[1]雷金艳,曹潘潘,张倩,等.特定密度的3D打印补偿膜在乳腺癌放疗中的临床应用[J].中国医学物理学杂志,2024,41(6):683-689.[doi:DOI:10.3969/j.issn.1005-202X.2024.06.005]
 LEI Jinyan,CAO Panpan,ZHANG Qian,et al.Clinical application of 3D printed bolus with specific density in breast cancer radiotherapy[J].Chinese Journal of Medical Physics,2024,41(6):683-689.[doi:DOI:10.3969/j.issn.1005-202X.2024.06.005]
点击复制

特定密度的3D打印补偿膜在乳腺癌放疗中的临床应用()
分享到:

《中国医学物理学杂志》[ISSN:1005-202X/CN:44-1351/R]

卷:
41卷
期数:
2024年第6期
页码:
683-689
栏目:
医学放射物理
出版日期:
2024-06-25

文章信息/Info

Title:
Clinical application of 3D printed bolus with specific density in breast cancer radiotherapy
文章编号:
1005-202X(2024)06-0683-07
作者:
雷金艳曹潘潘张倩李建锋代馨瑶王雪纯靳富彭海燕
重庆大学附属肿瘤医院放射物理中心, 重庆 400030
Author(s):
LEI Jinyan CAO Panpan ZHANG Qian LI Jianfeng DAI Xinyao WANG Xuechun JIN Fu PENG Haiyan
Radiation Physics Center, Chongqing University Cancer Hospital, Chongqing 400030, China
关键词:
乳腺癌3D打印补偿膜Catalyst HD剂量分布
Keywords:
Keywords: breast cancer 3D printed bolus Catalyst HD dose distribution
分类号:
R318;R811.1
DOI:
DOI:10.3969/j.issn.1005-202X.2024.06.005
文献标志码:
A
摘要:
目的:探究特定密度的3D打印补偿膜在乳腺癌放疗中的临床应用,并评估其对剂量分布和放疗摆位的影响。方法:随机选取行调强放疗的乳腺癌切除术后患者40例,使用3D打印补偿膜与常规补偿膜各20例,均采用发泡胶仰卧位固定。基于室内激光和体表标记进行常规摆位,每日Catalyst HD光学体表引导结合每周一次CBCT验证。记录不同补偿膜下的绝对剂量、患者皮肤表面剂量、手术切口、计划布野、靶区剂量(VCTV 50 Gy、VPTV 50 Gy)和危及器官受量,并计算适形度指数和均匀性指数;同时,记录患者的CBCT及Catalyst HD摆位误差。结果:不同补偿膜下的绝对剂量差异无统计学意义(P>0.05),3D打印补偿下的皮肤表面剂量显著高于常规补偿(P<0.05),二者分别为(54.83±0.44) Gy和(54.43±0.51) Gy。使用3D打印补偿膜的患者较常规补偿膜的适形度指数更高,二者分别为0.69±0.04和0.65±0.02。基于不同补偿膜,VCTV 50 Gy差异无统计学意义(P>0.05),3D打印补偿膜的患者VPTV 50 Gy略低于常规补偿膜,且危及器官受量更低(P<0.05),心脏Vmean分别为9.68%±3.24%和11.43%±3.60%。3D打印补偿膜的患者中,计划布野及手术切口对靶区剂量均存在影响,不包内乳的靶区剂量较包内乳更大(P<0.05)。当布野不包内乳时,不同手术切口仅对VPTV 50 Gy存在影响,且横梭形较斜竖形切口的VPTV 50 Gy更高(P<0.05),二者分别为95.58%±0.51%和95.44%±0.71%。3D打印与常规补偿膜的光学监测误差仅在左右方向存在差异,分别为(0.08±0.57) cm和(-0.15±0.46) cm(P<0.05)。结论:与常规补偿膜相比,3D打印补偿膜可提高剂量分布和光学监测误差;同时3D打印补偿膜下的手术切口和计划布野对靶区剂量均存在一定影响。
Abstract:
Abstract: Objective To investigate the clinical application of 3D printed bolus with specific density in breast cancer radiotherapy, and to evaluate its effects on dose distribution and positioning. Methods Forty post-mastectomy patients undergoing intensity-modulated radiotherapy were randomly enrolled for 3D printed bolus (n=20) and conventional bolus (n=20), and all patients were fixed in the supine position using styrofoam. Conventional positioning was performed based on in-room lasers and body markers, with daily Catalyst HD optical surface monitoring combined with weekly CBCT verification. The absolute dose, patients skin surface dose, surgical incision, planned field, target area doses (VCTV 50 Gy, VPTV 50 Gy) and organs-at-risk doses in patients with different boluses were recorded, and the conformity index and homogeneity index were calculated, and the setup errors using CBCT and Catalyst HD were also analyzed. Results The difference in absolute dose between different boluses was trivial, but the skin surface dose with 3D printed bolus was significantly higher than with conventional bolus [(54.83±0.44) Gy vs (54.43±0.51) Gy, P<0.05]. Patients with 3D printed boluses had a higher conformity index than with conventional boluses (0.69±0.04 vs 0.65±0.02). For different boluses, there was no significant difference in VCTV 50 Gy, while the VPTV 50 Gy and organs-at-risk doses were lower in those with 3D printed bolus than conventional boluses (P<0.05), with heart Vmean of 9.68%±3.24% vs 11.43%±3.60%. In patients with 3D printed boluses, both planned field arrangement and surgical incision affected the target doses, and the doses of the target area without an internal breast wrap was greater than those with internal breast wrap (P<0.05). When the field was not wrapped around the internal breast, the surgical incisions only affected VPTV 50 Gy, and the VPTV 50 Gy was greater with the transverse fusiform incision than with the oblique vertical incision (P<0.05), which were 95.58%±0.51% vs 95.44%±0.71%. The optical monitoring accuracies with different boluses differed only in the left-right direction (P<0.05), with (0.08±0.57) cm and (-0.15±0.46) cm for 3D printed and conventional boluses. Conclusion Compared with conventional bolus, 3D printed bolus can improve dose distribution and optical monitoring accuracy. The surgical incision and planned field arrangement under 3D-printed bolus would exert effects on target doses.

相似文献/References:

[1]柏 晗,王 丽,刘旭红,等.左乳腺5种不同主野分布IMRT计划的剂量比较分析[J].中国医学物理学杂志,2015,32(02):210.[doi:10.3969/j.issn.1005-202X.2015.02.013]
[2]张晋建,黎 静,文 婷,等.Arc CHECK在乳腺癌螺旋断层放疗计划剂量验证中的应用[J].中国医学物理学杂志,2015,32(02):244.[doi:10.3969/j.issn.1005-202X.2015.02.021]
[3]白 雪,杜向慧,王升晔,等.半野技术结合非共面照射技术在乳腺癌胸锁联合照射中的应用[J].中国医学物理学杂志,2014,31(03):4849.[doi:10.3969/j.issn.1005-202X.2014.03.001]
[4]赵永亮,储开岳,吴建亭,等.两种不同调强技术配合主动呼吸控制系统在乳腺癌放疗中的比较[J].中国医学物理学杂志,2014,31(03):4862.[doi:10.3969/j.issn.1005-202X.2014.03.004]
[5]花威,张西志,张先稳,等.RapidArc在乳腺癌术后放疗中的应用[J].中国医学物理学杂志,2013,30(02):3984.[doi:10.3969/j.issn.1005-202X.2013.02.005]
[6]徐金济,成俊,李伟,等.左侧乳腺癌改良根治术后逆向调强放射治疗的摆位误差分析[J].中国医学物理学杂志,2013,30(05):4358.[doi:10.3969/j.issn.1005-202X.2013.05.004]
[7]孟慧鹏,孙小喆,孙劲松,等.右乳腺癌术后放疗中VMAT和dIMRT的剂量学比较[J].中国医学物理学杂志,2013,30(05):4387.[doi:10.3969/j.issn.1005-202X.2013.05.011]
[8]刘旭红,陈晓,艾毅钦,等.早期左乳腺癌保乳术后仰卧位和俯卧位照射剂量学比较与分析[J].中国医学物理学杂志,2015,32(05):737.[doi:doi:10.3969/j.issn.1005-202X.2015.05.028]
 [J].Chinese Journal of Medical Physics,2015,32(6):737.[doi:doi:10.3969/j.issn.1005-202X.2015.05.028]
[9]张艺宝,吴昊,李莎,等.临床前验证与几何对比分析基于图谱库的危及器官自动勾画[J].中国医学物理学杂志,2015,32(06):761.[doi:doi:10.3969/j.issn.1005-202X.2015.06.001]
 [J].Chinese Journal of Medical Physics,2015,32(6):761.[doi:doi:10.3969/j.issn.1005-202X.2015.06.001]
[10]段云龙,杨涛,曲宝林,等.乳腺癌保乳切线野三维适形、切线野正向调强及逆向调强放疗的剂量学评估[J].中国医学物理学杂志,2015,32(06):841.[doi:doi:10.3969/j.issn.1005-202X.2015.06.017]
 [J].Chinese Journal of Medical Physics,2015,32(6):841.[doi:doi:10.3969/j.issn.1005-202X.2015.06.017]

备注/Memo

备注/Memo:
【收稿日期】2024-02-06 【基金项目】重庆市科卫联合医学科研项目(2022DBXM005);重庆市沙坪坝区技术创新与应用发展项目(2023129);重庆市教育委员会科学技术研究项目(KJQN202300126) 【作者简介】雷金艳,主管技师,研究方向:肿瘤放射治疗,E-mail: ljy321221@163.com 【通信作者】彭海燕,主管技师,研究方向:肿瘤放射治疗,E-mail: 13668175902@163.com
更新日期/Last Update: 2024-06-25