[1]袁瑜含,谷雪莲,崔建昆,等.人工静脉瓣膜周期变化中的力学性能研究[J].中国医学物理学杂志,2023,40(7):917-924.[doi:DOI:10.3969/j.issn.1005-202X.2023.07.020]
 YUAN Yuhan,GU Xuelian,CUI Jiankun,et al.Mechanical properties during periodical variation of prosthetic venous valve[J].Chinese Journal of Medical Physics,2023,40(7):917-924.[doi:DOI:10.3969/j.issn.1005-202X.2023.07.020]
点击复制

人工静脉瓣膜周期变化中的力学性能研究()
分享到:

《中国医学物理学杂志》[ISSN:1005-202X/CN:44-1351/R]

卷:
40卷
期数:
2023年第7期
页码:
917-924
栏目:
生物材料与力学
出版日期:
2023-07-15

文章信息/Info

Title:
Mechanical properties during periodical variation of prosthetic venous valve
文章编号:
1005-202X(2023)07-0917-08
作者:
袁瑜含1谷雪莲2崔建昆1陶玉蓉2袁振宇3
1.上海理工大学机械工程学院, 上海 200093; 2.上海理工大学健康科学与工程学院, 上海 200093; 3.上海蓝脉医疗科技有限公司, 上海 201318
Author(s):
YUAN Yuhan1 GU Xuelian2 CUI Jiankun1 TAO Yurong2 YUAN Zhenyu3
1. School of Mechanical Engineering, University of Shanghai for Science and Technology, Shanghai 200093, China 2. School of Health Science and Engineering, University of Shanghai for Science and Technology, Shanghai 200093, China 3. Shanghai Bluevastec Medical Technology Co., Ltd, Shanghai 201318, China
关键词:
人工静脉瓣膜力学性能慢性静脉功能不全有限元分析
Keywords:
Keywords: prosthetic venous valve mechanical property chronic venous insufficiency finite element analysis
分类号:
R318
DOI:
DOI:10.3969/j.issn.1005-202X.2023.07.020
文献标志码:
A
摘要:
目的:研究各参数在人工静脉瓣膜周期变化中对力学性能的影响,为人工静脉瓣膜的设计与优化提供参考依据。方法:采用光滑粒子动力学方法进行双向流固耦合仿真,研究包含2种材料(膨化聚四氟乙烯和猪心包),3种长度(5.0、7.5和10.0 mm)和3种厚度(0.050、0.075和0.100 mm)的瓣膜共18组模型,通过对比张开孔径、最大等效应力及回流情况判断3个因素对瓣膜的影响。结果:瓣膜材料为猪心包时达到更大的张开孔径,最大值为5.90 mm,所受最大等效应力更平稳;在张开阶段当瓣膜长度为7.5 mm、厚度为0.100 mm时,最大等效应力值最小;在关闭阶段瓣膜材料为猪心包,瓣膜长度为10.0 mm、厚度为0.100 mm时,最大等效应力最小;瓣膜厚度为0.075 mm能获得最好的抗回流效果。结论:在一定范围内减小瓣膜长度和厚度,增加瓣膜材料的弹性模量,能增加人工静脉瓣膜周期运动的稳定性。
Abstract:
Objective To investigate the effects of various parameters of prosthetic venous valve on its mechanical properties for providing reference for the design and optimization of prosthetic venous valve. Methods Smooth particle hydrodynamics method was used to evaluate the fluid-structure interaction between valve and blood. There were 18 groups of parameters, including 2 valve materials (ePTFE and porcine pericardium), 3 valve lengths (5.0, 7.5, 10.0 mm) and 3 valve thicknesses (0.050, 0.075, 0.100 mm). The opening aperture, maximum equivalent stress and anti-reflux function were analyzed for determining the effects of the 3 factors on the valve. Results The valve made of porcine pericardium had a larger opening aperture, with a maximum of 5.90 mm, and the maximum equivalent stress on the valve was more stable. When the valve was opening, the maximum equivalent stress value on the valve with a length of 7.5 mm and a thickness of 0.100 mm was the lowest. When the valve was closing, the maximum equivalent stress on the valve with a length of 10.0 mm and a thickness of 0.100 mm was the lowest. A valve thickness of 0.075 mm was best for anti-reflux function. Conclusion Reducing the length and thickness of the valve within a certain range and increasing the elastic modulus of the valve material can increase the stability of prosthetic venous valve.

相似文献/References:

[1]郭景振,宋成利,崔海坡.靶向药物洗脱支架扩张过程力学性能研究[J].中国医学物理学杂志,2015,32(01):129.[doi:10.3969/j.issn.1005-202X.2015.01.029]
[2]刘小利,袁小燕,牛晓东.拉伸法测Mersilk缝合线的力学性能[J].中国医学物理学杂志,2019,36(11):1342.[doi:DOI:10.3969/j.issn.1005-202X.2019.11.018]
 LIU Xiaoli,YUAN Xiaoyan,NIU Xiaodong.Mechanical properties of Mersilk suture measured by stretching method[J].Chinese Journal of Medical Physics,2019,36(7):1342.[doi:DOI:10.3969/j.issn.1005-202X.2019.11.018]
[3]姜豪,云忠,汤晓燕,等.直接心辅装置CRG/PVA复合水凝胶隔离层的制备与性能研究[J].中国医学物理学杂志,2021,38(2):238.[doi:DOI:10.3969/j.issn.1005-202X.2021.02.020]
 JIANG Hao,YUN Zhong,TANG Xiaoyan,et al.Preparation and properties of CRG/PVA composite hydrogel isolation layer for direct cardiac assist device[J].Chinese Journal of Medical Physics,2021,38(7):238.[doi:DOI:10.3969/j.issn.1005-202X.2021.02.020]

备注/Memo

备注/Memo:
【收稿日期】2023-02-08 【基金项目】上海介入医疗器械工程技术研究中心资助项目(18DZ2250900) 【作者简介】袁瑜含,硕士研究生,E-mail: yhyuan_0309@163.com 【通信作者】谷雪莲,副教授,E-mail: guxuelian@usst.edu.cn
更新日期/Last Update: 2023-07-15