|Table of Contents|

 Test and research on performance of an intravascular ultrasound system(PDF)

《中国医学物理学杂志》[ISSN:1005-202X/CN:44-1351/R]

Issue:
2018年第1期
Page:
70-75
Research Field:
医学信号处理与医学仪器
Publishing date:

Info

Title:
 Test and research on performance of an intravascular ultrasound system
Author(s):
 XUANYUAN Kai1 KE Gang1 HAN Zhile2
 1. Hubei Center for Medical Devices Quality Supervision and Testing, China Food and Drug Administration, Wuhan 430075, China;
2. Suzhou Institute of Biomedical Engineering and Technology, Chinese Academy of Sciences, Suzhou 215163, China
Keywords:
 Keywords: intravascular ultrasound system central frequency resolution test specification
PACS:
TB559;R445.1
DOI:
DOI:10.3969/j.issn.1005-202X.2018.01.013
Abstract:
 Abstract: Nowadays, there is no reference criterion for testing the performance of intravascular ultrasound system (IVUS), a specialized ultrasound diagnostic equipment. Here, with reference to the National Standard of PRC GB10152-2009 for B mode ultrasonic diagnostic equipment, we investigated the testing methods of resolution and center frequency for IVUS. After the testing method was determined, a testing system was established and the test and verification were carried out for the performance indexes such as center frequency and resolution. The results provide a good technical basis to establish the standard test specifications for the manufactures of IVUS.

References:

Memo

Memo:
-
Last Update: 2018-01-25