[1]祁勇嘉,葛云,陈清清,等.基于光学摆位系统的放疗定摆位体姿体态质控[J].中国医学物理学杂志,2026,43(7):861-865.[doi:DOI:10.3969/j.issn.1005-202X.2026.07.005]
 QI Yongjia,GE Yun,CHEN Qingqing,et al.Optical positioning system-based quality control for posture and body contour in radiotherapy positioning and setup[J].Chinese Journal of Medical Physics,2026,43(7):861-865.[doi:DOI:10.3969/j.issn.1005-202X.2026.07.005]
点击复制

基于光学摆位系统的放疗定摆位体姿体态质控()

《中国医学物理学杂志》[ISSN:1005-202X/CN:44-1351/R]

卷:
43卷
期数:
2026年第7期
页码:
861-865
栏目:
医学放射物理
出版日期:
2026-07-22

文章信息/Info

Title:
Optical positioning system-based quality control for posture and body contour in radiotherapy positioning and setup
文章编号:
1005-202X(2026)07-0861-05
作者:
祁勇嘉1葛云1陈清清2褚耀华2
1.南京大学电子科学与工程学院, 江苏 南京 210046; 2.苏州市立医院肿瘤放射治疗科, 江苏 苏州 215000
Author(s):
QI Yongjia1 GE Yun1 CHEN Qingqing2 CHU Yaohua2
1. School of Electronic Science and Engineering, Nanjing University, Nanjing 210046, China 2. Department of Radiation Oncology, Suzhou Municipal Hospital, Suzhou 215000, China
关键词:
放射治疗定摆位光学摆位系统体姿体态质控
Keywords:
Keywords: radiotherapy positioning and setup optical positioning system posture body contour quality control
分类号:
R318;R811.1
DOI:
DOI:10.3969/j.issn.1005-202X.2026.07.005
文献标志码:
A
摘要:
体姿体态的变化是放疗定摆位过程常见的误差来源,但是当前放疗定摆位设备种类繁多,原理各不相同,对于该过程精准性的判断较为困难。因此,采用联合定摆位的方式进行比较,并利用相应的质控平台对其准确性进行判断,对可能的误差原因进行溯源分析,对临床治疗具有重要意义。利用光学摆位系统可以实时获取摆位偏差的特点,本研究设计了一种放疗定摆位体姿体态质控方法,通过收集临床定摆位过程中的光学摆位系统数据,结合CT和CBCT图像,可以快速判断该过程的精准性以及体姿体态是否发生变化、辨别分析出误差产生的原因。在不同部位的病例情况下,本文方法都展现出良好的分析能力,能准确地发现问题,提升放疗质量。
Abstract:
Abstract: Variations in posture and body contour constitute common error sources during radiotherapy positioning and setup. However, there are various types of radiotherapy positioning and setup devices operating on different principles, making it difficult to accurately assess the positioning and setup precision. Therefore, it is of great significance for clinical treatment to compare combined positioning and setup methods, validate their accuracy using the corresponding quality control platforms, and trace the root causes of possible errors. By leveraging the capability of an optical positioning system to acquire positioning deviations in real time, a quality control method for posture and body contour in radiotherapy positioning and setup is proposed. Through the analysis of optical positioning system data collected during the clinical positioning and setup, and combined with CT and CBCT images, the proposed method can quickly assess positioning and setup precision, detect postural and contour variations, and identify and analyze the underlying causes of errors. The method exhibits strong analytical performance across various clinical cases and can accurately detect positioning and setup abnormalities, thereby improving radiotherapy quality.

相似文献/References:

[1]武 杰,袁杭英,欧阳蓓蕾.内旋式放射治疗用面罩固定装置的设计[J].中国医学物理学杂志,2015,32(02):265.[doi:10.3969/j.issn.1005-202X.2015.02.027]
[2]司 森,吴鑫全,包尚联,等.基于GPU的蒙特卡洛放疗剂量模拟[J].中国医学物理学杂志,2015,32(01):1.[doi:10.3969/j.issn.1005-202X.2015.01.001]
[3]王 文,程梦云,杨 琪,等.基于MCNP源子程序的放射治疗剂量计算验证方法[J].中国医学物理学杂志,2015,32(01):13.[doi:10.3969/j.issn.1005-202X.2015.01.004]
[4]张一戈,邱小平,陈维军,等.基于图像引导技术对头颈部肿瘤临床摆位误差的剂量学影响[J].中国医学物理学杂志,2015,32(01):25.[doi:10.3969/j.issn.1005-202X.2015.01.007]
[5]孟慧鹏,孙小喆,王 昊,等.胸部偏后背肿瘤RapidArc采用仰卧位设计治疗计划的可行性分析[J].中国医学物理学杂志,2015,32(01):31.[doi:10.3969/j.issn.1005-202X.2015.01.008]
[6]庞皓文,孙小扬.基于快速自由形变的弹性配准算法在放疗CT中的应用[J].中国医学物理学杂志,2015,32(03):404.[doi:10.3969/j.issn.1005-202X.2015.03.023]
[7]贾卫娟,周一兵,谭维维,等.利用模拟机及挡铅托架进行放疗前的治疗模拟[J].中国医学物理学杂志,2014,31(01):4622.[doi:10.3969/j.issn.1005-202X.2014.01.007]
[8]白 雪,杜向慧,王升晔,等.半野技术结合非共面照射技术在乳腺癌胸锁联合照射中的应用[J].中国医学物理学杂志,2014,31(03):4849.[doi:10.3969/j.issn.1005-202X.2014.03.001]
[9]周 钢,田 野,陆雪官,等.乳腺癌保乳术后三种瘤床同期加量放疗技术的剂量学比较[J].中国医学物理学杂志,2014,31(03):4853.[doi:10.3969/j.issn.1005-202X.2014.03.002]
[10]李贤富,谢 力,郭飞,等.医科达Precise直线加速器快速应急处理方法探讨[J].中国医学物理学杂志,2014,31(04):4984.[doi:10.3969/j.issn.1005-202X.2014.04.003]

备注/Memo

备注/Memo:
【收稿日期】2026-02-21 【基金项目】苏州市“科教兴卫”青年科技项目(KJXW2022029);苏州市临床重点病种诊疗技术专项(LCZX202209) 【作者简介】祁勇嘉,硕士,研究方向:医学物理,E-mail: qindd1009@163.com 【通信作者】褚耀华,主任技师,研究方向:放射治疗,E-mail: Chuyaohua522@163.com
更新日期/Last Update: 2026-07-22