Effects of design parameters of flexible electrode on electrode performance: a research based on
electromyography signal analysis(PDF)
《中国医学物理学杂志》[ISSN:1005-202X/CN:44-1351/R]
- Issue:
- 2019年第7期
- Page:
- 830-835
- Research Field:
- 医学信号处理与医学仪器
- Publishing date:
Info
- Title:
- Effects of design parameters of flexible electrode on electrode performance: a research based on
electromyography signal analysis
- Author(s):
- LIU Yexiao; XU Yulin; ZHAO Cuilian
- School of Mechatronic Engineering and Automation, Shanghai University, Shanghai 200444, China
- Keywords:
- flexible electrode; surface electromyography; design parameter; time-frequency domain
- PACS:
- R318;TH77
- DOI:
- DOI:10.3969/j.issn.1005-202X.2019.07.016
- Abstract:
- Abstract: Flexible electrode has received wide attention from researchers because of its advantages in wearable surface
electromyography signal acquisition. However, few studies have provided clear guidance and specifications for the design
parameters of flexible electrode. Herein the parameters that may affect the performance of the electrode are analyzed. Five kinds
of flexible dry electrodes of different specifications fabricated by flexible printing process and inkjet printing process are compared
with the commercial wet electrode. The electromyography signal acquisition experimentation based on ulnar wrist extensors is
designed to obtain 100 surface electromyography signal segments. Finally, the time-frequency domain characteristics of the signal
segment are analyzed, thereby determining the effect of the relevant parameters on the electrode performance. The results show
that the active flexible printed circuit board electrodes after increasing the electrode height have a better signal acquisition
performance.