Correlation between TQA data trends and output of helical tomotherapy(PDF)
《中国医学物理学杂志》[ISSN:1005-202X/CN:44-1351/R]
- Issue:
- 2020年第9期
- Page:
- 1120-1123
- Research Field:
- 医学人工智能
- Publishing date:
Info
- Title:
- Correlation between TQA data trends and output of helical tomotherapy
- Author(s):
- CHEN Jihong; ZHANG Xiuchun; BAI Penggang; LU Jun; CHEN Wenjuan; CHEN Kaiqiang; LI Qixin; HU Cairong; LIN Jinyong
- Department of Radiation Oncology, Fujian Cancer Hospital/Fujian Medical University Cancer Hospital, Fuzhou 350014, China
- Keywords:
- helical tomotherapy tomotherapy quality assurance static output beam quality
- PACS:
- R318;R811.1
- DOI:
- 10.3969/j.issn.1005-202X.2020.09.008
- Abstract:
- Objective To discuss the relationship between tomotherapy quality assurance (TQA) data trends and output of
helical tomotherapy (HT). Methods The correlations between static output and output energy variation (D20/D10) were
determined based on the retrospective analysis on the parameters and data trends of each module of TQA of HT in recent 3
years. Results The offset parameter of z axis in step-wedge static module had the best correlation with the static output of HT
(r=0.883, P<0.01). The exit detector flatness in basic dosimetry module was the most sensitive to energy changes (r=0.902),
followed by energy differences in step-wedge static module (r=0.897) and energy differences in step-wedge helical module (r
=0.852), and their sensitivities were 2.3×10-4, 3.1×10-4 and 5.7×10-4, respectively. Conclusion TQA can help user to monitor
the changes in the output and energy of HT, thereby carrying out necessary maintenance and dosimetric validation earlier.
Last Update: 2020-09-25