[1]张羽祺,葛云,蔡晶,等.鼻咽癌放疗中红外定位系统与锥形束CT的摆位精度比较[J].中国医学物理学杂志,2021,38(2):162-167.[doi:DOI:10.3969/j.issn.1005-202X.2021.02.007]
 ZHANG Yuqi,GE Yun,CAI Jing,et al.Comparison of optical positioning system and cone-beam computed tomography on positioning accuracy in radiotherapy for nasopharyngeal carcinoma[J].Chinese Journal of Medical Physics,2021,38(2):162-167.[doi:DOI:10.3969/j.issn.1005-202X.2021.02.007]
点击复制

鼻咽癌放疗中红外定位系统与锥形束CT的摆位精度比较()
分享到:

《中国医学物理学杂志》[ISSN:1005-202X/CN:44-1351/R]

卷:
38卷
期数:
2021年第2期
页码:
162-167
栏目:
医学放射物理
出版日期:
2021-02-02

文章信息/Info

Title:
Comparison of optical positioning system and cone-beam computed tomography on positioning accuracy in radiotherapy for nasopharyngeal carcinoma
文章编号:
1005-202X(2021)02-0162-06
作者:
张羽祺1葛云1蔡晶2储开岳2金建华2葛斌斌2陈颖1黄晓林1李伟峰1
1.南京大学电子科学与工程学院, 江苏 南京 210023; 2.南通市肿瘤医院放射科, 江苏 南通 226300
Author(s):
ZHANG Yuqi1 GE Yun1 CAI Jing2 CHU Kaiyue2 JIN Jianhua2 GE Binbin2 CHEN Ying1 HUANG Xiaolin1 LI Weifeng1
1. School of Electronic Science and Engineering, Nanjing University, Nanjing 210023, China 2. Department of Radiation Oncology, Nantong Tumor Hospital, Nantong 226300, China
关键词:
鼻咽癌红外定位系统锥形束CT摆位精度放疗摆位
Keywords:
Keywords: nasopharyngeal carcinoma optical positioning system cone-beam computed tomography positioning accuracy radiotherapy positioning
分类号:
R811.1;R730.55
DOI:
DOI:10.3969/j.issn.1005-202X.2021.02.007
文献标志码:
A
摘要:
目的:量化研究鼻咽癌放射治疗过程中红外定位系统OPS与锥形束CT(CBCT)摆位精度的差异。方法:随机选取南通市肿瘤医院2018年4月~11月收治的鼻咽癌患者21例,使用头颈肩热塑膜固定体位,CT扫描前,在热塑膜上粘贴OPS专用定位球6个。首次治疗时使用千伏级CBCT进行位置验证,并移床校正,记录CBCT移床值和移床后的OPS监测偏差值。选择CBCT肿瘤中心摆位值作为理想放疗摆位参考,对室内激光定位系统的肿瘤中心摆位值和OPS肿瘤中心摆位值进行比较和分析。当摆位误差在2.0 mm内,则被认定与CBCT摆位临床一致。结果:与室内激光定位系统摆位相比,OPS大大提高了鼻咽癌放疗摆位的精度。OPS与CBCT在鼻咽癌放疗摆位一致性结果为:患者左右方向(用x轴表示)95.2%、腹背方向(用y轴表示)71.4%、头脚方向(用z轴表示)100.0%。腹背偏差较大主要是因为放射治疗计划系统选择颈椎骨作为参考点,此时选择CBCT摆位将无法消除颈椎骨弯曲产生的偏差。结论:OPS实现了在鼻咽癌放疗中实时、准确和快速的患者摆位,代替室内激光定位系统大大提高了鼻咽癌放疗摆位精度。在鼻咽癌放疗摆位精度方面,OPS和CBCT表现出较高的一致性,而且在每次放疗中均可使用。
Abstract:
Abstract: Objective To quantitatively study the differences in positioning accuracy between optical positioning system (OPS) and cone-beam computed tomography (CBCT) in the radiotherapy for nasopharyngeal carcinoma. Methods Twenty-one patients with nasopharyngeal carcinoma who were admitted to Nantong Cancer Hospital were enrolled in the study. All the patients were immobilized with thermoplastic head-and-shoulder mask. Before CT scan, 6 OPS infrared positioning balls were affixed to the mask. kV-CBCT scan was performed to validate the positioning errors before the first treatment, and the couch was then shifted according to the CBCT validation results. The couch shift when using CBCT and the deviations detected by OPS after couch shift were recorded. The positioning results of the room laser system and those of OPS were recorded, and then compared with those of CBCT which were taken as the reference. When the positioning errors were no more than 2.0 mm, OPS was considered consistent with CBCT in positioning. Results Compared with the room laser system, OPS remarkably improved the positioning accuracy of nasopharyngeal carcinoma radiotherapy. The consistencies between OPS and CBCT were 95.2%, 71.4% and 100.0% in left-right, anterior-posterior and superior-inferior directions, respectively. The error caused by the bending of the cervical vertebrae when the cervical vertebrae was taken as the reference point in treatment planning system couldnt be avoided in positioning using CBCT, which led to a large deviation between CBCT and OPS in anterior-posterior direction. Conclusion OPS which achieves real-time, accurate and rapid patient positioning can replace the room laser system to greatly improve the positioning accuracy in nasopharyngeal carcinoma radiotherapy. OPS also has a high consistency with CBCT in positioning in nasopharyngeal carcinoma radiotherapy. Moreover, OPS can be used in all fractions.

相似文献/References:

[1]陈广涛,汪 淇,刘亚强,等.TrueBeam加速器射野准直器跟随功能在鼻咽癌固定野调强放疗和容积旋转调强放疗中的应用[J].中国医学物理学杂志,2014,31(03):4865.[doi:10.3969/j.issn.1005-202X.2014.03.005]
[2]潘宇辉,陈榕钦,柏朋刚,等.鼻咽癌调强放射治疗前后腮腺体积及空间几何位置变化研究[J].中国医学物理学杂志,2014,31(05):5120.[doi:10.3969/j.issn.1005-202X.2014.05.005]
[3]张俊俊,邱小平,李奇欣,等.ArcCheck系统在鼻咽癌容积旋转调强剂量验证中的应用[J].中国医学物理学杂志,2014,31(05):5136.[doi:10.3969/j.issn.1005-202X.2014.05.009]
[4]桑 勇,陈维军,狄小云,等.多叶光栅叶片宽度对鼻咽癌剂量学的影响[J].中国医学物理学杂志,2014,31(06):5240.[doi:10.3969/j.issn.1005-202X.2014.06.004]
[5]黄利生,庄婷婷,伍方财,等.鼻咽癌调强放疗Ib区剂量分析[J].中国医学物理学杂志,2014,31(06):5265.[doi:10.3969/j.issn.1005-202X.2014.06.010]
[6]易金玲,金献测,周永强,等.鼻咽癌IMRT与VMAT治疗的计划与剂量验证比较研究[J].中国医学物理学杂志,2013,30(01):3859.[doi:10.3969/j.issn.1005-202X.2013.01.006]
[7]金献测,易金玲,周永强,等.鼻咽癌同步加量调强放射治疗新的计划质量评估指标研究[J].中国医学物理学杂志,2013,30(02):3978.[doi:10.3969/j.issn.1005-202X.2013.02.004]
[8]蒋晓芹,段宝风,艾平,等.基于图谱库的自动轮廓勾画软件(ABAS)在鼻咽癌调强放疗中的应用[J].中国医学物理学杂志,2013,30(02):3997.[doi:10.3969/j.issn.1005-202X.2013.02.008]
[9]张达光,蒋胜鹏,李智华,等.鼻咽癌调强放疗中解剖结构改变及剂量分布变化的研究[J].中国医学物理学杂志,2013,30(04):4239.[doi:10.3969/j.issn.1005-202X.2013.04.006]
[10]杨海燕,王军良,周振山,等.AAA算法和PBC算法在鼻咽癌调强放疗中剂量分布的比较[J].中国医学物理学杂志,2013,30(04):4245.[doi:10.3969/j.issn.1005-202X.2013.04.008]

备注/Memo

备注/Memo:
【收稿日期】2020-07-12 【基金项目】江苏省基础研究计划自然科学基金(BK20190309) 【作者简介】张羽祺,在读硕士,研究方向:医学物理,E-mail: 2992282906@qq.com 【通信作者】葛云,博士,教授,研究方向:医学物理,E-mail: geyun@nju.edu.cn
更新日期/Last Update: 2021-02-02