[1]卢涛,白薇薇. 基于光声多普勒频谱标准偏差的自相关碳颗粒横向流速测量[J].中国医学物理学杂志,2018,35(5):598-601.[doi:DOI:10.3969/j.issn.1005-202X.2018.05.018]
 LU Tao,BAI Weiwei. Transverse flow velocity measurement of carbon particles based on photoacoustic Doppler standard deviation calculated with autocorrelation method[J].Chinese Journal of Medical Physics,2018,35(5):598-601.[doi:DOI:10.3969/j.issn.1005-202X.2018.05.018]
点击复制

 基于光声多普勒频谱标准偏差的自相关碳颗粒横向流速测量()
分享到:

《中国医学物理学杂志》[ISSN:1005-202X/CN:44-1351/R]

卷:
35卷
期数:
2018年第5期
页码:
598-601
栏目:
医学信号处理与医学仪器
出版日期:
2018-05-22

文章信息/Info

Title:
 Transverse flow velocity measurement of carbon particles based on photoacoustic Doppler standard deviation calculated with autocorrelation method
文章编号:
1005-202X(2018)05-0598-04
作者:
 卢涛白薇薇
 河南工业大学电气工程学院, 河南 郑州 450001
Author(s):
 LU Tao BAI Weiwei
 College of Electrical Engineering, Henan University of Technology, Zhengzhou 450001, China
关键词:
 光声多普勒横向流速测量频谱展宽标准偏差碳颗粒悬混液
Keywords:
 Keywords: photoacoustic Doppler transverse flow velocity measurement bandwidth broadening standard deviation carbon particle suspension
分类号:
Q631
DOI:
DOI:10.3969/j.issn.1005-202X.2018.05.018
文献标志码:
A
摘要:
 为了测量与超声探头信号接收轴线相垂直的碳颗粒悬混液的流速,采用多普勒频谱的标准偏差估计频带展宽,频谱标准偏差由自相关方法计算。光声信号由波长532 nm重复频率20 Hz的脉冲激光激励,由中心频率10 MHz的聚焦超声换能器采集。碳颗粒悬混液由微量注射泵驱动,时域光声信号经希尔伯特变换为复信号后进行自相关计算,多普勒频宽的标准偏差由若干序列A扫的自相关平均得到。对比之前基于序列A扫的互相关方法,所提出方法的优点是自相关中的信号时移大小可自定义,如时移量远小于信号扫描间隔,可避免信号混叠,避免互相关中对重复频率为数千赫兹脉冲激光的需求。该方法的可行性通过测量横向流速为5.0~8.4 mm/s的碳颗粒悬混液的频谱标准偏差得到初步验证,实验结果显示信号的自相关结果在测量范围内呈现线性分布趋势。
Abstract:
 Abstract: In order to measure the flow velocity of carbon particle suspension perpendicular to the receiving axis of ultrasound transducer, the standard deviation of photoacoutic Doppler frequency spectrum which was calculated with an autocorrelation method was used to estimate the bandwidth broadening. A 532 nm pulsed laser with the repetition frequency of 20 Hz was used as a pumping source to generate photoacoustic signals which were then detected using a focused ultrasound transducer with a central frequency of 10 MHz. The suspension of carbon particles was driven by a micro-injection syringe pump. The photoacoustic signals in time domain were transformed into complex photoacoustic signals using Hilbert transformation before autocorrelation calculation. The standard deviation of Doppler bandwidth was calculated by averaging the autocorrelation results of several sequential A scans. Compared with the previously reported cross-correlation method based on sequential A scans, the advantage of the proposed method is that the time delay in autocorrelation can be customized. If the time delay was chosen much less than the time interval between the sequential A scans, aliasing can be avoided, moreover, the requirement for high laser repetition frequency up to several kHz in cross-correlation was avoided. The feasibility of the proposed method is preliminarily demonstrated by measuring the standard deviation of the carbon particle suspension with a transversal flow velocity from 5.0 to 8.4 mm/s. The experimental results revealed that the autocorrelation results were approximately distributed linearly within the measuring range.

相似文献/References:

[1]王遥,霍万里,熊壮,等.TACE手术中不同站姿下铅眼镜和铅面罩对医生眼晶状体防护效果的蒙特卡洛模拟比较[J].中国医学物理学杂志,2016,33(6):553.[doi:DOI:10.3969/j.issn.1005-202X.2016.06.003]
 [J].Chinese Journal of Medical Physics,2016,33(5):553.[doi:DOI:10.3969/j.issn.1005-202X.2016.06.003]
[2]张新,谷晓芳,王培臣,等.轻离子束治疗设备注册检验关键技术问题[J].中国医学物理学杂志,2016,33(6):559.[doi:10.3969/j.issn.1005-202X.2016.06.004]
 [J].Chinese Journal of Medical Physics,2016,33(5):559.[doi:10.3969/j.issn.1005-202X.2016.06.004]
[3]江芬芬,王培,康盛伟,等. 热释光剂量片测量肺部肿瘤放疗剂量的方法[J].中国医学物理学杂志,2016,33(6):564.[doi:10.3969/j.issn.1005-202X.2016.06.005]
 [J].Chinese Journal of Medical Physics,2016,33(5):564.[doi:10.3969/j.issn.1005-202X.2016.06.005]
[4]刘洪源,彭威,杨锐,等. 锥形束CT离线校正肺癌摆位误差[J].中国医学物理学杂志,2016,33(6):573.[doi:10.3969/j.issn.1005-202X.2016.06.007]
 [J].Chinese Journal of Medical Physics,2016,33(5):573.[doi:10.3969/j.issn.1005-202X.2016.06.007]
[5]赵彪,潘香,杨毅,等. 右乳癌保乳术后瘤床同步X线和后程电子线补量调强放疗剂量学比较[J].中国医学物理学杂志,2016,33(6):576.[doi:10.3969/j.issn.1005-202X.2016.06.008]
 [J].Chinese Journal of Medical Physics,2016,33(5):576.[doi:10.3969/j.issn.1005-202X.2016.06.008]
[6]邓南,钱建庭,刁现芬,等. 基于宽带检测放疗X-光光声效应仿体实验[J].中国医学物理学杂志,2016,33(9):865.[doi:DOI:10.3969/j.issn.1005-202X.2016.09.001]
 [J].Chinese Journal of Medical Physics,2016,33(5):865.[doi:DOI:10.3969/j.issn.1005-202X.2016.09.001]
[7]张先稳,李军,张西志,等. 宫颈癌术后5野调强放疗4个变量组合的最佳治疗模式的剂量学[J].中国医学物理学杂志,2016,33(9):872.[doi:10.3969/j.issn.1005-202X.2016.09.002]
 [J].Chinese Journal of Medical Physics,2016,33(5):872.[doi:10.3969/j.issn.1005-202X.2016.09.002]
[8]胡健,李承军,徐利明,等. 床面倾斜对剂量验证通过率的影响[J].中国医学物理学杂志,2016,33(9):881.[doi:10.3969/j.issn.1005-202X.2016.09.003]
 [J].Chinese Journal of Medical Physics,2016,33(5):881.[doi:10.3969/j.issn.1005-202X.2016.09.003]
[9]陈亚正,肖明勇,孙春堂,等. 准直器角度对宫颈癌术后VMAT计划的影响[J].中国医学物理学杂志,2016,33(9):885.[doi:10.3969/j.issn.1005-202X.2016.09.004]
 [J].Chinese Journal of Medical Physics,2016,33(5):885.[doi:10.3969/j.issn.1005-202X.2016.09.004]
[10]李毅,唐丰文,张晓智. 基于四维CT和锥形束CT确定非小细胞肺癌放疗靶区外放边界[J].中国医学物理学杂志,2016,33(9):892.[doi:10.3969/j.issn.1005-202X.2016.09.005]
 [J].Chinese Journal of Medical Physics,2016,33(5):892.[doi:10.3969/j.issn.1005-202X.2016.09.005]

备注/Memo

备注/Memo:
 【收稿日期】2017-11-18
【基金项目】河南省自然科学基金(162300410054);河南省属高校基本科研业务费专项(2015RCJH13)
【作者简介】卢涛,博士,副教授,主要研究方向:光声成像与检测,E-mail: hautlutao@hotmail.com