|Table of Contents|

 Hardware design and implementation of portable electrocardiogram monitoring system(PDF)

《中国医学物理学杂志》[ISSN:1005-202X/CN:44-1351/R]

Issue:
2018年第2期
Page:
210-213
Research Field:
医学信号处理与医学仪器
Publishing date:

Info

Title:
 Hardware design and implementation of portable electrocardiogram monitoring system
Author(s):
 WU Min XIE Yun WU Yang
 School of Automation, Guangdong University of Technology, Guangzhou 510006, China
Keywords:
 analog devices 8232 programmable system-on-chip 4 with Bluetooth low energy portable electrocardiogram signal Bluetooth
PACS:
TP391
DOI:
DOI:10.3969/j.issn.1005-202X.2018.02.018
Abstract:
 A portable electrocardiogram monitoring system based on analog devices 8232 (AD8232) chip and programmable analog circuit of programmable system-on-chip 4 with Bluetooth low energy (PSOC4 BLE) was introduced. The built-in multiple operational amplifiers in AD8232 chip includes pre-amplifier, filter circuit and main amplifier circuit, which accord with the requirements of ECG signal acquisition. PSOC4 BLE programmable chip contains analog and digital conversion, Bluetooth and other programmable analog and digital circuits, which not only makes the hardware design very small, easy to carry, achieve real-time, low-power and wireless transmission function, but also achieves software filtering. The processed waveform verifies the feasibility of the proposed design.

References:

Memo

Memo:
-
Last Update: 2018-01-29