保乳术后放疗中钛夹动度与乳腺大小及钛夹位置的相关性()
分享到:

《中国医学物理学杂志》[ISSN:1005-202X/CN:44-1351/R]

卷:
39卷
期数:
2022年第5期
页码:
529-534
栏目:
医学放射物理
出版日期:
2022-05-27

文章信息/Info

Title:
Correlations of mobility of titanium clips in radiotherapy after breast-conserving surgery with breast size and positions of titanium clips
文章编号:
1005-202X(2022)05-0529-06
作者:
罗娇娇1彭海燕1罗焕丽1周宪1曾晓华2靳富1
1.重庆大学附属肿瘤医院肿瘤放射治疗中心, 重庆 400030; 2.重庆大学附属肿瘤医院乳腺肿瘤中心, 重庆 400030
Author(s):
LUO Jiaojiao1 PENG Haiyan1 LUO Huanli1 ZHOU Xian1 ZENG Xiaohua2 JIN Fu1
1. Department of Radiation Oncology, Cancer Hospital Affiliated to Chongqing University, Chongqing 400030, China 2. Department of Breast Cancer, Cancer Hospital Affiliated to Chongqing University, Chongqing 400030, China
关键词:
乳腺癌保乳术后放疗锥形束CT钛夹
Keywords:
Keywords: breast cancer radiotherapy after breast-conserving surgery cone-beam computed tomography titanium clip
分类号:
R811.1
DOI:
DOI:10.3969/j.issn.1005-202X.2022.05.001
文献标志码:
A
摘要:
【摘要】目的:分析保乳术后放疗患者钛夹动度,并探讨其与乳腺大小、钛夹位置的相关性。方法:随机选取保乳术后放疗患者15例,每周两次锥形束CT。测量乳腺基底面直径(D)和乳高轴(H),以瘤床质心为原点将临床靶区分4个象限。记录各象限内钛夹在左右、腹背、头脚方向动度(MLR、MAP、MSI)以及其与临床靶区最内、最前、最上的距离(DSLR、DSAP、DSSI)。结果:MLR、MAP、MSI分别为(2.2±3.0)、(-1.1±3.6)、(0.8±4.7) mm;系统误差Σ在左、右、腹、背、头、脚分别为1.7、2.2、2.0、2.4、2.9、3.0 mm,随机误差σ分别为2.4、4.0、3.2、4.0、4.7、4.7 mm;靶区对应外放5.9、8.3、7.2、8.8、10.5、10.9 mm。一象限内,当D×H<99.89 cm2,MAP和D×H强相关(r=0.805),MLR、MAP均和DSLR、DSAP、DSSI强正相关(r=0.94, 0.94, 0.91;0.87, 0.91, 0.92),MSI和DSLR、DSAP强正相关(r=0.91, 0.94);四象限内,当D×H<90.71 cm2,MAP和DSLR,MSI和DSAP均强负相关(r=-0.96;-0.95),MLR和DSLR强正相关(r=0.91)。结论:钛夹动度有各向异性,以SI方向外扩最大,并易受乳腺大小、钛夹位置影响。
Abstract:
Abstract: Objective To analyze the mobility of titanium clips in patients receiving radiotherapy after breast-conserving surgery, and explore its correlations with the size of the breast and the positions of the titanium clips. Methods Fifteen patients receiving radiotherapy after breast-conserving surgery were randomly enrolled, and each patient underwent CBCT twice a week. The basal surface diameter (D) and the breast height axis (H) of the mammary gland were measured, and the clinical target area was divided into 4 quadrants with the centroid of the tumor bed as the origin. In each quadrant, the mobility (MLR, MAP, MSI) of the titanium clip in the left-right (LR), anterior-posterior (AP) and superior-inferior (SI) directions, and the distances (DSLR, DSAP, DSSI) from the titanium clip to the innermost, foremost and uppermost of the clinical target area were recorded. Results MLR, MAP and MSI were (2.2±3.0), (-1.1±3.6) and (0.8±4.7) mm, respectively. The systematic errors Σ in the left, right, abdomen, back, head and feet directions were 1.7, 2.2, 2.0, 2.4, 2.9 and 3.0 mm, and the random errors σ were 2.4, 4.0, 3.2, 4.0, 4.7 and 4.7 mm and the corresponding margins of target areas were 5.9, 8.3, 7.2, 8.8, 10.5 and 10.9 mm. In the first quadrant, when D×H<99.89 cm2, MAP and D×H were strongly correlated (r=0.805) and MLR and MAP were strongly positively correlated with DSLR, DSAP, and DSSI (r=0.94, 0.94, 0.91 0.87, 0.91, 0.92) and MSI was strongly positively correlated with DSLR and DSAP (r=0.91, 0.94). In the fourth quadrant, when D×H<90.71 cm2, there were strongly negative correlations between MAP and DSLR (r=-0.96), between MSI and DSAP (r=-0.95), while MLR had a strong positive correlation with DSLR (r=0.91). Conclusion The mobility of the titanium clips is anisotropic, with the largest margin in the SI direction, and it is easily affected by the size of the breast and the positions of the titanium clips.

相似文献/References:

[1]柏 晗,王 丽,刘旭红,等.左乳腺5种不同主野分布IMRT计划的剂量比较分析[J].中国医学物理学杂志,2015,32(02):210.[doi:10.3969/j.issn.1005-202X.2015.02.013]
[2]张晋建,黎 静,文 婷,等.Arc CHECK在乳腺癌螺旋断层放疗计划剂量验证中的应用[J].中国医学物理学杂志,2015,32(02):244.[doi:10.3969/j.issn.1005-202X.2015.02.021]
[3]白 雪,杜向慧,王升晔,等.半野技术结合非共面照射技术在乳腺癌胸锁联合照射中的应用[J].中国医学物理学杂志,2014,31(03):4849.[doi:10.3969/j.issn.1005-202X.2014.03.001]
[4]赵永亮,储开岳,吴建亭,等.两种不同调强技术配合主动呼吸控制系统在乳腺癌放疗中的比较[J].中国医学物理学杂志,2014,31(03):4862.[doi:10.3969/j.issn.1005-202X.2014.03.004]
[5]花威,张西志,张先稳,等.RapidArc在乳腺癌术后放疗中的应用[J].中国医学物理学杂志,2013,30(02):3984.[doi:10.3969/j.issn.1005-202X.2013.02.005]
[6]徐金济,成俊,李伟,等.左侧乳腺癌改良根治术后逆向调强放射治疗的摆位误差分析[J].中国医学物理学杂志,2013,30(05):4358.[doi:10.3969/j.issn.1005-202X.2013.05.004]
[7]孟慧鹏,孙小喆,孙劲松,等.右乳腺癌术后放疗中VMAT和dIMRT的剂量学比较[J].中国医学物理学杂志,2013,30(05):4387.[doi:10.3969/j.issn.1005-202X.2013.05.011]
[8]刘旭红,陈晓,艾毅钦,等.早期左乳腺癌保乳术后仰卧位和俯卧位照射剂量学比较与分析[J].中国医学物理学杂志,2015,32(05):737.[doi:doi:10.3969/j.issn.1005-202X.2015.05.028]
 [J].Chinese Journal of Medical Physics,2015,32(5):737.[doi:doi:10.3969/j.issn.1005-202X.2015.05.028]
[9]张艺宝,吴昊,李莎,等.临床前验证与几何对比分析基于图谱库的危及器官自动勾画[J].中国医学物理学杂志,2015,32(06):761.[doi:doi:10.3969/j.issn.1005-202X.2015.06.001]
 [J].Chinese Journal of Medical Physics,2015,32(5):761.[doi:doi:10.3969/j.issn.1005-202X.2015.06.001]
[10]段云龙,杨涛,曲宝林,等.乳腺癌保乳切线野三维适形、切线野正向调强及逆向调强放疗的剂量学评估[J].中国医学物理学杂志,2015,32(06):841.[doi:doi:10.3969/j.issn.1005-202X.2015.06.017]
 [J].Chinese Journal of Medical Physics,2015,32(5):841.[doi:doi:10.3969/j.issn.1005-202X.2015.06.017]

备注/Memo

备注/Memo:
【收稿日期】2021-12-03 【基金项目】国家自然科学基金(11805025, 81972857) 【作者简介】罗娇娇,研究方向:乳腺肿瘤放射治疗,E-mail: 1146783223@qq.com 【通信作者】靳富,博士,研究方向:肿瘤放射治疗,E-mial: jfazj@126.com
更新日期/Last Update: 2022-05-27