Feasibility of RT-Mind auto-segmentation technique in radiotherapy for nasopharyngeal carcinoma
文章编号:
1005-202X(2022)06-0661-05
作者:
薛涛1吴迪1卢晓岩2张恒2秦浩人3李海鹏3孙婉君2王辉1
1.天津市人民医院放疗科, 天津 300121; 2.天津市人民医院肿瘤科, 天津 300121; 3.天津中医药大学中西医结合学院, 天津300193
Author(s):
XUE Tao1 WU Di1 LU Xiaoyan2 ZHANG Heng2 QIN Haoren3 LI Haipeng3 SUN Wanjun2 WANG Hui1
1. Department of Radiotherapy, Tianjin Peoples Hospital, Tianjin 300121, China 2. Department of Oncology, Tianjin Peoples Hospital, Tianjin 300121, China 3. School of Integrative Medicine, Tianjin University of Traditional Chinese Medicine, Tianjin 300193, China
关键词:
鼻咽癌RT-Mind自动勾画临床靶区危及器官
Keywords:
nasopharyngeal carcinoma RT-Mind auto-segmentation clinical target volume organ-at-risk
分类号:
R739.6;R811.1
DOI:
DOI:10.3969/j.issn.1005-202X.2022.06.001
文献标志码:
A
摘要:
目的:探讨RT-Mind软件在鼻咽癌放疗临床靶区(CTV)、危及器官(OARs)自动勾画的可行性,为临床应用提供依据。方法:回顾性选取28例应用调强技术治疗的鼻咽癌患者,将放疗医师手动勾画CTV、OARs(脑干、脊髓、左右晶体、视交叉、左右视神经、左右腮腺、左右颞叶、左右颞颌关节、下颌骨)做为参考标准,再使用RT-Mind软件自动勾画CTV及OARs。对比手动与自动勾画在CTV和OARs区域的Dice相似性系数(DSC)、Jaccard系数(JAC)、敏感性指数(SI)、包容性系数(lncI)、质心偏差(DC)、Hausdorff距离(HD)等参数,从而评估自动勾画效果。结果:CTV的DSC、JAC、SI、lncl、DC、HD分别为:0.78±0.04、0.70±0.05、0.85±0.08、0.87±0.04、(7.76±5.03) mm、(12.3±1.16) mm,OARs中DC、HD值均在1 cm之内。结论:RT-Mind软件能够基本满足临床要求,能够较为准确地实现鼻咽癌患者CTV和OARs的自动勾画。由于病人既往病史的个体差异,放射治疗医师必须根据临床需要,对自动勾画的CTV和OARs进行修改后,才能用于治疗。若依据临床需求进一步完善自动勾画的个性化定制,相信RT-Mind软件在解剖结构复杂的鼻咽癌放疗中能够辅助放疗医生提高工作效率,更好地为患者服务。
Abstract:
Abstract: Objective To explore the feasibility of RT-Mind software in the automatic segmentations of clinical target volume (CTV) and organs-at-risk (OAR) for radiotherapy for nasopharyngeal carcinoma, so as to provide a basis for its clinical application. Methods Twenty-eight cases of nasopharyngeal carcinoma patients treated with radiotherapy were enrolled retrospectively. Taking manual segementations of CTV and OAR (brain stem, spinal cord, lens, optic chiasm, optic nerves, parotid glands, temporal lobes, temporomandibular joints, and mandible) by radiotherapists as reference standards, RT-Mind software was used to carry out the auto-segmentations of CTV and OAR. The Dice similarity coefficient (DSC), Jaccard index (JAC), sensitivity index (SI), inclusiveness index (Incl), deviation of centroid (DC), and Hausdorff distance (HD) between manual and automatic segmentations of CTV and OAR were calculated for evaluating the performance of automatic segmentation. Results The DSC, JAC, SI, Incl, DC, and HD of CTV were 0.78±0.04, 0.70±0.05, 0.85±0.08, 0.87±0.04, (7.76±5.03) mm, and (12.3±1.16) mm, respectively. The DC and HD of OAR were all less than 1 cm. Conclusion RT-Mind software can basically meet the clinical requirements, and can more accurately realize the automatic segementations of OAR and CTV in patients with nasopharyngeal carcinoma. Due to individual differences in the patients medical history, radiotherapy physicians need to make certain modifications to the auto-segmentation results of OAR and CTV according to clinical needs. If the personalized customization of auto-segmentation can be further improved according to clinical requirements, RT-Mind software can assist doctors in improving their work efficiency in the radiotherapy of nasopharyngeal carcinoma with complex anatomical structures, and serving patients better.

相似文献/References:

[1]陈广涛,汪 淇,刘亚强,等.TrueBeam加速器射野准直器跟随功能在鼻咽癌固定野调强放疗和容积旋转调强放疗中的应用[J].中国医学物理学杂志,2014,31(03):4865.[doi:10.3969/j.issn.1005-202X.2014.03.005]
[2]潘宇辉,陈榕钦,柏朋刚,等.鼻咽癌调强放射治疗前后腮腺体积及空间几何位置变化研究[J].中国医学物理学杂志,2014,31(05):5120.[doi:10.3969/j.issn.1005-202X.2014.05.005]
[3]张俊俊,邱小平,李奇欣,等.ArcCheck系统在鼻咽癌容积旋转调强剂量验证中的应用[J].中国医学物理学杂志,2014,31(05):5136.[doi:10.3969/j.issn.1005-202X.2014.05.009]
[4]桑 勇,陈维军,狄小云,等.多叶光栅叶片宽度对鼻咽癌剂量学的影响[J].中国医学物理学杂志,2014,31(06):5240.[doi:10.3969/j.issn.1005-202X.2014.06.004]
[5]黄利生,庄婷婷,伍方财,等.鼻咽癌调强放疗Ib区剂量分析[J].中国医学物理学杂志,2014,31(06):5265.[doi:10.3969/j.issn.1005-202X.2014.06.010]
[6]易金玲,金献测,周永强,等.鼻咽癌IMRT与VMAT治疗的计划与剂量验证比较研究[J].中国医学物理学杂志,2013,30(01):3859.[doi:10.3969/j.issn.1005-202X.2013.01.006]
[7]金献测,易金玲,周永强,等.鼻咽癌同步加量调强放射治疗新的计划质量评估指标研究[J].中国医学物理学杂志,2013,30(02):3978.[doi:10.3969/j.issn.1005-202X.2013.02.004]
[8]蒋晓芹,段宝风,艾平,等.基于图谱库的自动轮廓勾画软件(ABAS)在鼻咽癌调强放疗中的应用[J].中国医学物理学杂志,2013,30(02):3997.[doi:10.3969/j.issn.1005-202X.2013.02.008]
[9]张达光,蒋胜鹏,李智华,等.鼻咽癌调强放疗中解剖结构改变及剂量分布变化的研究[J].中国医学物理学杂志,2013,30(04):4239.[doi:10.3969/j.issn.1005-202X.2013.04.006]
[10]杨海燕,王军良,周振山,等.AAA算法和PBC算法在鼻咽癌调强放疗中剂量分布的比较[J].中国医学物理学杂志,2013,30(04):4245.[doi:10.3969/j.issn.1005-202X.2013.04.008]

备注/Memo

备注/Memo:
【收稿日期】2021-10-26 【基金项目】国家自然科学基金(81972847, 81573089) 【作者简介】薛涛,工程师,研究方向:肿瘤放射物理,E-mail: tjxt1168@sina.com 【通信作者】王辉,博士,主任医师,研究方向:肿瘤放射治疗,E-mail: ezxwanghui@163.com
更新日期/Last Update: 2022-06-27