[1]方春锋,徐伟,宋圆源,等.两台医科达直线加速器匹配后束流特性比较[J].中国医学物理学杂志,2020,37(1):1-7.[doi:DOI:10.3969/j.issn.1005-202X.2020.01.001]
 FANG Chunfeng,XU Wei,SONG Yuanyuan,et al.Comparison of beam characteristics after beam matching in two Elekta linacs[J].Chinese Journal of Medical Physics,2020,37(1):1-7.[doi:DOI:10.3969/j.issn.1005-202X.2020.01.001]
点击复制

两台医科达直线加速器匹配后束流特性比较()
分享到:

《中国医学物理学杂志》[ISSN:1005-202X/CN:44-1351/R]

卷:
37
期数:
2020年第1期
页码:
1-7
栏目:
医学放射物理
出版日期:
2020-01-25

文章信息/Info

Title:
Comparison of beam characteristics after beam matching in two Elekta linacs
文章编号:
1005-202X(2020)01-0001-06
作者:
方春锋1徐伟12宋圆源1解琨1蒋中凯1王子申1赵治宇1徐寿平12
1.河北一洲肿瘤医院放疗科, 河北 涿州 072750; 2.解放军总医院第一医学中心放疗科, 北京 100853
Author(s):
FANG Chunfeng1 XU Wei12 SONG Yuanyuan1 XIE Kun1 JIANG Zhongkai1 WANG Zishen1 ZHAO Zhiyu1 XU Shouping12
1. Department of Radiation Oncology, Hebei Yizhou Cancer Hospital, Zhuozhou 072750, China; 2. Department of Radiation Oncology, the First Medical Center of Chinese PLA General Hospital, Beijing 100853, China
关键词:
直线加速器束流特性X射线电子束
Keywords:
Keywords: linac beam characteristic X-ray electron beam
分类号:
R811.1;R312
DOI:
DOI:10.3969/j.issn.1005-202X.2020.01.001
文献标志码:
A
摘要:
目的:比较分析两台医科达直线加速器匹配后的束流特性,为临床上实现治疗计划在两台加速器上互换执行提供依据和参考。方法:利用IBA公司Blue Phantom2水箱采集两台加速器X射线束及电子束等相关数据并对其进行比较分析。结果:两台加速器6 MV各射野%dd(10)X偏差在±0.1%之内,10 MV %dd(10)X偏差在±0.3%之内。其相对应射野条件下两档能量束流平坦度与对称性差异均在±1.5%之内,半影的最大绝对偏差为0.5 mm。两档X射线能量60°楔形野的%dd(10)X最大差异为0.8%。电子线Rp、R50、E0方面,两台加速器中各档电子线的差异皆在±1.2%之内,dmax最大绝对偏差为0.9 mm。各能量射野输出因子之间虽有偏差,但差异均较为微小。结论:两台加速器的束流特性显示出良好的数据匹配度,将为治疗计划在两台加速器上实现互换执行提供临床依据。
Abstract:
Abstract: Objective To compare and analyze the beam characteristics after beam matching in two Elekta linacs and provide the basis and reference for realizing the interchangeable implementation of treatment plans on two linacs. Methods The data related to X-ray and electron beams generated by two accelerators were obtained using IBA Blue Phantom2 and the obtained data were compared and analyzed. Results The %dd(10)X deviations for 6 and 10 MV between two accelerators was within ±0.1% and ±0.3%, respectively. The differences in flatness and symmetry between the two energy levels under the corresponding fields were within ±1.5%, and the maximum absolute deviation of the penumbras was 0.5 mm. The maximum difference in %dd(10)X between the two energy levels in 60° wedge fields was 0.8%. The differences of the electron beams generated by two accelerators in the Rp, R50 and E0 were within ±1.2%, and the maximum dmax deviation was 0.9 mm. Although there were certain deviations in the output factors of each field for different energy levels, the deviations were trivial. Conclusion The beam characteristics of two linacs show a good matching of beam data, which will provide a solid foundation and necessary reference for the interchangeable implementation of clinical plans on two linacs.

相似文献/References:

[1]李军,张西志,花威,等.VARIAN直线加速器IGRT全碳纤维治疗床与虚拟治疗床对高能X射线剂量的影响[J].中国医学物理学杂志,2015,32(04):479.[doi:10.3969/j.issn.1005-202X.2015.04.006]
 [J].Chinese Journal of Medical Physics,2015,32(1):479.[doi:10.3969/j.issn.1005-202X.2015.04.006]
[2]岳海振,张艺宝,刘卓伦,等. 直线加速器均整和非均整模式下射线质和射野输出因子的蒙特卡罗模拟与实测值比较[J].中国医学物理学杂志,2016,33(6):548.[doi:DOI:10.3969/j.issn.1005-202X.2016.06.002]
 [J].Chinese Journal of Medical Physics,2016,33(1):548.[doi:DOI:10.3969/j.issn.1005-202X.2016.06.002]
[3]周鹏,胡南,刘岩海,等.医用直线加速器无均整器下6 MV X线能谱特性的蒙特卡罗研究[J].中国医学物理学杂志,2016,33(8):761.[doi:10.3969/j.issn.1005-202X.2016.08.002]
 [J].Chinese Journal of Medical Physics,2016,33(1):761.[doi:10.3969/j.issn.1005-202X.2016.08.002]
[4]喻伟,徐胜,马宏宾,等. 基于IAEA398号报告进行直线加速器输出量校准的实践与探索[J].中国医学物理学杂志,2017,34(3):235.[doi:10.3969/j.issn.1005-202X.2017.03.004]
 [J].Chinese Journal of Medical Physics,2017,34(1):235.[doi:10.3969/j.issn.1005-202X.2017.03.004]
[5]董晓庆,林清,胡杰,等. 病人皮肤剂量监测仪在放射治疗的临床应用[J].中国医学物理学杂志,2017,34(4):363.[doi:DOI:10.3969/j.issn.1005-202X.2017.04.009]
[6]门阔,戴建荣,陈辛元,等. 双能锥形束CT线性混合技术提高相对电子密度的准确性[J].中国医学物理学杂志,2017,34(5):433.[doi:DOI:10.3969/j.issn.1005-202X.2017.05.001]
[7]夏祥1,李永宝,吴朝霞,等. Varian RapidARC直线加速器6 MeV-X线虚拟源模型的建立[J].中国医学物理学杂志,2017,34(6):561.[doi:DOI:10.3969/j.issn.1005-202X.2017.06.005]
 [J].Chinese Journal of Medical Physics,2017,34(1):561.[doi:DOI:10.3969/j.issn.1005-202X.2017.06.005]
[8]闫冰,吴爱东. 非晶硅电子射野影像装置在直线加速器日检中的应用[J].中国医学物理学杂志,2017,34(10):988.[doi:DOI:10.3969/j.issn.1005-202X.2017.10.004]
 [J].Chinese Journal of Medical Physics,2017,34(1):988.[doi:DOI:10.3969/j.issn.1005-202X.2017.10.004]
[9]李建锋,黄顺平. 实验细胞直线加速器照射方法与剂量学探讨[J].中国医学物理学杂志,2017,34(12):1211.[doi:DOI:10.3969/j.issn.1005-202X.2017.12.005]
 LI Jianfeng,HUANG Shunping. Irradiation method of experimental cell on linear accelerator and dosimetry study[J].Chinese Journal of Medical Physics,2017,34(1):1211.[doi:DOI:10.3969/j.issn.1005-202X.2017.12.005]
[10]唐志全,蒋泽,彭旭东,等. 医科达加速器的剂量监测与控制系统[J].中国医学物理学杂志,2018,35(1):9.[doi:DOI:10.3969/j.issn.1005-202X.2018.01.003]
 TANG Zhiquan,JIANG Ze,PENG Xudong,et al. Dose monitoring and control system of Elekta linear accelerator[J].Chinese Journal of Medical Physics,2018,35(1):9.[doi:DOI:10.3969/j.issn.1005-202X.2018.01.003]

备注/Memo

备注/Memo:
【收稿日期】2019-09-20 【基金项目】国家重点研发计划(2017YFC0112105);国家自然科学基金青年科学基金(61601012) 【作者简介】方春锋,物理师,研究方向:医学物理,E-mail: xiaofang0520@126.com 【通信作者】徐寿平,高级工程师,研究方向:肿瘤放射物理与精确放疗,E-mail: shouping_xu@yahoo.com
更新日期/Last Update: 2020-01-14