[1]黄经纬,曹乐,阚秀,等.基于自适应免疫-最小一乘算法的生物电阻抗特征参数提取[J].中国医学物理学杂志,2021,38(8):1005-1012.[doi:DOI:10.3969/j.issn.1005-202X.2021.08.017]
 HUANG Jingwei,CAO Le,KAN Xiu,et al.Extraction of bioimpedance characteristic parameters based on adaptive immune-least absolute deviation algorithm[J].Chinese Journal of Medical Physics,2021,38(8):1005-1012.[doi:DOI:10.3969/j.issn.1005-202X.2021.08.017]
点击复制

基于自适应免疫-最小一乘算法的生物电阻抗特征参数提取()
分享到:

《中国医学物理学杂志》[ISSN:1005-202X/CN:44-1351/R]

卷:
38卷
期数:
2021年第8期
页码:
1005-1012
栏目:
医学生物物理
出版日期:
2021-08-02

文章信息/Info

Title:
Extraction of bioimpedance characteristic parameters based on adaptive immune-least absolute deviation algorithm
文章编号:
1005-202X(2021)08-1005-08
作者:
黄经纬曹乐阚秀李敏
上海工程技术大学电子电气工程学院, 上海 201620
Author(s):
HUANG Jingwei CAO Le KAN Xiu LI Min
School of Electronic and Electrical Engineering, Shanghai University of Engineering Science, Shanghai 201620, China
关键词:
生物阻抗自适应免疫算法最小一乘法Cole-Cole参数
Keywords:
Keywords: bioimpedance adaptive immune algorithm least absolute deviation algorithm Cole-Cole parameter
分类号:
R318
DOI:
DOI:10.3969/j.issn.1005-202X.2021.08.017
文献标志码:
A
摘要:
针对Cole-Cole参数的提取,本文提出自适应免疫-最小一乘算法。首先,通过制定自适应策略,并增设记忆单元,设计本算法对Cole-Cole参数的提取流程;其次,在标准生物阻抗数据集基础上分别添加3种不同噪声,并设置评价指标,对比分析自适应免疫-最小一乘算法与另外3种算法对Cole-Cole圆弧的拟合精度及Cole-Cole参数的提取结果;最后,基于自适应免疫-最小一乘算法,对其在噪声数据集下提取Cole-Cole参数的耗时长短及初值依赖性做出考量。实验结果表明,该算法相较于另外3种算法能够精准、快速地提取Cole-Cole参数,且对初值的依赖性不高。 【关键词】生物阻抗;自适应免疫算法;最小一乘法;Cole-Cole参数
Abstract:
Abstract: An adaptive immune-least absolute deviation algorithm is proposed for extracting Cole-Cole parameters. The extraction of Cole-Cole parameters by the proposed algorithm is realized by formulating adaptive policies and adding memory units. Then, 3 different kinds of noises are added to the standard data sets of the bioimpedance, and related evaluation indicators were set. Based on the fitting accuracy of Cole-Cole arcs and the extraction results of Cole-Cole parameters, the proposed adaptive immune-least absolute deviation algorithm is compared with other 3 algorithms. Finally, the time-consuming and dependence on the initial values of the proposed algorithm for extracting Cole-Cole parameters from the noise data sets are analyzed. Experimental results show that the Cole-Cole parameters can be extracted accurately and quickly by the proposed algorithm compared with other 3 algorithms, and that the dependence on the initial values is relatively low.

相似文献/References:

[1]谢延媛,周 宇,任彬彬,等.电热闭合过程中肠道生物阻抗变化与闭合效果[J].中国医学物理学杂志,2015,32(05):622.[doi:doi:10.3969/j.issn.1005-202X.2015.05.003]
[2]周延顺,于致远,郭永生,等.便携式多生理参数检测仪的设计[J].中国医学物理学杂志,2024,41(4):486.[doi:DOI:10.3969/j.issn.1005-202X.2024.04.014]
 ZHOU Yanshun,YU Zhiyuan,GUO Yongsheng,et al.Design of a portable multi-physiological parameter detector[J].Chinese Journal of Medical Physics,2024,41(8):486.[doi:DOI:10.3969/j.issn.1005-202X.2024.04.014]

备注/Memo

备注/Memo:
【收稿日期】2020-12-15 【基金项目】国家自然科学基金(61703270) 【作者简介】黄经纬,在读硕士研究生,研究领域:生理信号处理和医疗电子器件研发,E-mail: huangjingweiha@163.com 【通信作者】曹乐,博士,讲师,研究领域:智能物联网应用、医疗设备研发及MEMS器件动态特性等,E-mail: caole00012@163.com
更新日期/Last Update: 2021-07-31