Analysis of hemolytic performance of continuous blood pump
Author(s):
ZHANG Jinli1 CHENG Yunzhang1 ZHENG Qiwen2
1. School of Medical Instrument and Food Engineering, University of Shanghai for Science and Technology, Shanghai 200093, China; 2. ForQaly Medical (Shanghai) Company Limited, Shanghai 201138, China
Keywords:
continuous blood pump hemolysis computational fluid dynamics hemolysis experiment
PACS:
R318.6
DOI:
10.3969/j.issn.1005-202X.2018.09.017
Abstract:
Hemolytic performance is one of the main factors affecting the reliability of blood pump. Effective evaluation of hemolytic performance provides a strong basis for the design and improvement of blood pump. Herein, the commonly used methods for hemolysis evaluation in the research and development of blood pump were summarized, including hemolysis estimation and hemolysis experiment. Firstly, the hemolysis estimation was described, including the theoretical numerical estimation model and the hemolysis analysis based on the computational fluid dynamics technique. After the model prototype was developed, the hemolysis experiments were introduced, including cardiopulmonary bypass simulation and animal experiment. Secondly, the advantages and limitations of the two methods were compared, and the comparison results showed that hemolysis estimation was preferred in the pre-development of blood pump, while hemolysis experiment was suitable for hemolytic evaluation with model prototype in the middle and later stages. The hemolysis estimation laid the foundation for hemolysis experiment which verified the results of hemolysis estimation. The hemolysis experiment and numerical estimation were mutual connection and mutual promotion, but they could not be completely replaced. Different researchers and research institutes should choose the appropriate evaluation method according to their own conditions. Finally, the problems and the development trends in the evaluation of blood pump hemolysis were discussed.

References:

Memo

Memo:
-
Last Update: 2018-09-28