[1]罗斌,李贤富,曹璐,等.统计过程控制方法在后装192Ir放射源空气比释动能强度验证中的应用[J].中国医学物理学杂志,2023,40(1):19-23.[doi:DOI:10.3969/j.issn.1005-202X.2023.01.004]
 LUO Bin,LI Xianfu,et al.Application of statistical process control method in verification of air-kerma strength of 192Ir brachytherapy source[J].Chinese Journal of Medical Physics,2023,40(1):19-23.[doi:DOI:10.3969/j.issn.1005-202X.2023.01.004]
点击复制

统计过程控制方法在后装192Ir放射源空气比释动能强度验证中的应用()
分享到:

《中国医学物理学杂志》[ISSN:1005-202X/CN:44-1351/R]

卷:
40卷
期数:
2023年第1期
页码:
19-23
栏目:
医学放射物理
出版日期:
2023-01-07

文章信息/Info

Title:
Application of statistical process control method in verification of air-kerma strength of 192Ir brachytherapy source
文章编号:
1005-202X(2023)01-0019-05
作者:
罗斌12李贤富12曹璐2阳华东2许鹏2郭飞2谢力2
1.医学影像四川省重点实验室/四川省肿瘤学重点学科, 四川 南充 637000; 2.川北医学院附属医院肿瘤科, 四川 南充 637000
Author(s):
LUO Bin1 2 LI Xianfu1 2 CAO Lu2 YANG Huadong2 XU Peng2 GUO Fei2 XIE Li2
1. Sichuan Key Laboratory of Medical Imaging/Key Disciplines of Oncology in Sichuan Province, Nanchong 637000, China 2. Department of Oncology, Affiliated Hospital of North Sichuan Medical College, Nanchong 637000, China
关键词:
统计过程控制方法后装192Ir空气比释动能强度过程能力指数
Keywords:
Keywords: statistical process control method brachytherapy 192Ir air-kerma strength process capability index
分类号:
R811.1;TH71
DOI:
DOI:10.3969/j.issn.1005-202X.2023.01.004
文献标志码:
A
摘要:
目的:运用统计过程控制(SPC)方法分析研究后装192Ir放射源空气比释动能强度(Sk)验证的精准性及稳定性。方法:首先使用PTW SOURCECHECK 4π井型电离室按照国标WS 262-2017的要求对192Ir放射源Sk进行质控验证。再通过Minitab软件对测量值和计划系统计算值之间的偏差绘制均值-极差图,计算相应的上下控制线UCL、LCL及其过程能力指数Cp、Cpk,并判断该方法的精准稳定性。结果:25组偏差值的分布为-1.584%~0.915%,达到了WS 262-2017中小于±5%、AAPM-TG43中小于±3%的限值要求。各分组的均值都处于-1.518%~0.973%的上下控制线范围内,Cp、Cpk取值分别为1.66、1.51,整个测量过程处于较稳定的控制水平内。结论:SPC方法能够对192Ir放射源Sk的质控验证结果即偏差值进行较好的监测与评估,发现其中的异常点,确保了后装治疗中剂量传递的准确性。
Abstract:
Abstract: Objective To analyze the accuracy and stability of statistical process control (SPC) method in the verification of the air-kerma strength (Sk) of 192Ir brachytherapy source. Methods After completing the quality control verification of 192Ir radioactive source Sk with PTW SOURCECHECK 4π well-type ionization chamber according to the requirements of national standard WS 262-2017, Minitab software was used to draw the mean value-range chart for the deviation between the measured value and the calculated value of the planning system. The corresponding upper and lower control lines and process capability indexes (Cp, Cpk) were calculated for evaluating and the accuracy and stability of SPC method. Results The distribution of the deviation values in the 25 groups ranged from -1.584% to 0.915%, reaching the limit requirements of less than ±5% in WS 262-2017 and less than ±3% in AAPM-TG43. The mean values of each group were within the range of upper and lower control lines (from -1.518% to 0.973%). Cp and Cpk were 1.66 and 1.51, respectively. The whole measurement process was at a relatively stable control level. Conclusion SPC method can better monitor and evaluate the quality control verification results (deviation values) of 192Ir brachytherapy source Sk, and find out the abnormal points, thereby ensuring the accuracy of dose delivery in brachytherapy.

相似文献/References:

[1]刘礼东,李军烽,李书舟,等.统计过程控制方法在加速器日常输出剂量稳定性中的研究[J].中国医学物理学杂志,2021,38(2):138.[doi:DOI:10.3969/j.issn.1005-202X.2021.02.002]
 LIU Lidong,LI Junfeng,LI Shuzhou,et al.Statistical process control method in evaluating the stability of daily output dose of linear accelerator[J].Chinese Journal of Medical Physics,2021,38(1):138.[doi:DOI:10.3969/j.issn.1005-202X.2021.02.002]

备注/Memo

备注/Memo:
【收稿日期】2022-06-26 【基金项目】四川卫生计生委项目(19PJ037);川北医学院附属医院科研项目(2022ZD004) 【作者简介】罗斌,硕士,物理师,主要从事放射物理工作,E-mail: 2283712905@qq.com 【通信作者】李贤富,教授,物理师、加速器工程师,研究方向:放射肿瘤学、肿瘤放射治疗物理与肿瘤治疗工程技术,E-mail: lixianfu13@163.com
更新日期/Last Update: 2023-01-07