方舱CT防护设计快速估算方法研究及防护监测分析()
分享到:

《中国医学物理学杂志》[ISSN:1005-202X/CN:44-1351/R]

卷:
38卷
期数:
2021年第10期
页码:
1209-1215
栏目:
医学放射物理
出版日期:
2021-10-27

文章信息/Info

Title:
A method to rapidly estimate the protective design for CT shelter and analysis of radiation?rotection?onitoring
文章编号:
1005-202X(2021)10-1209-07
作者:
凌玉婷1康立丽1杨绍洲2蓝坤明2高扬1
1.南方医科大学生物医学工程学院, 广东 广州 510515; 2.南方医科大学南方医院增城分院设备信息科, 广东 广州 511356
Author(s):
LING Yuting1 KANG Lili1 YANG Shaozhou2 LAN Kunming2 GAO Yang1
1. School of Biomedical Engineering, Southern Medical University, Guangzhou 510515, China 2. Department of Equipment and Information, Zengcheng Branch of Nanfang Hospital, Southern Medical University, Guangzhou 511356, China
关键词:
方舱CT剂量分布屏蔽物质铅当量防护检测
Keywords:
Keywords: CT shelter dose distribution shielding material lead equivalent protection detection
分类号:
R318
DOI:
DOI:10.3969/j.issn.1005-202X.2021.10.005
文献标志码:
A
摘要:
目的:探讨方便、快捷的屏蔽计算方法用于方舱CT机房的屏蔽设计,使其防护检测结果符合周围剂量当量率要求的国家标准。方法:在不考虑工作负荷情况下,利用周围剂量当量率目标管理值和CT剂量分布曲线进行屏蔽透射因子B的计算,从而快速估算各方向屏蔽体的设计值。根据实际防护检测结果结合机房防护进行防护效果分析。结果:优化了CT设备方舱在不同目标管理值、不同机房布局条件下的快速屏蔽估算方法,在确保施工质量的条件下使其符合《放射诊断放射防护要求》(GBZ 130-2020)对防护的要求。结论:优化的方舱式CT影像设备机房屏蔽计算方法可在不考虑CT工作负荷情况下方便快捷地对各方向屏蔽体进行设计。
Abstract:
Abstract: Objective To discuss a fast and convenient shielding estimation method for CT shelter, making the radiation protection conform to the national standards for peripheral dose equivalent rate. Methods Without regard to work load, the management objective values of peripheral dose equivalent rate and the CT dose distribution curves were used to calculate the shielding transmission factor B, thereby quickly estimating the design value of the shield in every direction. Based on the test results of radiation protection in practice and the protection condition of X-ray room, the protective effect was analyzed. Results The method for the rapid shielding estimation for CT shelters with different management objective values and different layouts was optimized, and the optimized method made CT shelter meet the requirements for radiation protection in "Radiation Diagnostic Radiological Protection Requirements" (GBZ 130-2020) under the guarantee of construction quality. Conclusion The optimized shielding estimation method for CT shelter can realize the rapid design of the shield in every direction, without regard to CT work load.

相似文献/References:

[1]林晓汕,蔡维洵,李伟周,等.应用Matrixx验证活动挡铅形成的电子束照射野[J].中国医学物理学杂志,2014,31(02):4751.[doi:10.3969/j.issn.1005-202X.2014.02.008]
[2]苏善宁,曾自力.中下段食管癌三维适形放疗不同照射方法对剂量分布的影响[J].中国医学物理学杂志,2014,31(06):5249.[doi:10.3969/j.issn.1005-202X.2014.06.006]
[3]王矫健,徐利明,全红,等.多叶准直器叶片到位偏差对调强放射治疗剂量的影响[J].中国医学物理学杂志,2013,30(03):4121.[doi:10.3969/j.issn.1005-202X.2013.03.009]
[4]姜瑞瑶,熊霏,黄国锋,等.基于胶片测量方法研究金属植入物对放疗剂量分布的影响[J].中国医学物理学杂志,2013,30(04):4235.[doi:10.3969/j.issn.1005-202X.2013.04.005]
[5]高艳,刘雅洁,李壮玲,等.鼻咽癌常规二维、三维适形和调强适形放疗颞叶剂量学比较[J].中国医学物理学杂志,2013,30(05):4353.[doi:10.3969/j.issn.1005-202X.2013.05.003]
[6]陈杨生.食管癌放射治疗胸部组织校正与否对剂量学影响的研究[J].中国医学物理学杂志,2013,30(05):4360.[doi:10.3969/j.issn.1005-202X.2013.05.005]
[7]刘凌湘,张 超,邹国荣.乳腺癌改良根治术后逆向与混合调强放疗剂量学比较[J].中国医学物理学杂志,2015,32(05):742.[doi:doi:10.3969/j.issn.1005-202X.2015.05.029]
 [J].Chinese Journal of Medical Physics,2015,32(10):742.[doi:doi:10.3969/j.issn.1005-202X.2015.05.029]
[8]程秀艳,胡逸民.基于一种插值算法的金球位置追踪方法[J].中国医学物理学杂志,2016,33(3):304.[doi:DOI:10.3969/j.issn.1005-202X.2016.03.018]
 [J].Chinese Journal of Medical Physics,2016,33(10):304.[doi:DOI:10.3969/j.issn.1005-202X.2016.03.018]
[9]叶逢松,吴魁,吴爱菊.瓦里安IGRT治疗床对调强放疗剂量的影响[J].中国医学物理学杂志,2016,33(4):430.[doi:10.3969/j.issn.1005-202X.2016.04.022]
 [J].Chinese Journal of Medical Physics,2016,33(10):430.[doi:10.3969/j.issn.1005-202X.2016.04.022]
[10]唐涛,庞皓文,杨波,等.应用Mapcheck验证后装剂量分布[J].中国医学物理学杂志,2016,33(7):687.[doi:10.3969/j.issn.1005-202X.2016.07.009]
 [J].Chinese Journal of Medical Physics,2016,33(10):687.[doi:10.3969/j.issn.1005-202X.2016.07.009]

备注/Memo

备注/Memo:
【收稿日期】2021-05-15 【基金项目】广东省医学科研基金(201711815831371) 【作者简介】凌玉婷,就读于南方医科大学生物医学工程(医学物理师)专业,E-mail: 312307790@qq.com 【通信作者】康立丽,博士,副教授,研究方向:医疗设备质量保证测试与放射防护评价,E-mail: 1035685370@qq.com
更新日期/Last Update: 2021-10-28