[1]王语汇,王东燕,周俊东,等.基于红外定位系统放疗实时摆位预警系统的摆位数据误差统计[J].中国医学物理学杂志,2021,38(12):1477-1481.[doi:DOI:10.3969/j.issn.1005-202X.2021.12.005]
 WANG Yuhui,WANG Dongyan,ZHOU Jundong,et al.Statistical analysis of positioning errors based on OPS radiotherapy real-time positioning early warning system[J].Chinese Journal of Medical Physics,2021,38(12):1477-1481.[doi:DOI:10.3969/j.issn.1005-202X.2021.12.005]
点击复制

基于红外定位系统放疗实时摆位预警系统的摆位数据误差统计()
分享到:

《中国医学物理学杂志》[ISSN:1005-202X/CN:44-1351/R]

卷:
38卷
期数:
2021年第12期
页码:
1477-1481
栏目:
医学放射物理
出版日期:
2021-12-24

文章信息/Info

Title:
Statistical analysis of positioning errors based on OPS radiotherapy real-time positioning early warning system
文章编号:
1005-202X(2021)12-1477-05
作者:
王语汇1王东燕2周俊东2吴传锋2葛云1陈颖1黄晓林1李伟峰1
1.南京大学电子科学与工程学院, 江苏 南京 210023; 2.苏州市立医院东区放疗科, 江苏 苏州 215000
Author(s):
WANG Yuhui1 WANG Dongyan2 ZHOU Jundong2 WU Chuanfeng2 GE Yun1 CHEN Ying1 HUANG Xiaolin1 LI Weifeng1
1. School of Electronic Science and Engineering, Nanjing University, Nanjing 210023, China 2. Department of Radiotherapy, Suzhou Municipal Hospital (East District), Suzhou 215000, China
关键词:
红外定位系统放疗摆位误差统计实时预警
Keywords:
Keywords: optical positioning system radiotherapy positioning error statistics real-time early warning
分类号:
R318;R811.1
DOI:
DOI:10.3969/j.issn.1005-202X.2021.12.005
文献标志码:
A
摘要:
目的:利用红外定位系统(OPS)放疗实时摆位预警系统,对放疗摆位累积产生的海量数据,进行与设备性能及体态体姿相关的误差统计与监测,进而实时对摆位进行指导。方法:首先选取264例腹部癌症摆位数据、195例头颈部癌症摆位数据以及186例头部癌症摆位数据;然后搭建归一转换系统,量化标注不同摆位方法产生的不同摆位结果及差异,将现有的摆位结果数据统一转换到一个坐标系下;最后基于放疗摆位累积产生的海量数据,开发模型预警系统,实时监控等中心注册、相机移动、激光线偏移、体膜变形以及锥形束CT本身机械问题5种摆位误差。结果:OPS放疗实时摆位预警系统能够实时监控摆位数据,在腹部264例摆位数据中,5种误差分别为5、1、11、4、7例;在头颈部195例摆位数据中,5种误差分别为13、4、40、0、27例;在头部186例摆位数据中,5种误差分别为10、3、25、0、19例。结论:OPS放疗实时摆位预警系统能较为准确实时预警摆位质量和摆位精度,当摆位出现较大误差时,该系统能够迅速报警,及时指出摆位存在的问题,第一时间提醒、纠正存在的问题。
Abstract:
Abstract: Objective To analyze and monitor the errors related to equipment performance and posture by OPS radiotherapy real-time positioning early warning system based on the massive data accumulated by radiotherapy positioning, thereby guiding the positioning in real time. Methods The positioning data in the study were selected from 264 cases of abdominal cancer, 195 cases of head and neck cancer and 186 cases of head cancer. A normalization conversion system was build to quantify and label different positioning results and differences produced by different positioning methods, and then to convert the existing positioning results to a coordinate system. Finally, based on the massive data generated by radiotherapy positioning, an early warning system was developed for the real-time monitoring of 5 kinds of positioning errors which were caused by isocenter registration, camera movement, laser line deviation, body membrane deformation or CBCT mechanical problems. Results The OPS radiotherapy real-time positioning early warning system could monitor the positioning data in real time. The numbers of cases with 5 kinds of positioning errors were 5, 1, 11, 4 and 7 among 264 cases of abdominal cases, 13, 4, 40, 0 and 27 among 195 cases of head and neck cancer, and 10, 3, 25, 0 and 19 among 186 cases of head cancer. Conclusion OPS radiotherapy real-time positioning early warning system can provide more accurate real-time early warning of the quality and accuracy of the positioning. When there is a large positioning error, the system can quickly alarm, point out the problems existing in the positioning in time, and remind and correct it immediately.

相似文献/References:

[1]杨晓梅,储开岳,金建华,等.胸部定位参考线配合红外定位系统在鼻咽癌放疗中的应用[J].中国医学物理学杂志,2017,34(2):147.[doi:10.3969/j.issn.1005-202X.2017.02.007]
 Application of chest positioning reference line combined with infrared positioning system in[J].Chinese Journal of Medical Physics,2017,34(12):147.[doi:10.3969/j.issn.1005-202X.2017.02.007]
[2]黄新阳,周俊东,吴传锋,等.基于红外定位系统的放疗摆位偏差预警系统对头颈肩膜形变的分析与研究[J].中国医学物理学杂志,2022,39(4):421.[doi:DOI:10.3969/j.issn.1005-202X.2022.04.005]
 HUANG Xinyang,ZHOU Jundong,WU Chuanfeng,et al.Analyzing the deformation of head-neck-shoulder thermoplastic mask by OPS-based early warning system for setup errors in radiotherapy[J].Chinese Journal of Medical Physics,2022,39(12):421.[doi:DOI:10.3969/j.issn.1005-202X.2022.04.005]
[3]张羽祺,葛云,蔡晶,等.鼻咽癌放疗中红外定位系统与锥形束CT的摆位精度比较[J].中国医学物理学杂志,2021,38(2):162.[doi:DOI:10.3969/j.issn.1005-202X.2021.02.007]
 ZHANG Yuqi,GE Yun,CAI Jing,et al.Comparison of optical positioning system and cone-beam computed tomography on positioning accuracy in radiotherapy for nasopharyngeal carcinoma[J].Chinese Journal of Medical Physics,2021,38(12):162.[doi:DOI:10.3969/j.issn.1005-202X.2021.02.007]

备注/Memo

备注/Memo:
【收稿日期】2021-06-26 【基金项目】国家自然科学基金(81671777) 【作者简介】王语汇,硕士,研究方向:医学影像物理,E-mail: wanghyuhuiqd@163.com
更新日期/Last Update: 2021-12-24