[1]刘园源,蔡永林.鼻咽癌调强放射治疗中心点Y轴精度分析[J].中国医学物理学杂志,2020,37(8):955-957.[doi:DOI:10.3969/j.issn.1005-202X.2020.08.003]
 LIU Yuanyuan,CAI Yonglin.Y-axis accuracy based on the center point of radiotherapy for nasopharyngeal carcinoma[J].Chinese Journal of Medical Physics,2020,37(8):955-957.[doi:DOI:10.3969/j.issn.1005-202X.2020.08.003]
点击复制

鼻咽癌调强放射治疗中心点Y轴精度分析()
分享到:

《中国医学物理学杂志》[ISSN:1005-202X/CN:44-1351/R]

卷:
37
期数:
2020年第8期
页码:
955-957
栏目:
医学放射物理
出版日期:
2020-08-27

文章信息/Info

Title:
Y-axis accuracy based on the center point of radiotherapy for nasopharyngeal carcinoma
文章编号:
1005-202X(2020)08-0955-03
作者:
刘园源1蔡永林2
1.梧州市红十字会医院放疗一区, 广西 梧州 543002;2.梧州市红十字会医院检验科, 广西 梧州 543002
Author(s):
LIU Yuanyuan1 CAI Yonglin2
1. the First Radiotherapy District, Wuzhou Red Cross Hospital, Wuzhou 543002, China 2. Department of Laboratory, Wuzhou Red Cross Hospital, Wuzhou 543002, China
关键词:
鼻咽癌调强放射治疗治疗中心点Y轴精度
Keywords:
Keywords: nasopharyngeal carcinoma intensity-modulated radiotherapy treatment center Y-axis accuracy
分类号:
R739.6;R811.1
DOI:
DOI:10.3969/j.issn.1005-202X.2020.08.003
文献标志码:
A
摘要:
目的:分析鼻咽癌(NPC)调强放射治疗中心点Y轴的精度情况。方法:选取200例NPC患者为观察对象,将其随机分为A、B两组,每组100例。A组采用双弧容积旋转调强放射治疗(VMAT)逆向优化设计,B组采用传统的调强放射治疗(IMRT)逆向优化设计,两组均采用Eclipse13.5计划系统进行优化设计。比较两组优化设计放疗中心点与重建放射图像中心点Y轴精度情况。结果:A组Y轴移动的发生率为43.00%,B组为58.00%,A组低于B组(P<0.05)。A组Y轴移动的方向向下移动的占比为53.49%,B组Y轴移动方向向上移动占比为68.97%。在摆位误差方面,A组上移摆位误差为(1.07±1.01) mm,B组为(2.15±1.38) mm,A组下移摆位误差为(1.12±1.13) mm,B组为(2.07±1.24) mm,A组均低于B组(P<0.05)。在治疗中心点1层,A组占比为38.00%,B组占比为44.00%,两组无统计学差异(P>0.05)。在治疗中心点2层,A组占比为62.00%,B组占比为56.00%,两组无统计学差异(P>0.05)。中心点层数移动方面,A组移动主要发生在2层,发生率为69.35%。B组1层移动的发生率为36.36%,2层移动的发生率为75.00%。位于治疗中心点1层,B组的移动方向主要是向上移动。位于治疗中心点2层,A组上移的发生率为46.51%,下移的发生率为53.49%。B组在这一层中,主要是上移,发生率为51.72%。结论:NPC调强放射治疗过程中,采用VMAT逆向优化设计更具优势,能够提高调强放射治疗过程中治疗中心点Y轴的精度,具有应用和推广价值。
Abstract:
Abstract: Objective To analyze the Y-axis precision based on the center point of radiotherapy for nasopharyngeal carcinoma (NPC). Methods A total of 200 NPC patients were enrolled and randomly divided into groups A and B, with 100 cases in each group. Double-arc volumetric modulated arc therapy (VMAT) reverse optimization was adopted in group A, while traditional intensity-modulated radiotherapy reverse optimization was adopted in group B. Eclipse13.5 planning system was used for optimization design in two groups. Y-axis accuracy was analyzed based on the center point of optimized radiotherapy and that in reconstruction images. Results The incidence of Y-axis movement in group A was 43.00%, lower than 58.00% in group B (P<0.05). The direction of Y-axis movement in group A was downward, accounting for 53.49%, while the Y-axis in group B was moved upward, accounting for 68.97%. The positioning errors (moving upward) were (1.07±1.01) mm and (2.15±1.38) mm in group A and group B, and the positioning errors (moving downward) were (1.12±1.13) mm and (2.07±1.24) mm in group A and group B (P<0.05). The proportions of treatment center located at layer 1 were 38.00% and 44.00% in group A and group B, respectively, without statistical differences (P>0.05) and the proportions of treatment center located at layer 2 were 62.00% and 56.00%, respectively, without statistical differences (P>0.05). The center point movement in group A mainly occurred at layer 2, with an incidence of 69.35%. The incidence of the center point movement in group B was 36.36% at layer 1 and 75.00% at layer 2. The center point located at layer 1 in group B was mainly moved upward. The incidences of upward movement and downward movement of center point located at layer 2 were 46.51% and 53.49% in group A, and the movement of center point located at the layer 2 in group B was mainly moved upward, with an incidence of 51.72%. Conclusion During the radiotherapy for NPC, VMAT reverse optimization has more advantages and can improve the accuracy of Y-axis of the treatment center during radiotherapy, which is worthy of application and promotion.

相似文献/References:

[1]史贵连,叶福丽.PTW729二维矩阵的平面剂量验证[J].中国医学物理学杂志,2015,32(02):239.[doi:10.3969/j.issn.1005-202X.2015.02.020]
[2]戴相昆,杨丰蔚,杜 镭,等.等效均匀剂量优化法在宫颈癌调强计划中的应用[J].中国医学物理学杂志,2015,32(02):290.[doi:10.3969/j.issn.1005-202X.2015.02.033]
[3]巩汉顺,王金媛,丛小虎,等.晚期胰腺癌适形、简化调强及螺旋断层放疗的剂量学评估[J].中国医学物理学杂志,2015,32(01):8.[doi:10.3969/j.issn.1005-202X.2015.01.003]
[4]曾 彪,张九堂,席许平,等.单中心上下半野技术在胸中上段食管癌放射治疗中的剂量学分布[J].中国医学物理学杂志,2015,32(03):357.[doi:10.3969/j.issn.1005-202X.2015.03.012]
[5]史贵连,叶福丽.肺癌调强放疗计划的设计[J].中国医学物理学杂志,2015,32(03):361.[doi:10.3969/j.issn.1005-202X.2015.03.013]
[6]陈炳周,路俊哲,侯 氢.不同设计方式的调强逆向治疗计划比较[J].中国医学物理学杂志,2015,32(03):391.[doi:10.3969/j.issn.1005-202X.2015.03.020]
[7]周 钢,田 野,陆雪官,等.同步加量技术应用于脑胶质瘤术后调强放疗的剂量学研究[J].中国医学物理学杂志,2014,31(02):4727.[doi:10.3969/j.issn.1005-202X.2014.02.002]
[8]陈广涛,汪 淇,刘亚强,等.TrueBeam加速器射野准直器跟随功能在鼻咽癌固定野调强放疗和容积旋转调强放疗中的应用[J].中国医学物理学杂志,2014,31(03):4865.[doi:10.3969/j.issn.1005-202X.2014.03.005]
[9]宋 威,赵 迪,鹿 红,等.射野方向优化技术在胸中上段食管癌调强放疗中的应用[J].中国医学物理学杂志,2014,31(05):5109.[doi:10.3969/j.issn.1005-202X.2014.05.003]
[10]潘宇辉,陈榕钦,柏朋刚,等.鼻咽癌调强放射治疗前后腮腺体积及空间几何位置变化研究[J].中国医学物理学杂志,2014,31(05):5120.[doi:10.3969/j.issn.1005-202X.2014.05.005]
[11]桑 勇,陈维军,狄小云,等.多叶光栅叶片宽度对鼻咽癌剂量学的影响[J].中国医学物理学杂志,2014,31(06):5240.[doi:10.3969/j.issn.1005-202X.2014.06.004]
[12]黄利生,庄婷婷,伍方财,等.鼻咽癌调强放疗Ib区剂量分析[J].中国医学物理学杂志,2014,31(06):5265.[doi:10.3969/j.issn.1005-202X.2014.06.010]
[13]易金玲,金献测,周永强,等.鼻咽癌IMRT与VMAT治疗的计划与剂量验证比较研究[J].中国医学物理学杂志,2013,30(01):3859.[doi:10.3969/j.issn.1005-202X.2013.01.006]
[14]高艳,刘雅洁,李壮玲,等.鼻咽癌常规二维、三维适形和调强适形放疗颞叶剂量学比较[J].中国医学物理学杂志,2013,30(05):4353.[doi:10.3969/j.issn.1005-202X.2013.05.003]
[15]习勉,梁健,孙文钊,等.鼻咽癌调强放射治疗后放射诱发肿瘤发生风险的预测分析[J].中国医学物理学杂志,2016,33(1):6.[doi:DOI:10.3969/j.issn.1005-202X.2016.01.002]
 [J].Chinese Journal of Medical Physics,2016,33(8):6.[doi:DOI:10.3969/j.issn.1005-202X.2016.01.002]
[16]周祥,张书旭.鼻咽癌自适应放疗时腮腺保护的研究进展[J].中国医学物理学杂志,2016,33(3):258.[doi:DOI:10.3969/j.issn.1005-202X.2016.03.008]
 [J].Chinese Journal of Medical Physics,2016,33(8):258.[doi:DOI:10.3969/j.issn.1005-202X.2016.03.008]
[17]莫小眯,肖爱农,杨立,等.鼻咽癌调强放疗中头颈部轮廓变化对摆位误差的影响[J].中国医学物理学杂志,2016,33(5):468.[doi:10.3969/j.issn.1005-202X.2016.05.008]
 [J].Chinese Journal of Medical Physics,2016,33(8):468.[doi:10.3969/j.issn.1005-202X.2016.05.008]
[18]胡邦,唐承佩,刘小芳,等. 基于等效均匀剂量的目标函数对鼻咽癌调强放疗中颞叶的保护[J].中国医学物理学杂志,2017,34(5):472.[doi:DOI:10.3969/j.issn.1005-202X.2017.05.008]
[19]田伟,戴建荣. 子野小数机器跳数取整对鼻咽癌静态调强放射治疗计划剂量学的影响[J].中国医学物理学杂志,2018,35(10):1128.[doi:DOI:10.3969/j.issn.1005-202X.2018.10.003]
 TIAN Wei,DAI Jianrong. Dosimetric effects of rounding subfield monitor units up for step-and-shoot IMRT plan of nasopharyngeal carcinoma[J].Chinese Journal of Medical Physics,2018,35(8):1128.[doi:DOI:10.3969/j.issn.1005-202X.2018.10.003]
[20]黄家文,张梅芳,刘利彬,等. 鼻咽癌放疗旋转与平移误差的相关性[J].中国医学物理学杂志,2018,35(12):1365.[doi:DOI:10.3969/j.issn.1005-202X.2018.12.001]
 HUANG Jiawen,ZHANG Meifang,LIU Libin,et al. Correlation between rotational errors and translational errors in intensity-modulated radiotherapy for nasopharyngeal carcinoma[J].Chinese Journal of Medical Physics,2018,35(8):1365.[doi:DOI:10.3969/j.issn.1005-202X.2018.12.001]

备注/Memo

备注/Memo:
【收稿日期】2020-02-11 【基金项目】国家自然科学基金(81860601) 【作者简介】刘园源,主治医师,研究方向:鼻咽癌放化疗,E-mail: xy2014121403@163.com
更新日期/Last Update: 2020-08-27