Requirements for technical review of CBCT image-guided radiotherapy system(PDF)
《中国医学物理学杂志》[ISSN:1005-202X/CN:44-1351/R]
- Issue:
- 2018年第9期
- Page:
- 993-996
- Research Field:
- 医学放射物理
- Publishing date:
Info
- Title:
- Requirements for technical review of CBCT image-guided radiotherapy system
- Author(s):
- GU Xiaofang
- Center for Medical Device Evaluation, China Food and Drug Administration, Beijing 100081, China
- Keywords:
- cone beam computed tomography; fan beam computed tomography; image-guided system; technical review; electrical
safety; performance indexes
- PACS:
- R318.6;R197.39
- DOI:
- DOI:10.3969/j.issn.1005-202X.2018.09.001
- Abstract:
- The differences in technical principles and clinical applications between kV-cone beam computed tomography (CBCT)
image-guided radiotherapy system and fan beam CT for diagnosis were analyzed, including differences in data acquisition, image
processing, spatial resolution, imaging dose, image noise, density resolution and so on. The differences in the requirements for
electrical safety and performance indexes in the technical review of CBCT and fan beam CT were further analyzed. Then the
requirements for CBCT user manual and registration unit division were also discussed. This analysis aims to provide some
references for manufacturers to apply for the registration of CBCT image-guided radiotherapy system.
Last Update: 2018-09-28