Technical evaluation of temperature control equipment for extracorporeal membrane oxygenation
Author(s):
HE Liping
Center for Medical Device Evaluation, National Medical Products Administration, Beijing 100076, China
Keywords:
extracorporeal membrane oxygenation temperature control effectiveness safety
PACS:
R318.6
DOI:
10.3969/j.issn.1005-202X.2025.05.014
Abstract:
The study introduces the technical characteristics (product type, usage mode, temperature control mode, use environment, etc.) of temperature control equipment for extracorporeal membrane oxygenation, and evaluates effectiveness and safety of the product based on its technical characteristics. Main contributions of the study are as follows: (1) clarifying key technical indicators for effectiveness and safety, as well as common risk sources; (2) putting forward the requirements for the researches on product performance, cleaning and disinfection, and stability.