[1]袁玲,刘颖,徐瑶瑶,等.SRS MapCHECK与EDOSE在立体定向放疗计划剂量验证中的对比[J].中国医学物理学杂志,2022,39(1):23-27.[doi:DOI:10.3969/j.issn.1005-202X.2022.01.004]
 YUAN Ling,LIU Ying,XU Yaoyao,et al.Comparison of dose verification by SRS MapCHECK versus EDOSE for stereotactic radiotherapy[J].Chinese Journal of Medical Physics,2022,39(1):23-27.[doi:DOI:10.3969/j.issn.1005-202X.2022.01.004]
点击复制

SRS MapCHECK与EDOSE在立体定向放疗计划剂量验证中的对比()
分享到:

《中国医学物理学杂志》[ISSN:1005-202X/CN:44-1351/R]

卷:
39卷
期数:
2022年第1期
页码:
23-27
栏目:
医学放射物理
出版日期:
2022-01-17

文章信息/Info

Title:
Comparison of dose verification by SRS MapCHECK versus EDOSE for stereotactic radiotherapy
文章编号:
1005-202X(2022)01-0023-05
作者:
袁玲刘颖徐瑶瑶张红利张晓鹏王鑫石兴源
广州医科大学附属第五医院放疗科, 广东 广州 510700
Author(s):
YUAN Ling LIU Ying XU Yaoyao ZHANG Hongli ZHANG Xiaopeng WANG Xin SHI Xingyuan
Department of Radiotherapy, the Fifth Affiliated Hospital of Guangzhou Medical University, Guangzhou 510700, China
关键词:
立体定向放疗SRS MapCHECKEDOSEγ通过率剂量验证
Keywords:
Keywords: stereotactic radiotherapy SRS MapCHECK EDOSE gamma passing rate dose verification
分类号:
R811.1
DOI:
DOI:10.3969/j.issn.1005-202X.2022.01.004
文献标志码:
A
摘要:
目的:使用SRS MapCHECK和EDOSE对立体定向放疗(SRT)计划进行剂量验证,并比较它们的验证结果。方法:首先比较两种系统在不同方野(2 cm×2 cm、4 cm×4 cm、6 cm×6 cm、10 cm×10 cm)的γ通过率(3%/3 mm),然后选择29例SRT计划,分析两种系统在不同γ标准下的绝对剂量通过率,最后分析EDOSE建立的4种不同物理模型对SRT计划剂量验证的影响。结果:SRS MapCHECK在所有方野和29例SRT计划的绝对剂量γ通过率均高于EDOSE,两者的验证结果有统计学差异(P<0.05);SRS MapCHECK及EDOSE在29例SRT计划的γ通过率(2%/2 mm)分别是为98.60%±2.14%和96.53%±2.41%。基于EDOSE的不同物理模型验证结果有统计学差异,平均γ通过率(2%/2 mm)偏差为1.8%~5.1%。结论:SRS MapCHECK和EDOSE系统均满足SRT计划剂量验证的要求,且SRS MapCHECK的剂量验证通过率优于EDOSE;不同的EDOSE物理模型对SRT计划剂量验证有影响。
Abstract:
Abstract: Objective To carry out dose verification for stereotactic radiotherapy (SRT) using SRS MapCHECK and EDOSE, and to compare the results of dose verification between SRS MapCHECK and EDOSE. Methods The gamma passing rate (3%/3 mm) for different square fields (2 cm×2 cm, 4 cm×4 cm, 6 cm×6 cm, 10 cm×10 cm) using SRS MapCHECK and EDOSE for dose verification were compared, and then, 29 SRT plan were selected to analyze the absolute dose passing rate between SRS MapCHECK and EDOSE at different gamma criteria. Finally, the effects of 4 different physical models established in EDOSE system on the dose verification for SRT plan were analyzed. Results The gamma passing rates of absolute dose obtained by SRS MapCHECK in all square fields and 29 cases of SRT plan were higher those of EDOSE, and the verification results between two systems were statistically different (P<0.05). The gamma passing rates (2%/2 mm) of SRS MapCHECK and EDOSE for 29 cases of SRT plan were 98.60%±2.14% and 96.53%±2.41%, respectively. The verification results among different EDOSE physical models were statistically different, and the deviation of average gamma passing rate (2%/2 mm) was 1.8%-5.1%. Conclusion Both SRS MapCHECK and EDOSE satisfy the requirements of dose verification for SRT plan, but the passing rate of dose verification of SRS MapCHECK is higher than that of EDOSE. The dose verification for SRT plan can be affected by different EDOSE physical models.

相似文献/References:

[1]任信信,戴建荣,张 岳,等.基于专家库的γ射线立体定向放疗计划优化方法[J].中国医学物理学杂志,2015,32(03):301.[doi:10.3969/j.issn.1005-202X.2015.03.001]
[2]葛瑞刚,解传滨,戴相昆,等.D2SRS两种照射技术与CyberKnife在脑转移瘤立体定向放疗中的剂量学研究[J].中国医学物理学杂志,2020,37(3):265.[doi:DOI:10.3969/j.issn.1005-202X.2020.03.001]
 GE Ruigang,XIE Chuanbin,DAI Xiangkun,et al.Dosimetric comparison of two kinds of radiation techniques of D2SRS and CyberKnife in stereotactic radiotherapy for brain metastases[J].Chinese Journal of Medical Physics,2020,37(1):265.[doi:DOI:10.3969/j.issn.1005-202X.2020.03.001]
[3]刘登洪,王雪桃,钟仁明.几何不确定性鲁棒优化在肝癌立体定向放疗中的应用[J].中国医学物理学杂志,2022,39(2):162.[doi:DOI:10.3969/j.issn.1005-202X.2022.02.006]
 LIU Denghong,WANG Xuetao,ZHONG Renming.Application of geometric uncertainty-based robust optimization in stereotactic body radiotherapy for liver cancer[J].Chinese Journal of Medical Physics,2022,39(1):162.[doi:DOI:10.3969/j.issn.1005-202X.2022.02.006]
[4]祝小惠,梁志文,曹婷,等.SRS MapCHECK半导体矩阵用于射波刀脊柱计划剂量验证的应用研究[J].中国医学物理学杂志,2022,39(4):415.[doi:DOI:10.3969/j.issn.1005-202X.2022.04.004]
 ZHU Xiaohui,LIANG Zhiwen,CAO Ting,et al.Application of SRS MapCHECK semiconductor detector array in dose verification for spine CyberKnife plan[J].Chinese Journal of Medical Physics,2022,39(1):415.[doi:DOI:10.3969/j.issn.1005-202X.2022.04.004]

备注/Memo

备注/Memo:
【收稿日期】2021-07-21 【基金项目】广州市科技计划项目(201905010004) 【作者简介】袁玲,硕士,研究方向:放射物理,E-mail: 1210710050@qq.com 【通信作者】石兴源,研究方向:肿瘤治疗,E-mail: 18549183@qq.com
更新日期/Last Update: 2022-01-17