[1]石婷婷,倪昕晔,韩济华,等.基于Monaco计划系统的Inc参数在乳腺癌放疗计划中的剂量学研究[J].中国医学物理学杂志,2026,43(3):281-287.[doi:DOI:10.3969/j.issn.1005-202X.2026.03.001]
 SHI Tingting,NI Xinye,HAN Jihua,et al.Dosimetric study of Inc parameters based on the Monaco treatment planning system for breast cancer radiotherapy[J].Chinese Journal of Medical Physics,2026,43(3):281-287.[doi:DOI:10.3969/j.issn.1005-202X.2026.03.001]
点击复制

基于Monaco计划系统的Inc参数在乳腺癌放疗计划中的剂量学研究()

《中国医学物理学杂志》[ISSN:1005-202X/CN:44-1351/R]

卷:
43卷
期数:
2026年第3期
页码:
281-287
栏目:
医学放射物理
出版日期:
2026-03-27

文章信息/Info

Title:
Dosimetric study of Inc parameters based on the Monaco treatment planning system for breast cancer radiotherapy
文章编号:
1005-202X(2026)03-0281-07
作者:
石婷婷1倪昕晔2韩济华1张艳1张晓晔1朱志坚1洪君1
1.南京医科大学附属淮安第一医院放疗科, 江苏 淮安 223001; 2.南京医科大学第三附属医院(常州市第二人民医院), 江苏 常州 213003
Author(s):
SHI Tingting1 NI Xinye2 HAN Jihua1 ZHANG Yan1 ZHANG Xiaoye1 ZHU Zhijian1 HONG Jun1
1. Department of Radiotherapy, the Affiliated Huaian No.1 Peoples Hospital of Nanjing Medical University, Huaian 223001, China 2. The Third Affiliated Hospital of Nanjing Medical University (Changzhou Second Peoples Hospital), Changzhou 213003, China
关键词:
乳腺癌剂量验证机架角度容积旋转调强放射治疗γ通过率
Keywords:
Keywords: breast cancer dose verification gantry angle volumetric modulated arc therapy gamma pass rate
分类号:
R318;R811.1
DOI:
DOI:10.3969/j.issn.1005-202X.2026.03.001
文献标志码:
A
摘要:
目的:分析不同机架角增量参数对保乳术后乳腺癌放疗计划的影响。方法:选取左乳乳腺癌病例26例,在Monaco5.11计划系统上为每位患者分别设计3组不同机角增量的容积旋转调强放射治疗(VMAT)计划,分别为小Inc组(10°)、中Inc组(25°)、大Inc组(40°)。比较各组计划靶区、危及器官及计划验证通过率。结果:乳腺癌保乳术后,中、小Inc值(10°、25°)的VMAT计划靶区PTV覆盖率、CI、HI、计划验证通过率更优,危及器官受量也更低,患侧肺V5、V20、Dmean,心脏V10、Dmean等受量在Inc值为10°、25°时更小。结论:在设计保乳术后乳腺癌VMAT计划时,将机架角增量设置为10°~25°,能使危及器官得到最好的保护,还能保证较高的绝对γ通过率。
Abstract:
Abstract: Objective To analyze the effects of different angular increments (Inc) on volumetric modulated arc therapy (VMAT) plans for breast cancer after breast-conserving surgery. Methods A total of 26 patients with left-sided breast cancer were enrolled. For each patients, 3 VMAT plans with varying increments of gantry angle were generated using the MonacoTM (v5.11) treatment planning system: a small Inc group (10°), a medium Inc group (25°), and a large Inc group (40°). The 3 groups were compared regarding target coverage, organ-at-risk (OAR) doses, and plan verification pass rates. Results After breast-conserving surgery for breast cancer, VMAT plans with small-to-medium Inc values (10° and 25°) exhibited superior PTV coverage, conformity index, homogeneity index, and plan verification pass rate. Furthermore, plans with Inc values of 10° and 25°generally resulted in lower doses to OAR, with smaller V5, V20, and Dmean of the ipsilateral lung, as well as V10 and Dmean of the heart. Conclusion When designing VMAT plans for breast cancer after breast-conserving surgery, setting the increment of gantry angle between 10° and 25° can achieve the optimal OAR sparing while ensuring a high absolute gamma pass rate.

相似文献/References:

[1]柏 晗,王 丽,刘旭红,等.左乳腺5种不同主野分布IMRT计划的剂量比较分析[J].中国医学物理学杂志,2015,32(02):210.[doi:10.3969/j.issn.1005-202X.2015.02.013]
[2]张富利,王雅棣,许卫东,等.应用两种三维探测器阵列进行螺旋断层调强放疗计划剂量验证[J].中国医学物理学杂志,2015,32(02):218.[doi:10.3969/j.issn.1005-202X.2015.02.015]
[3]张晋建,黎 静,文 婷,等.Arc CHECK在乳腺癌螺旋断层放疗计划剂量验证中的应用[J].中国医学物理学杂志,2015,32(02):244.[doi:10.3969/j.issn.1005-202X.2015.02.021]
[4]曾 彪,张九堂,席许平.单中心上下半野调强放射治疗的物理剂量验证[J].中国医学物理学杂志,2015,32(01):17.[doi:10.3969/j.issn.1005-202X.2015.01.005]
[5]任 强,王 玉,曹瑞芬,等.基于非晶硅电子射野影像装置的精确剂量刻度方法[J].中国医学物理学杂志,2015,32(01):38.[doi:10.3969/j.issn.1005-202X.2015.01.010]
[6]白 雪,杜向慧,王升晔,等.半野技术结合非共面照射技术在乳腺癌胸锁联合照射中的应用[J].中国医学物理学杂志,2014,31(03):4849.[doi:10.3969/j.issn.1005-202X.2014.03.001]
[7]赵永亮,储开岳,吴建亭,等.两种不同调强技术配合主动呼吸控制系统在乳腺癌放疗中的比较[J].中国医学物理学杂志,2014,31(03):4862.[doi:10.3969/j.issn.1005-202X.2014.03.004]
[8]刘 浩,王 新,李公平,等.用二维电离室矩阵验证宫颈癌腔内放疗剂量分布[J].中国医学物理学杂志,2014,31(04):4988.[doi:10.3969/j.issn.1005-202X.2014.04.004]
[9]张富利,蒋华勇,王雅棣,等.应用一维、二维、三维探测器阵列进行螺旋断层调强放疗计划剂量验证[J].中国医学物理学杂志,2014,31(06):5230.[doi:10.3969/j.issn.1005-202X.2014.06.002]
[10]易金玲,金献测,周永强,等.鼻咽癌IMRT与VMAT治疗的计划与剂量验证比较研究[J].中国医学物理学杂志,2013,30(01):3859.[doi:10.3969/j.issn.1005-202X.2013.01.006]
[11]唐慧敏,杨振,龚学余,等.不同计划系统的剂量学比较[J].中国医学物理学杂志,2016,33(3):286.[doi:DOI:10.3969/j.issn.1005-202X.2016.03.013]
 [J].Chinese Journal of Medical Physics,2016,33(3):286.[doi:DOI:10.3969/j.issn.1005-202X.2016.03.013]

备注/Memo

备注/Memo:
【收稿日期】2025-10-18 【基金项目】国家自然科学基金(62371243);江苏省重点研发计划社会发展项目(BE2022720);江苏省自然科学基金面上项目(BK20231190);江苏省医学重点学科建设单位(JSDW202237);常州市社会发展项目(CE20235063);常州市应用基础研究专项(CJ20244020) 【作者简介】石婷婷,工程师,研究方向:生物医学工程(医学物理), E-mail: hayystt@163.com 【通信作者】倪昕晔,博士,教授,研究方向:生物医学工程(医学物理),E-mail: nxy@njmu.edu.cn
更新日期/Last Update: 2026-03-27