[1]闫冰,吴爱林,薛旭东,等.ArcCheck探测螺旋断层放疗误差的敏感性[J].中国医学物理学杂志,2021,38(6):666-671.[doi:DOI:10.3969/j.issn.1005-202X.2021.06.002]
 YAN Bing,WU Ailin,XUE Xudong,et al.Sensitivity of ArcCheck to detect dose delivery errors in helical tomotherapy[J].Chinese Journal of Medical Physics,2021,38(6):666-671.[doi:DOI:10.3969/j.issn.1005-202X.2021.06.002]
点击复制

ArcCheck探测螺旋断层放疗误差的敏感性()
分享到:

《中国医学物理学杂志》[ISSN:1005-202X/CN:44-1351/R]

卷:
38卷
期数:
2021年第6期
页码:
666-671
栏目:
医学放射物理
出版日期:
2021-06-29

文章信息/Info

Title:
Sensitivity of ArcCheck to detect dose delivery errors in helical tomotherapy
文章编号:
1005-202X(2021)06-0666-06
作者:
闫冰1吴爱林2薛旭东1张红雁1吴爱东1
1.中国科学技术大学附属第一医院/安徽省立医院放疗科, 安徽 合肥 230001; 2.中国科学技术大学附属第一医院西区/安徽省肿瘤医院放疗科, 安徽 合肥 230001
Author(s):
YAN Bing1 WU Ailin2 XUE Xudong1 ZHANG Hongyan1 WU Aidong1
1. Department of Radiation Oncology, the First Affiliated Hospital of University of Science and Technology of China/Anhui Provincial Hospital, Hefei 230001, China 2. Department of Radiation Oncology, the First Affiliated Hospital of University of Science and Technology of China (West District)/Anhui Provincial Cancer Hospital, Hefei 230001, China
关键词:
鼻咽癌螺旋断层放疗ArcCheck质量保证Gamma分析
Keywords:
Keywords: nasopharyngeal carcinoma helical tomotherapy ArcCheck quality assurance Gamma analysis
分类号:
R318;R815.2
DOI:
DOI:10.3969/j.issn.1005-202X.2021.06.002
文献标志码:
A
摘要:
目的:针对螺旋断层放疗(HT)系统,分析ArcCheck验证患者计划通过率对治疗床运动速度误差、机架旋转周期误差、机架起始角度误差、多叶准直器叶片开启时间误差的敏感性。方法:选取9例行HT的鼻咽癌患者计划,由自编程序生成与原计划相对应的误差计划。应用点剂量及Gamma分析,计算得出上述误差的临床可检测误差值。结果:所有患者的原计划采用3%/2 mm、2%/1 mm标准时的平均Gamma通过率分别为97.49%±1.08%、73.38%±4.31%。应用3%/2 mm通过率标准时,可检测的最小误差分别为治疗床运动速度:-1.58%、1.38%(上下限阈值),机器旋转周期:-1.68%、1.31%(上下限阈值),机架起始角度误差:2.50°,多叶准直器叶片开启时间:1.62%。而应用2%/1 mm标准时,检测精度得到提升,可检测的最小误差分别为治疗床运动速度:-0.69%、1.27%,机器旋转周期:-0.69%、0.69%,机架起始角度误差:2.06°,多叶准直器叶片开启时间:0.52%。结论:ArcCheck与电离室测量点剂量联用与单独使用ArcCheck Gamma通过率相比并未表现出明显优势,ArcCheck可以检测出临床相关的照射误差,更为严格的Gamma通过率标准可以明显提高误差计划的检测精度。
Abstract:
Abstract: Objective To measure the passing rates of helical tomotherapy (HT) plan with ArcCheck, and to explore the sensitivity of ArcCheck to detect the errors which are introduced in couch movement, gantry rotation period, gantry starting angle and leaf open time. Methods The plans of 9 patients with nasopharyngeal carcinoma who underwent HT were analyzed. For each patient, error plan corresponding to the original plan was generated using a self-compiled program. Point dose and Gamma analysis were used for calculating clinically detectable errors. Results The average Gamma passing rates of the 9 original HT plans were 97.49%±1.08% and 73.38%±4.31% under the criteria of 3%/2 mm and 2%/1 mm, respectively. When adopting the criterion of 3%/2mm, the detectable errors were -1.58% and 1.38% (upper and lower thresholds) for couch movement, -1.68% and 1.31% (upper and lower thresholds) for gantry rotation period, 2.50° for gantry starting angle, 1.62% for leaf open time. When adopting the criterion of 2%/1mm, the detection accuracy was improved, and the corresponding minimum detectable errors were -0.69% and 1.27% for couch movement, -0.69% and 0.69% for gantry rotation period, 2.06°for gantry starting angle, and 0.52% for leaf open time, respectively. Conclusion Compared with ArcCheck only, the combination of ion chamber measurements with ArcCheck do not improve error detection accuracy. ArcCheck can be used to detect clinically relevant dose delivery errors. Selecting more strict criteria for Gamma passing rate calculation can significantly improve the detection accuracy of error plan.

相似文献/References:

[1]张富利,王雅棣,许卫东,等.应用两种三维探测器阵列进行螺旋断层调强放疗计划剂量验证[J].中国医学物理学杂志,2015,32(02):218.[doi:10.3969/j.issn.1005-202X.2015.02.015]
[2]巩汉顺,王金媛,丛小虎,等.晚期胰腺癌适形、简化调强及螺旋断层放疗的剂量学评估[J].中国医学物理学杂志,2015,32(01):8.[doi:10.3969/j.issn.1005-202X.2015.01.003]
[3]陈广涛,汪 淇,刘亚强,等.TrueBeam加速器射野准直器跟随功能在鼻咽癌固定野调强放疗和容积旋转调强放疗中的应用[J].中国医学物理学杂志,2014,31(03):4865.[doi:10.3969/j.issn.1005-202X.2014.03.005]
[4]朱夫海,吴伟章,王 勇,等.肺癌螺旋断层放疗计划设计的初步研究[J].中国医学物理学杂志,2014,31(04):4979.[doi:10.3969/j.issn.1005-202X.2014.04.002]
[5]潘宇辉,陈榕钦,柏朋刚,等.鼻咽癌调强放射治疗前后腮腺体积及空间几何位置变化研究[J].中国医学物理学杂志,2014,31(05):5120.[doi:10.3969/j.issn.1005-202X.2014.05.005]
[6]张俊俊,邱小平,李奇欣,等.ArcCheck系统在鼻咽癌容积旋转调强剂量验证中的应用[J].中国医学物理学杂志,2014,31(05):5136.[doi:10.3969/j.issn.1005-202X.2014.05.009]
[7]张富利,蒋华勇,王雅棣,等.应用一维、二维、三维探测器阵列进行螺旋断层调强放疗计划剂量验证[J].中国医学物理学杂志,2014,31(06):5230.[doi:10.3969/j.issn.1005-202X.2014.06.002]
[8]桑 勇,陈维军,狄小云,等.多叶光栅叶片宽度对鼻咽癌剂量学的影响[J].中国医学物理学杂志,2014,31(06):5240.[doi:10.3969/j.issn.1005-202X.2014.06.004]
[9]黄利生,庄婷婷,伍方财,等.鼻咽癌调强放疗Ib区剂量分析[J].中国医学物理学杂志,2014,31(06):5265.[doi:10.3969/j.issn.1005-202X.2014.06.010]
[10]易金玲,金献测,周永强,等.鼻咽癌IMRT与VMAT治疗的计划与剂量验证比较研究[J].中国医学物理学杂志,2013,30(01):3859.[doi:10.3969/j.issn.1005-202X.2013.01.006]
[11]黄栋,田晓云,刘海,等.鼻咽癌螺旋断层放疗MegaVoltage CT引导下摆位误差分析[J].中国医学物理学杂志,2016,33(2):204.[doi:10.3969/j.issn.1005-202X.2016.02.020]

备注/Memo

备注/Memo:
【收稿日期】2021-02-10 【基金项目】国家自然科学基金(11805198);安徽省自然科学基金(1808085QH281);安徽省学术和技术带头人后备人选科研项目(2020H230) 【作者简介】闫冰,硕士,工程师,研究方向:放射物理,E-mail: cpreo@126.com 【通信作者】吴爱东,博士,正高级工程师,研究方向:放射物理,E-mail: flkaidongwu@163.com
更新日期/Last Update: 2021-06-29