[1]康盛伟,王培,祁国海,等.新型国产二维矩阵剂量验证系统的临床测试[J].中国医学物理学杂志,2020,37(8):945-950.[doi:DOI:10.3969/j.issn.1005-202X.2020.08.001]
 KANG Shengwei,WANG Pei,et al.Clinical testing of a novel domestic two-dimensional matrix dosimetric verification system[J].Chinese Journal of Medical Physics,2020,37(8):945-950.[doi:DOI:10.3969/j.issn.1005-202X.2020.08.001]
点击复制

新型国产二维矩阵剂量验证系统的临床测试()
分享到:

《中国医学物理学杂志》[ISSN:1005-202X/CN:44-1351/R]

卷:
37
期数:
2020年第8期
页码:
945-950
栏目:
医学放射物理
出版日期:
2020-08-27

文章信息/Info

Title:
Clinical testing of a novel domestic two-dimensional matrix dosimetric verification system
文章编号:
1005-202X(2020)08-0945-06
作者:
康盛伟12王培12祁国海12刘操3龚岚3黎杰12肖明勇12唐斌12郎锦义12
1.四川省肿瘤医院[?研究所/四川省癌症防治中心, 四川 成都 610041; 2.电子科技大学医学院, 四川 成都 610041; 3.中国测试技术研究院, 四川 成都 610000
Author(s):
KANG Shengwei1 2 WANG Pei1 2 QI Guohai1 2 LIU Cao3 GONG Lan3 LI Jie1 2 XIAO Mingyong1 2 TANG Bin1 2 LANG Jin-yi1 2
1. Sichuan Cancer Hospital & Institute, Sichuan Cancer Center, Chengdu 610041, China 2. School of Medicine, University of Electronic Science and Technology of China, Chengdu 610041, China 3. National Institute of Measurement and testing technology, Chengdu, 610000, China
关键词:
二维矩阵剂量验证临床测试
Keywords:
Keywords: two-dimensional matrix dose verification clinical testing
分类号:
R318
DOI:
DOI:10.3969/j.issn.1005-202X.2020.08.001
文献标志码:
A
摘要:
目的:对新型国产二维矩阵剂量验证系统在临床条件下进行测试,检验其是否能够满足临床使用需要。方法:参照GB15213-94对用来检测国产二维矩阵剂量验证系统的医用直线加速器进行检测调整,使其达到国家标准。使用新型国产二维矩阵剂量验证系统,对标准照射野下的绝对剂量重复性,标准照射野下的剂量线性,平坦度、对称性,真实病例放疗计划验证进行测试。结果:标准照射野下的绝对剂量重复性检测,其变异系数小于0.7%,符合测试要求;标准照射野下的剂量线性检测与电离室检测结果相比,无明显差异;平坦度检测±3%以内、对称性检测±2%以内,均满足临床使用要求;真实病例计划验证γ通过率均大于98%,完全满足临床放疗计划验证要求。结论:新型国产二维矩阵剂量验证系统具备点剂量、面剂量测量功能,能够对加速器基本剂量性能进行检测,达到临床使用要求;能够实现放疗计划系统的DICOM数据导入,与实际测量结果比较分析,达到临床计划验证要求。
Abstract:
Abstract: Objective To test a new type of domestic two-dimensional (2D) matrix dosimetric verification system in the clinical condition for evaluating whether the system can meet the clinical requirements. Methods The medical linear accelerator which was used to test the domestic 2D matrix dosimetric verification system was firstly tested and adjusted according to GB15213-94, thus ensuring that the accelerator met the national standard. Then the new type of domestic 2D matrix dose verification system is used for testing the repeatability of absolute dose measurement and dose linearity in the standard radiation field, flatness and symmetry, and validating actual radiotherapy plans. Results The test on the repeatability of absolute dose measurement in the standard radiation field showed that the coefficient of variation was less than 0.7%, which was consistent with the test requirements. There was no significant difference in the test results of dose linearity in the standard radiation field between 2D matrix dosimetric verification system and ionization chamber. The test also revealed that both flatness (within ±3%) and symmetry (within ±2%) met the clinical requirements. Moreover, the validation of actual treatment plans showed that the gamma passing rates were all above 98%, which fully satisfied the clinical requirements. Conclusion The novel domestic 2D matrix dosimetric system can be used to not only measure point and planar doses for the evaluation of the basic dose performances of the accelerator, but also realize the import of the DICOM data exported from the treatment planning system which are then compared with actual measurements, thereby meeting the clinical plan verification requirements.

相似文献/References:

[1]张富利,王雅棣,许卫东,等.应用两种三维探测器阵列进行螺旋断层调强放疗计划剂量验证[J].中国医学物理学杂志,2015,32(02):218.[doi:10.3969/j.issn.1005-202X.2015.02.015]
[2]曾 彪,张九堂,席许平.单中心上下半野调强放射治疗的物理剂量验证[J].中国医学物理学杂志,2015,32(01):17.[doi:10.3969/j.issn.1005-202X.2015.01.005]
[3]任 强,王 玉,曹瑞芬,等.基于非晶硅电子射野影像装置的精确剂量刻度方法[J].中国医学物理学杂志,2015,32(01):38.[doi:10.3969/j.issn.1005-202X.2015.01.010]
[4]刘 浩,王 新,李公平,等.用二维电离室矩阵验证宫颈癌腔内放疗剂量分布[J].中国医学物理学杂志,2014,31(04):4988.[doi:10.3969/j.issn.1005-202X.2014.04.004]
[5]张富利,蒋华勇,王雅棣,等.应用一维、二维、三维探测器阵列进行螺旋断层调强放疗计划剂量验证[J].中国医学物理学杂志,2014,31(06):5230.[doi:10.3969/j.issn.1005-202X.2014.06.002]
[6]易金玲,金献测,周永强,等.鼻咽癌IMRT与VMAT治疗的计划与剂量验证比较研究[J].中国医学物理学杂志,2013,30(01):3859.[doi:10.3969/j.issn.1005-202X.2013.01.006]
[7]方明明,周希法,卢绪菁,等.使用Compass系统进行调强验证的一些探讨[J].中国医学物理学杂志,2013,30(01):3870.[doi:10.3969/j.issn.1005-202X.2013.01.008]
[8]陆佳扬,林珠,陈志坚,等.基于Delta4对Truebeam容积调强放疗(VMAT)计划的验证评估[J].中国医学物理学杂志,2013,30(03):4118.[doi:10.3969/j.issn.1005-202X.2013.03.008]
[9]谭丽娜,孙晓欢,马奎,等.三维剂量验证系统Delta4在容积旋转调强计划剂量验证中的应用[J].中国医学物理学杂志,2013,30(06):4497.[doi:10.3969/j.issn.1005-202X.2013.06.007]
[10]陈上河,王 石,吴朝霞,等.基于蒙特卡罗模拟的CyberKnife病人治疗计划剂量验证[J].中国医学物理学杂志,2015,32(04):451.[doi:10.3969/j.issn.1005-202X.2015.04.001]
 [J].Chinese Journal of Medical Physics,2015,32(8):451.[doi:10.3969/j.issn.1005-202X.2015.04.001]

备注/Memo

备注/Memo:
【收稿日期】2020-02-22 【基金项目】四川省卫计委科研课题(17PJ585);国家重大科学仪器设备开发专项(2013YQ090811) 【作者简介】康盛伟,助理研究员,在读博士,从事辐射医学物理研究,E-mail: cn-ksw@qq.com 【通信作者】郎锦义,博士,主任医师,E-mail: langjy610@163.com
更新日期/Last Update: 2020-08-27