[1]柳叶潇,徐昱琳,赵翠莲.基于肌电信号分析研究柔性电极设计参数对电极性能的影响[J].中国医学物理学杂志,2019,36(7):830-835.[doi:DOI:10.3969/j.issn.1005-202X.2019.07.016]
 LIU Yexiao,XU Yulin,ZHAO Cuilian.Effects of design parameters of flexible electrode on electrode performance: a research based onelectromyography signal analysis[J].Chinese Journal of Medical Physics,2019,36(7):830-835.[doi:DOI:10.3969/j.issn.1005-202X.2019.07.016]
点击复制

基于肌电信号分析研究柔性电极设计参数对电极性能的影响()
分享到:

《中国医学物理学杂志》[ISSN:1005-202X/CN:44-1351/R]

卷:
36卷
期数:
2019年第7期
页码:
830-835
栏目:
医学信号处理与医学仪器
出版日期:
2019-07-25

文章信息/Info

Title:
Effects of design parameters of flexible electrode on electrode performance: a research based on electromyography signal analysis
文章编号:
1005-202X(2019)07-0830-06
作者:
柳叶潇徐昱琳赵翠莲
上海大学机电工程与自动化学院,上海200444
Author(s):
LIU Yexiao XU Yulin ZHAO Cuilian
School of Mechatronic Engineering and Automation, Shanghai University, Shanghai 200444, China
关键词:
柔性电极表面肌电设计参数时频域
Keywords:
flexible electrode surface electromyography design parameter time-frequency domain
分类号:
R318;TH77
DOI:
DOI:10.3969/j.issn.1005-202X.2019.07.016
文献标志码:
A
摘要:
柔性电极因其穿戴式表面肌电采集的优势而得到研究学者的广泛关注,但很少有研究针对柔性电极的设计参数提 供明确的指导和规范。本实验首先对可能影响电极性能的参数进行了分析;然后利用柔性印刷工艺与喷墨印刷工艺制作 5种不同规格的柔性干电极,并将其与商用湿电极进行对比,设计基于尺侧腕伸肌的肌电采集实验,得到100个表面肌电 信号段;最后对信号段的时频域特性进行分析,从而确定相关参数对电极性能的影响。结果表明增加电极高度后的有源 柔性印刷电路板电极具有更好的信号采集性能。
Abstract:
Abstract: Flexible electrode has received wide attention from researchers because of its advantages in wearable surface electromyography signal acquisition. However, few studies have provided clear guidance and specifications for the design parameters of flexible electrode. Herein the parameters that may affect the performance of the electrode are analyzed. Five kinds of flexible dry electrodes of different specifications fabricated by flexible printing process and inkjet printing process are compared with the commercial wet electrode. The electromyography signal acquisition experimentation based on ulnar wrist extensors is designed to obtain 100 surface electromyography signal segments. Finally, the time-frequency domain characteristics of the signal segment are analyzed, thereby determining the effect of the relevant parameters on the electrode performance. The results show that the active flexible printed circuit board electrodes after increasing the electrode height have a better signal acquisition performance.

相似文献/References:

[1]王改云,薛漭潮,王磊杨. 一种可穿戴式心电图动态监护系统[J].中国医学物理学杂志,2018,35(4):431.[doi:DOI:10.3969/j.issn.1005-202X.2018.04.012]
 WANG Gaiyun,XUE Mangchao,WANG Leiyang. A wearable ECG dynamic monitoring system[J].Chinese Journal of Medical Physics,2018,35(7):431.[doi:DOI:10.3969/j.issn.1005-202X.2018.04.012]

备注/Memo

备注/Memo:
【收稿日期】2019-02-20 【基金项目】上海市科学技术委员会资助项目(16441909000) 【作者简介】柳叶潇,硕士研究生,研究方向:生物肌电信号采集,E-mail: liuyx1234@126.com;徐昱琳,副教授,研究方向:智能控制、 机器人技术,E-mail: xuyulin@shu.edu.cn 【通信作者】赵翠莲,教授,研究方向:生机电一体化与康复工程,E-mail: clzhao@shu.edu.cn
更新日期/Last Update: 2019-07-25