[1]朱远湖,佘广南,王学涛,等.一种新型热塑膜定位器在放疗体位固定中的摆位精度和剂量精度[J].中国医学物理学杂志,2023,40(7):801-807.[doi:DOI:10.3969/j.issn.1005-202X.2023.07.002]
 ZHU Yuanhu,SHE Guangnan,et al.Positioning accuracy and dose accuracy in radiotherapy using a novel positioner for thermoplastic immobilization[J].Chinese Journal of Medical Physics,2023,40(7):801-807.[doi:DOI:10.3969/j.issn.1005-202X.2023.07.002]
点击复制

一种新型热塑膜定位器在放疗体位固定中的摆位精度和剂量精度()
分享到:

《中国医学物理学杂志》[ISSN:1005-202X/CN:44-1351/R]

卷:
40卷
期数:
2023年第7期
页码:
801-807
栏目:
其他(激光医学等)
出版日期:
2023-07-15

文章信息/Info

Title:
Positioning accuracy and dose accuracy in radiotherapy using a novel positioner for thermoplastic immobilization
文章编号:
1005-202X(2023)07-0801-07
作者:
朱远湖12佘广南3王学涛2谭翔2朱琳2何强2简婉薇2张白霖2孟皓宇2甄鑫1
1.南方医科大学生物医学工程学院, 广东 广州 510515; 2.广州中医药大学第二附属医院放射治疗区, 广东 广州 510006; 3.中山大学附属第六医院数字医院建设处, 广东 广州 510655
Author(s):
ZHU Yuanhu1 2 SHE Guangnan3 WANG Xuetao2 TAN Xiang2 ZHU Lin2 HE Qiang2 JIAN Wanwei2 ZHANG Bailin2 MENG Haoyu2 ZHEN Xin1
1. School of Biomedical Engineering, Southern Medical University, Guangzhou 510515, China 2. Department of Radiation Therapy, The Second Affiliated Hospital of Guangzhou University of Chinese Medicine, Guangzhou 510006, China 3. Department of Digital Hospital Construction, The Sixth Affiliated Hospital, Sun Yat-sen University, Guangzhou 510655, China
关键词:
放射治疗体位固定热塑膜定位器
Keywords:
Keywords: radiotherapy positioning thermoplastic film positioner
分类号:
R318;R811.1
DOI:
DOI:10.3969/j.issn.1005-202X.2023.07.002
文献标志码:
A
摘要:
目的:研究放疗热塑膜固定中有无热塑膜定位器的摆位时间差异、体位精度差异和剂量精度差异。方法:随机选取需应用热塑膜固定的放疗患者3例,每个病人有一半的治疗次数不使用定位器,有一半的治疗次数使用定位器。记录摆位时间、MVCT扫描匹配后的三维方向摆位误差、三维空间矢量位移误差和剂量精度误差,应用2.5∑+0.7σ来计算计划靶区(PTV)外扩边界。结果:无热塑膜定位器组和有热塑膜定位器组的误差在X轴方向分别为6.90(2.85, 12.10) mm和2.60(1.45, 8.10) mm(P=0.007),Y轴方向分别为9.30(4.05, 16.15) mm和5.00(2.60, 9.50) mm(P=0.002),Z轴方向分别为3.40(2.05, 4.10) mm和1.90(1.30, 3.65) mm (P=0.017)。无热塑膜定位器组和有热塑膜定位器组在X、Y、Z轴的PTV外扩边界分别为9.15、14.62、2.15 mm和6.50、7.04、2.97 mm。结论:热塑膜定位器在热塑膜固定中的应用能够减少放疗患者三维空间摆位误差,并且减少PTV外扩边界。
Abstract:
Abstract: Objective To investigate the differences in positioning time, positional accuracy, and dose accuracy between the use and non-use of thermoplastic film positioners in radiotherapy immobilization. Methods Three patients requiring thermoplastic immobilization for radiotherapy were enrolled, with half of the treatments using a thermoplastic film positioner and the other half without. The positioning time, three-dimensional directional positioning errors after MVCT scanning matching, three-dimensional spatial vector displacement errors, and dose accuracy errors were recorded. The PTV margin was calculated using 2.5∑+0.7σ. Results The errors of no-positioner group and positioner group were 6.90 (2.85, 12.10) mm versus 2.60 (1.45, 8.10) mm in the X-axis direction (P=0.007), 9.30 (4.05, 16.15) mm versus 5.00 (2.60, 9.50) mm in the Y-axis direction (P=0.002), 3.40 (2.05, 4.10) mm versus 1.90 (1.30, 3.65) mm in the Z-axis direction (P=0.017), respectively. The PTV margins in the X, Y and Z axes were 9.15, 14.62, 2.15 mm in no-positioner group, and 6.50, 7.04, 2.97 mm in positioner group, respectively. Conclusion The application of positioners in thermoplastic immobilization can reduce the three-dimensional spatial positioning errors and minimize PTV margin.

相似文献/References:

[1]武 杰,袁杭英,欧阳蓓蕾.内旋式放射治疗用面罩固定装置的设计[J].中国医学物理学杂志,2015,32(02):265.[doi:10.3969/j.issn.1005-202X.2015.02.027]
[2]司 森,吴鑫全,包尚联,等.基于GPU的蒙特卡洛放疗剂量模拟[J].中国医学物理学杂志,2015,32(01):1.[doi:10.3969/j.issn.1005-202X.2015.01.001]
[3]王 文,程梦云,杨 琪,等.基于MCNP源子程序的放射治疗剂量计算验证方法[J].中国医学物理学杂志,2015,32(01):13.[doi:10.3969/j.issn.1005-202X.2015.01.004]
[4]张一戈,邱小平,陈维军,等.基于图像引导技术对头颈部肿瘤临床摆位误差的剂量学影响[J].中国医学物理学杂志,2015,32(01):25.[doi:10.3969/j.issn.1005-202X.2015.01.007]
[5]孟慧鹏,孙小喆,王 昊,等.胸部偏后背肿瘤RapidArc采用仰卧位设计治疗计划的可行性分析[J].中国医学物理学杂志,2015,32(01):31.[doi:10.3969/j.issn.1005-202X.2015.01.008]
[6]庞皓文,孙小扬.基于快速自由形变的弹性配准算法在放疗CT中的应用[J].中国医学物理学杂志,2015,32(03):404.[doi:10.3969/j.issn.1005-202X.2015.03.023]
[7]贾卫娟,周一兵,谭维维,等.利用模拟机及挡铅托架进行放疗前的治疗模拟[J].中国医学物理学杂志,2014,31(01):4622.[doi:10.3969/j.issn.1005-202X.2014.01.007]
[8]白 雪,杜向慧,王升晔,等.半野技术结合非共面照射技术在乳腺癌胸锁联合照射中的应用[J].中国医学物理学杂志,2014,31(03):4849.[doi:10.3969/j.issn.1005-202X.2014.03.001]
[9]周 钢,田 野,陆雪官,等.乳腺癌保乳术后三种瘤床同期加量放疗技术的剂量学比较[J].中国医学物理学杂志,2014,31(03):4853.[doi:10.3969/j.issn.1005-202X.2014.03.002]
[10]李贤富,谢 力,郭飞,等.医科达Precise直线加速器快速应急处理方法探讨[J].中国医学物理学杂志,2014,31(04):4984.[doi:10.3969/j.issn.1005-202X.2014.04.003]
[11]王颖,陈飞.头颈部面罩和胸部体罩固定对胸上段食管癌放疗剂量分布的影响[J].中国医学物理学杂志,2022,39(2):157.[doi:DOI:10.3969/j.issn.1005-202X.2022.02.005]
 WANG Ying,CHEN Fei.Effects of immobilizations with head and neck mask or chest body covering on dose distribution in the radiotherapy for upper thoracic esophageal carcinoma[J].Chinese Journal of Medical Physics,2022,39(7):157.[doi:DOI:10.3969/j.issn.1005-202X.2022.02.005]

备注/Memo

备注/Memo:
【收稿日期】2023-02-19 【基金项目】国家自然科学基金(81874216);广东省基础与应用基础研究基金(2022A1515011410);广州市基础研究计划基础与应用基础研究项目(81874216);广州市科技计划项目(202102010264) 【作者简介】朱远湖,工程师,研究方向:肿瘤放射物理学、肿瘤放射治疗技术等,E-mail: 1275841716@qq.com 【通信作者】甄鑫,博士,教授,研究方向:肿瘤放射物理学、医学图像处理、医学人工智能等,E-mail: xinzhen@smu.edu.cn
更新日期/Last Update: 2023-07-15