Preliminary study for automatically verifying treatment isocenter based on markers(PDF)

《中国医学物理学杂志》[ISSN:1005-202X/CN:44-1351/R]

Issue:
2025年第1期
Page:
1-6
Research Field:
医学放射物理
Publishing date:

Info

Title:
Preliminary study for automatically verifying treatment isocenter based on markers
Author(s):
L?Dongxia1 WANG Wenhua2 QIN Wei3 WANG Min3 WEI Xiaowei3 SHI Feiyue3 JIANG Hongbing1 4
1. Department of Clinical Medical Equipment, Nanjing First Hospital, Nanjing Medical University, Nanjing 210006, China 2. Shanghai Datu Medical?echnology, Shanghai 200062, China 3. Tumor Radiotherapy Center, Nanjing First Hospital, Nanjing Medical University, Nanjing 210006, China 4. Nanjing Emergency Medical Center, Nanjing 210003, China
Keywords:
Keywords: radiotherapy metal marker treatment isocenter CT simulator
PACS:
R318;R815.6
DOI:
DOI:10.3969/j.issn.1005-202X.2025.01.001
Abstract:
Abstract: Objective To propose a novel method for verifying the isocenter in radiotherapy based on markers and conduct a preliminary test. Methods A feasibility experiment was conducted on wooden box phantom for radiotherapy resetting. Fifteen groups of displacement data were randomly generated, corresponding to the position deviations of the isocenter in the radiotherapy plan relative to the original isocenter. According to each set of displacement data, with the aid of movable lasers and a CT scanning couch, CT scanning was performed with two sets of markers (3 per set) affixed to the phantom which were corresponding to the original and treatment isocenters, respectively. In the Eclipse treatment planning system, the coordinate data of the original and treatment isocenters were manually verified, and the difference of coordinate data was calculated to obtain the actual displacement value. The treatment isocenter position was finally confirmed by comparing with the actual displacement. In addition, the study attempts to use threshold segmentation algorithm to automatically detect metal markers and obtain coordinate values of the original isocenter on patient-positioned CT images. In the wooden box experiment, the absolute value of the difference between the actual displacement value and the planned displacement value (?) was used to represent the position accuracy of treatment isocenter, and the deviation value obtained with threshold segmentation algorithm for isocenter detection was ?. Results The ? in the X (left-right), Y (superior-inferior) and Z (anterior-superior) directions was (0.83±0.37) mm, 0 (0, 0.5) mm and (0.45±0.29) mm, respectively. The ? in the X, Y and Z directions was (0.46±0.22) mm, 0 (0, 0.5) mm and (0.33±0.29) mm, respectively. The mean values of ? in 3 directions were all lower than 2 mm, within the range of permissible clinical positioning error. Conclusion The method of automatically verifying treatment isocenter position based on markers is feasible, and the study provides a useful reference for radiotherapy resetting using CT simulator.

References:

Memo

Memo:
-
Last Update: 2025-01-19