中国医学物理学杂志,2022,39(10):1189-1193.[doi:DOI:10.3969/j.issn.1005-202X.2022.10.001]
 XIAO Liangjie,FANG Yongwen,et al.Analysis of setup errors in patients with esophageal or lung cancer using CBCT[J].Chinese Journal of Medical Physics,2022,39(10):1189-1193.[doi:DOI:10.3969/j.issn.1005-202X.2022.10.001]
点击复制

基于锥形束CT分析食管癌和肺癌患者摆位误差()
分享到:

《中国医学物理学杂志》[ISSN:1005-202X/CN:44-1351/R]

卷:
39卷
期数:
2022年第10期
页码:
1189-1193
栏目:
医学放射物理
出版日期:
2022-11-02

文章信息/Info

Title:
Analysis of setup errors in patients with esophageal or lung cancer using CBCT
文章编号:
1005-202X(2022)10-1189-05
作者:
肖亮杰12方涌文1王宇留1林晓生1许森奎12姚文燕1赵磊1方键蓝1
1.中山大学肿瘤防治中心放疗科, 广东 广州 510060; 2.广东省食管癌研究所, 广东 广州 510060
Author(s):
XIAO Liangjie1 2 FANG Yongwen1 WANG Yuliu1 LIN Xiaosheng1 XU Senkui1 2 YAO Wenyan1 ZHAO Lei1 FANG Jianlan1
1. Department of Radiotherapy, Sun Yat-sen University Cancer Center, Guangzhou 510060, China 2. Guangdong Institute of Esophageal Cancer, Guangzhou 510060, China
关键词:
锥形束CT食管癌肺癌六维床摆位误差
Keywords:
Keywords: cone-beam CT esophageal cancer lung cancer six-dimensional couch setup error
分类号:
R815.6;R735.1
DOI:
DOI:10.3969/j.issn.1005-202X.2022.10.001
文献标志码:
A
摘要:
【摘要】目的:应用锥形束CT(CBCT)观察食管癌和肺癌六维方向的摆位误差,分析六维床对食管癌平移和旋转方向校正的必要性及其临床意义。方法:选取2019年10月~2021年12月在中山大学肿瘤防治中心放疗的胸部肿瘤患者,食管癌组、肺癌组各85例,收集所有患者前3次CBCT扫描,通过六维方向与计划中的定位CT图像进行配准,获得头脚方向(SI)、左右方向(LR)、前后方向(AP)、绕Z轴旋转(RTN)、绕X轴旋转(PITCH)、绕Y轴旋转(ROLL)6个方向的摆位误差。原始数据取绝对值,采用Kolmogorov-Smirnov Z非参数秩和检验进行统计分析。结果:食管癌组和肺癌组在SI、LR、AP平移方向摆位误差取绝对值分别为:0.30(0.10, 0.50)、0.10(0.10, 0.30)、0.10(0.10, 0.20)和0.20(0.10, 0.40)、0.20(0.10, 0.30)、0.10(0.10, 0.30) cm;在RTN、PITCH、ROLL旋转方向的摆位误差取绝对值分别为0.50°(0.30°, 0.90°)、0.90°(0.40°, 1.60°)、0.90°(0.40°, 1.40°)和0.40°(0.10°, 0.70°)、0.70°(0.20°, 1.30°)、0.50°(0.10°, 1.00°),食管癌组和肺癌组中SI、RTN、PITCH、ROLL方向比较,差异均有统计学意义(P<0.05),而LR、AP方向比较,差异无统计学意义(P>0.05)。结论:基于CBCT引导下,六维配准技术配合六维床能降低胸部肿瘤患者放疗时摆位误差,特别是长靶区如食管癌方面,校正旋转摆位误差有明显优势。
Abstract:
Abstract: Objective To analyze the 6-dimensional setup errors of esophageal cancer and lung cancer with the use of cone-beam CT (CBCT), and to discuss the necessity and clinical significance of 6-dimensional couch for translational and rotational errors correction in esophageal cancer. Methods Patients with thoracic tumors treated with radiotherapy at Sun Yat-sen University Cancer Center from October 2019 to December 2021 were enrolled, with 85 cases each in esophageal cancer group and lung cancer group. The first 3 CBCT scans of all patients were collected for obtaining the 6-dimensional setup errors, including superior-inferior (SI), left-right (LR), anterior-posterior (AP) directional setup errors, and rotational setup errors around Z-axis (RTN), X-axis (PITCH), Y-axis (ROLL). The absolute values of the raw data were analyzed by Kolmogorov-Smirnov Z nonparametric rank sum test. Results The absolute values of SI, LR, and AP setup errors were 0.30 (0.10, 0.50), 0.10 (0.10, 0.30), 0.10 (0.10, 0.20) cm in esophageal cancer group, and 0.20 (0.10, 0.40), 0.20 (0.10, 0.30), 0.10 (0.10, 0.30) cm in lung cancer group. The RTN, PITCH, and ROLL setup errors were 0.50° (0.30°, 0.90°), 0.90° (0.40°, 1.60°), 0.90° (0.40°, 1.40°) in esophageal cancer group, and 0.40° (0.10°, 0.70°), 0.70° (0.20°, 1.30°), 0.50° (0.10°, 1.00°) in lung cancer group. There were statistically significant differences between esophageal cancer group and lung cancer group in SI, RTN, PITCH and ROLL setup errors (P<0.05), but not in LR and AP setup errors (P>0.05). Conclusion Guided by CBCT, 6-dimensional registration technique with 6-dimensional couch can reduce the setup errors during radiotherapy for thoracic tumors, especially for long target areas such as esophageal cancer, and has obvious advantages in correcting the rotational setup errors.

相似文献/References:

[1]张俊俊,杨 振,柏朋刚,等.运动模体中靶区长度与锥形束CT扫描速度的相关性[J].中国医学物理学杂志,2015,32(02):259.[doi:10.3969/j.issn.1005-202X.2015.02.025]
[2]杨一威,商海焦,单国平,等.两种不同优化算法在胸部肿瘤调强放射治疗计划中的对比研究[J].中国医学物理学杂志,2015,32(02):276.[doi:10.3969/j.issn.1005-202X.2015.02.030]
[3]张一戈,邱小平,陈维军,等.基于图像引导技术对头颈部肿瘤临床摆位误差的剂量学影响[J].中国医学物理学杂志,2015,32(01):25.[doi:10.3969/j.issn.1005-202X.2015.01.007]
[4]曾 彪,张九堂,席许平,等.单中心上下半野技术在胸中上段食管癌放射治疗中的剂量学分布[J].中国医学物理学杂志,2015,32(03):357.[doi:10.3969/j.issn.1005-202X.2015.03.012]
[5]陈恩乐,吴 魁,董 事,等.改变多叶准直器角度对调强放疗计划效率的影响[J].中国医学物理学杂志,2015,32(03):437.[doi:10.3969/j.issn.1005-202X.2015.03.030]
[6]李 毅,李文荣,苏 进,等.基于CBCT非小细胞肺癌外放边界研究[J].中国医学物理学杂志,2014,31(04):5006.[doi:10.3969/j.issn.1005-202X.2014.04.008]
[7]翁邓胡,王 建,尹中明,等.基于锥形束CT研究肺癌图像引导放疗的内靶区外放边界值[J].中国医学物理学杂志,2014,31(04):5012.[doi:10.3969/j.issn.1005-202X.2014.04.009]
[8]宋 威,赵 迪,鹿 红,等.射野方向优化技术在胸中上段食管癌调强放疗中的应用[J].中国医学物理学杂志,2014,31(05):5109.[doi:10.3969/j.issn.1005-202X.2014.05.003]
[9]查元梓,吴晔明,陆冬青,等.应用锥形束CT研究儿童神经母细胞瘤放射治疗的摆位误差[J].中国医学物理学杂志,2014,31(05):5128.[doi:10.3969/j.issn.1005-202X.2014.05.007]
[10]沈奕晨,聂生东.锥形束CT与螺旋CT图像配准算法及其在放疗中的应用[J].中国医学物理学杂志,2014,31(05):5151.[doi:10.3969/j.issn.1005-202X.2014.05.013]
[11]巩汉顺,徐寿平,刘博,等.食管癌患者CBCT图像剂量计算准确性的分析评估[J].中国医学物理学杂志,2019,36(5):497.[doi:DOI:10.3969/j.issn.1005-202X.2019.05.001]
 GONG Hanshun,XU Shouping,LIU Bo,et al. Accuracy of dose calculation with CBCT images in patients with esophageal cancer[J].Chinese Journal of Medical Physics,2019,36(10):497.[doi:DOI:10.3969/j.issn.1005-202X.2019.05.001]

备注/Memo

备注/Memo:
【收稿日期】2022-05-19 【基金项目】国家自然科学基金(81874220);广东省医学科学技术研究基金(A2020621);广东省食管癌研究所科技计划(Q202107) 【作者简介】肖亮杰,技师,研究方向:放疗技术摆位误差分析,E-mail: xiaolj@sysucc.org.cn 【通信作者】方键蓝,技师,研究方向:放射治疗技术,E-mail: fangjl@sysucc.org.cn
更新日期/Last Update: 2022-10-27