新型立体定向放射外科治疗设备—Zap-X()
分享到:

《中国医学物理学杂志》[ISSN:1005-202X/CN:44-1351/R]

卷:
39卷
期数:
2022年第1期
页码:
76-80
栏目:
医学信号处理与医学仪器
出版日期:
2022-01-17

文章信息/Info

Title:
A novel stereotactic radiosurgery system Zap-X
文章编号:
1005-202X(2022)01-0076-05
作者:
王金媛1戴相昆1郑庆增2鞠忠建1潘隆盛3曲宝林1王成成1徐寿平1
1.中国人民解放军总医院第一医学中心放射治疗科, 北京 100853; 2.北京老年医院肿瘤放疗科, 北京 100095; 3.中国人民解放军总医院第一医学中心神经外科, 北京 100853
Author(s):
WANG Jinyuan1 DAI Xiangkun1 ZHENG Qingzeng2 JU Zhongjian1 PAN Longsheng3 QU Baolin1 WANG Chengcheng1 XU Shouping1
1. Department of Radiation Oncology, the First Medical Center of the PLA General Hospital, Beijing 100853, China 2. Department of Radiation Oncology, Beijing Geriatric Hospital, Beijing 100095, China 3. Department of Neurosurgery, the First Medical Center of the PLA General Hospital, Beijing 100853, China
关键词:
Zap-X系统立体定向放射外科自屏蔽
Keywords:
Keywords: Zap-X system stereotactic radiosurgery self-shielding
分类号:
R318.6
DOI:
DOI:10.3969/j.issn.1005-202X.2022.01.013
文献标志码:
A
摘要:
精确的剂量照射是立体定向放射外科治疗成功的关键。Zap-X系统是一款具有自屏蔽功能的辐射治疗设备,用于脑部及头颈部肿瘤的立体定向放射外科治疗。本研究针对Zap-X系统的相关技术特点进行简述,分别从机械结构、自屏蔽系统、机械运动系统、辐射准直器系统、治疗计划系统、kV图像引导系统、MV影像验证系统等方面着重介绍了Zap-X系统的组成,阐述该系统的治疗过程,并总结对比其临床应用,最后对该系统的未来发展方向提出展望。其旋转式准直器的设计加上自身屏蔽防护的设计极大地降低了设备安装成本,可移动式MV探测器能实时监测治疗过程中的剂量准确性,采用直线加速器代替放射源,安全性、稳定性得到一定程度的提高。作为系统的初代产品,该设备还有诸多技术环节有待进一步改善。
Abstract:
Abstract: Precise dose delivery is the key to the success of stereotactic radiosurgery. Zap-X system is a kind of radiotherapy equipment equipped with self-shielding function for the stereotactic radiosurgery of tumors in the brain, head and neck. Herein the technical characteristics of Zap-X system are briefly described, introducing the composition of Zap-X system from mechanical structure, self-shielding system, mechanical movement system, radiation collimator system, treatment planning system, kV image guidance system and MV image verification system. The treatment process of the system is explained and its clinical application is evaluated. Finally, the future development of the system is discussed. The designs of the rotating collimator and self-shielding protection greatly reduce the installation cost of the equipment. The removable megavoltage radiation detectors can monitor the dose accuracy during the treatment process in real time. The use of linear accelerator instead of radioactive source makes it safe and stable to a certain extent. As the first-generation product of the system, more technical aspects still need to be further improved.

相似文献/References:

[1]张平,邓官华,戴鹏,等. RapidArc与HybridArc技术在大体积脑转移瘤立体定向放射外科中的剂量学比较[J].中国医学物理学杂志,2019,36(8):887.[doi:DOI:10.3969/j.issn.1005-202X.2019.08.005]
 ZHANG Ping,DENG Guanhua,DAI Peng,et al. Stereotactic radiosurgery for large brain metastases: a dosimetric comparison of volumetric modulated arc therapy generated with RapidArc versus HybridArc[J].Chinese Journal of Medical Physics,2019,36(1):887.[doi:DOI:10.3969/j.issn.1005-202X.2019.08.005]
[2]戴鹏,张平,罗龙辉,等.RapidArc与Novalis Knife在多发脑转移瘤SRS中的剂量学比较[J].中国医学物理学杂志,2020,37(10):1242.[doi:DOI:10.3969/j.issn.1005-202X.2020.10.006]
 DAI Peng,ZHANG Ping,LUO Longhui,et al.Dosimetric comparison between RapidArc and Novalis Knife in stereotactic radiosurgery for multiple brain metastases[J].Chinese Journal of Medical Physics,2020,37(1):1242.[doi:DOI:10.3969/j.issn.1005-202X.2020.10.006]
[3]陈知然,王晓红,李向斌,等.多发脑转移瘤容积旋转调强立体定向放射外科计划对比[J].中国医学物理学杂志,2021,38(12):1482.[doi:DOI:10.3969/j.issn.1005-202X.2021.12.006]
 CHEN Zhiran,WANG Xiaohong,LI Xiangbin,et al.Comparison of volumetric modulated arc therapy stereotactic radiosurgery plans of multiple brain metastases[J].Chinese Journal of Medical Physics,2021,38(1):1482.[doi:DOI:10.3969/j.issn.1005-202X.2021.12.006]
[4]蔡俊涛.立体定向放射外科中图像引导技术的研究现状[J].中国医学物理学杂志,2023,40(7):822.[doi:DOI:10.3969/j.issn.1005-202X.2023.07.005]
 CAI Juntao.Advances in image-guided techniques in stereotactic radiosurgery[J].Chinese Journal of Medical Physics,2023,40(1):822.[doi:DOI:10.3969/j.issn.1005-202X.2023.07.005]

备注/Memo

备注/Memo:
【收稿日期】2021-08-12 【基金项目】国家自然科学基金(81801799) 【作者简介】王金媛,硕士,主管技师,研究方向:医学图像处理及立体定向放射外科,E-mail: kyjinyuan@126.com 【通信作者】徐寿平,博士,高级工程师,研究方向:光子/质子精确放疗和自适应放疗,E-mail: shouping_xu@yahoo.com
更新日期/Last Update: 2022-01-17