[1]王达飞,黄新阳,许益芬,等.基于红外定位系统的激光灯摆位精度研究[J].中国医学物理学杂志,2021,38(3):308-311.[doi:DOI:10.3969/j.issn.1005-202X.2021.03.007]
 WANG Dafei,HUANG Xinyang,XU Yifen,et al.Setup accuracy of laser system: a study based on optical positioning system[J].Chinese Journal of Medical Physics,2021,38(3):308-311.[doi:DOI:10.3969/j.issn.1005-202X.2021.03.007]
点击复制

基于红外定位系统的激光灯摆位精度研究()
分享到:

《中国医学物理学杂志》[ISSN:1005-202X/CN:44-1351/R]

卷:
38卷
期数:
2021年第3期
页码:
308-311
栏目:
医学放射物理
出版日期:
2021-03-30

文章信息/Info

Title:
Setup accuracy of laser system: a study based on optical positioning system
文章编号:
1005-202X(2021)03-0308-04
作者:
王达飞1黄新阳2许益芬1朱国民1陈胜东1葛云2
1.宜兴市肿瘤医院放疗科, 江苏 宜兴 214206; 2.南京大学电子科学与工程学院, 江苏 南京 210046
Author(s):
WANG Dafei1 HUANG Xinyang2 XU Yifen1 ZHU Guomin1 CHEN Shengdong1 GE Yun2
1. Department of Radiotherapy, Cancer Hospital of Yixing City, Yixing 214206, China 2. School of Electronic Science and Engineering, Nanjing University, Nanjing 210046, China
关键词:
肿瘤放射治疗光学定位系统室内激光系统摆位精度
Keywords:
Keywords: tumor radiotherapy optical positioning system indoor laser system setup accuracy
分类号:
R811.1
DOI:
DOI:10.3969/j.issn.1005-202X.2021.03.007
文献标志码:
A
摘要:
目的:通过光学定位系统(OPS)评估室内激光系统在放射治疗时对不同部位患者的摆位精度。方法:为验证室内激光系统摆位的不稳定性,本研究对象为404例常规放射治疗的头颈部、胸部、腹部和骨盆区域肿瘤患者。所有患者均采用室内激光系统方法摆位,同时记录OPS上显示的X(左右)、Y(前后)、Z(上下)3个方向的摆位误差。结果和结论:室内激光系统在放射治疗时对不同部位患者的摆位精度不同,摆位精度及稳定性较差,尤其是胸部区域,建议应用OPS直接进行放射治疗摆位。
Abstract:
Abstract: Objective To evaluate the setup accuracy of indoor laser system in radiotherapy of tumors at different sites by optical positioning system (OPS). Methods A total of 404 patients with tumors of head and neck, chest, abdomen and pelvic who were treated with conventional radiotherapy were enrolled in the study to verify the instability of indoor laser system for patient setup. The patients positions were adjusted by indoor laser system, and meanwhile the setup errors displaying in OPS in X (left-right), Y (anterior-posterior) and Z (superior-inferior) axes were recorded. Results and Conclusion The setup accuracy of indoor laser system is various in patients with tumors at different sites, and the results reveal that indoor laser system has a lower setup accuracy and poor stability, especially for chest tumors. It is recommended to perform radiotherapy setup directly using OPS.

相似文献/References:

[1]武 杰,袁杭英,欧阳蓓蕾.内旋式放射治疗用面罩固定装置的设计[J].中国医学物理学杂志,2015,32(02):265.[doi:10.3969/j.issn.1005-202X.2015.02.027]
[2]司 森,吴鑫全,包尚联,等.基于GPU的蒙特卡洛放疗剂量模拟[J].中国医学物理学杂志,2015,32(01):1.[doi:10.3969/j.issn.1005-202X.2015.01.001]
[3]王 文,程梦云,杨 琪,等.基于MCNP源子程序的放射治疗剂量计算验证方法[J].中国医学物理学杂志,2015,32(01):13.[doi:10.3969/j.issn.1005-202X.2015.01.004]
[4]张一戈,邱小平,陈维军,等.基于图像引导技术对头颈部肿瘤临床摆位误差的剂量学影响[J].中国医学物理学杂志,2015,32(01):25.[doi:10.3969/j.issn.1005-202X.2015.01.007]
[5]孟慧鹏,孙小喆,王 昊,等.胸部偏后背肿瘤RapidArc采用仰卧位设计治疗计划的可行性分析[J].中国医学物理学杂志,2015,32(01):31.[doi:10.3969/j.issn.1005-202X.2015.01.008]
[6]庞皓文,孙小扬.基于快速自由形变的弹性配准算法在放疗CT中的应用[J].中国医学物理学杂志,2015,32(03):404.[doi:10.3969/j.issn.1005-202X.2015.03.023]
[7]贾卫娟,周一兵,谭维维,等.利用模拟机及挡铅托架进行放疗前的治疗模拟[J].中国医学物理学杂志,2014,31(01):4622.[doi:10.3969/j.issn.1005-202X.2014.01.007]
[8]白 雪,杜向慧,王升晔,等.半野技术结合非共面照射技术在乳腺癌胸锁联合照射中的应用[J].中国医学物理学杂志,2014,31(03):4849.[doi:10.3969/j.issn.1005-202X.2014.03.001]
[9]周 钢,田 野,陆雪官,等.乳腺癌保乳术后三种瘤床同期加量放疗技术的剂量学比较[J].中国医学物理学杂志,2014,31(03):4853.[doi:10.3969/j.issn.1005-202X.2014.03.002]
[10]李贤富,谢 力,郭飞,等.医科达Precise直线加速器快速应急处理方法探讨[J].中国医学物理学杂志,2014,31(04):4984.[doi:10.3969/j.issn.1005-202X.2014.04.003]

备注/Memo

备注/Memo:
【收稿日期】2020-10-14 【作者简介】王达飞,主任医师,主要研究方向:放射治疗,E-mail: dafeiwang@163.com
更新日期/Last Update: 2021-03-30