[1]张光伟,钟鹤立,吴何苟,等.光学表面监测系统引导实现左侧乳腺癌深吸气屏气治疗临床应用及评估[J].中国医学物理学杂志,2021,38(7):793-797.[doi:DOI:10.3969/j.issn.1005-202X.2021.07.001]
 ZHANG Guangwei,ZHONG Heli,WU Hegou,et al.Application and evaluation of OSMS guided DIBH radiotherapy for left-sided breast cancer[J].Chinese Journal of Medical Physics,2021,38(7):793-797.[doi:DOI:10.3969/j.issn.1005-202X.2021.07.001]
点击复制

光学表面监测系统引导实现左侧乳腺癌深吸气屏气治疗临床应用及评估()
分享到:

《中国医学物理学杂志》[ISSN:1005-202X/CN:44-1351/R]

卷:
38卷
期数:
2021年第7期
页码:
793-797
栏目:
医学放射物理
出版日期:
2021-07-26

文章信息/Info

Title:
Application and evaluation of OSMS guided DIBH radiotherapy for left-sided breast cancer
文章编号:
1005-202X(2021)07-0793-05
作者:
张光伟钟鹤立吴何苟刘剑锋张定杨东樊娟李彬徐广然
深圳市人民医院(暨南大学第二临床医学院,南方科技大学第一附属医院)肿瘤放疗科, 广东 深圳 518020
Author(s):
ZHANG Guangwei ZHONG Heli WU Hegou LIU Jianfeng ZHANG Ding YANG Dong FAN Juan LI Bin XU Guangran
Department of Radiation Oncology, Shenzhen Peoples Hospital (the Second Clinical Medical College of Jinan University, the First Affiliated Hospital of Southern University of Science and Technology), Shenzhen 518020, China
关键词:
光学表面监测系统放射治疗左侧乳腺癌深吸气屏气自动摆位
Keywords:
Keywords: optical surface monitoring system radiotherapy left-sided breast cancer deep inspiration breath hold automatic positioning
分类号:
R811.1
DOI:
DOI:10.3969/j.issn.1005-202X.2021.07.001
文献标志码:
A
摘要:
目的:通过分析摆位精度、摆位时间和心脏位置评估以光学表面监测系统(OSMS)引导实现左侧乳腺癌深吸气屏气(DIBH)治疗的临床价值。方法:30例左侧乳腺癌保乳术后患者,运用OSMS的“Move Couch”自动摆位功能,以OSMS引导摆位、DIBH和治疗监控,实现DIBH技术治疗。分析CBCT配准靶区六维绝对误差数据和心脏位置的误差数据,记录摆位时间。结果:绝对摆位误差在x、y、z平移方向和Rx、Ry、Rz旋转方向分别为(0.11±0.08)、(0.14±0.11)、(0.13±0.10) cm和(0.79±0.66)、(0.69±0.57)、(0.73±0.64)°。心脏位置在x、y、z方向平移误差分别为(0.01±0.30)、(0.05±0.34)、(0.01±0.09) cm。摆位时间为(4.1±1.2) min。结论:左侧乳腺癌DIBH放疗技术可将心脏向下、后方推移,远离放疗靶区,减少心脏受量。基于OSMS实施左侧乳腺癌DIBH治疗实现了较好的靶区和心脏位置准确性,OSMS的自动摆位功能可帮助提高摆位精度,减少摆位时间。
Abstract:
Abstract: Objective To evaluate the clinical value of optical surface monitoring system (OSMS) guided deep inspiration breath hold (DIBH) radiotherapy for left-sided breast cancer by analyzing positioning accuracy, setup time and cardiac position. Methods Thirty patients with left-sided breast cancer after breast-conserving surgery were enrolled in the study. The Move Couch function of OSMS was used for automatic positioning, and DIBH radiotherapy was realized based on OSMS guided positioning, DIBH and treatment monitoring. After CBCT registration, the six-dimensional absolute errors of target areas and the errors of cardiac position were analyzed, and meanwhile the setup time which was defined as the time from the patients contact with the couch to the time before CBCT scanning or before accelerator beam on (without CBCT scanning) was recorded. Results The absolute setup errors were (0.11±0.08), (0.14±0.11), (0.13±0.10) cm in x, y, z directions, respectively, and (0.79±0.66)°, (0.69±0.57)°, (0.73±0.64)° in Rx, Ry, Rz directions, respectively. The averages of cardiac position deviations were (0.01±0.30), (0.05±0.34), (0.01±0.09) cm in x, y, z directions. The recorded setup time was (4.1±1.2) min. Conclusion DIBH technique for left-sided breast cancer makes the heart distant from treatment target area by moving the heart to posterior and inferior directions, thereby lowering the radiation dose delivered to the heart. OSMS guided DIBH radiotherapy for left-sided breast cancer results in a good positioning accuracy of target areas and the heart. The automatic positioning function of OSMS contributes to improving positioning accuracy and reducing setup time.

相似文献/References:

[1]武 杰,袁杭英,欧阳蓓蕾.内旋式放射治疗用面罩固定装置的设计[J].中国医学物理学杂志,2015,32(02):265.[doi:10.3969/j.issn.1005-202X.2015.02.027]
[2]司 森,吴鑫全,包尚联,等.基于GPU的蒙特卡洛放疗剂量模拟[J].中国医学物理学杂志,2015,32(01):1.[doi:10.3969/j.issn.1005-202X.2015.01.001]
[3]王 文,程梦云,杨 琪,等.基于MCNP源子程序的放射治疗剂量计算验证方法[J].中国医学物理学杂志,2015,32(01):13.[doi:10.3969/j.issn.1005-202X.2015.01.004]
[4]张一戈,邱小平,陈维军,等.基于图像引导技术对头颈部肿瘤临床摆位误差的剂量学影响[J].中国医学物理学杂志,2015,32(01):25.[doi:10.3969/j.issn.1005-202X.2015.01.007]
[5]孟慧鹏,孙小喆,王 昊,等.胸部偏后背肿瘤RapidArc采用仰卧位设计治疗计划的可行性分析[J].中国医学物理学杂志,2015,32(01):31.[doi:10.3969/j.issn.1005-202X.2015.01.008]
[6]庞皓文,孙小扬.基于快速自由形变的弹性配准算法在放疗CT中的应用[J].中国医学物理学杂志,2015,32(03):404.[doi:10.3969/j.issn.1005-202X.2015.03.023]
[7]贾卫娟,周一兵,谭维维,等.利用模拟机及挡铅托架进行放疗前的治疗模拟[J].中国医学物理学杂志,2014,31(01):4622.[doi:10.3969/j.issn.1005-202X.2014.01.007]
[8]白 雪,杜向慧,王升晔,等.半野技术结合非共面照射技术在乳腺癌胸锁联合照射中的应用[J].中国医学物理学杂志,2014,31(03):4849.[doi:10.3969/j.issn.1005-202X.2014.03.001]
[9]周 钢,田 野,陆雪官,等.乳腺癌保乳术后三种瘤床同期加量放疗技术的剂量学比较[J].中国医学物理学杂志,2014,31(03):4853.[doi:10.3969/j.issn.1005-202X.2014.03.002]
[10]李贤富,谢 力,郭飞,等.医科达Precise直线加速器快速应急处理方法探讨[J].中国医学物理学杂志,2014,31(04):4984.[doi:10.3969/j.issn.1005-202X.2014.04.003]
[11]张光伟,吴何苟,刘剑锋,等.光学表面监测系统自动摆位功能对乳腺癌放疗患者治疗精度及摆位时间的影响[J].中国医学物理学杂志,2021,38(4):411.[doi:DOI:10.3969/j.issn.1005-202X.2021.04.003]
 ZHANG Guangwei,WU Hegou,LIU Jianfeng,et al.Effects of automatic setup function of optical surface monitoring system on treatment accuracy and setup time of breast cancer patients undergoing radiotherapy[J].Chinese Journal of Medical Physics,2021,38(7):411.[doi:DOI:10.3969/j.issn.1005-202X.2021.04.003]

备注/Memo

备注/Memo:
【收稿日期】2021-03-15 【基金项目】国家自然科学基金(61871374);深圳市科技研发基金(JCYJ20170307145422811) 【作者简介】张光伟,主管技师,研究方向:肿瘤放射治疗技术,E-mail: 59283444@qq.com 【通信作者】钟鹤立,高级工程师,研究方向:肿瘤放射物理,E-mail: 747775309@qq.com
更新日期/Last Update: 2021-07-26